Cargando…

Endovascular Repair of Aortic Arch Aneurysm with Surgeon-Modified Fenestrated Stent Graft

The authors describe a technique of treating patients with aortic arch aneurysm using surgeon-modified fenestrated stent graft (SMFSG). The technique is demonstrated in a 80-year-old patient whose aneurysm was successfully excluded with a SMFSG using Cook Alpha thoracic stent graft. The device was d...

Descripción completa

Detalles Bibliográficos
Autor principal: Manunga, Jesse
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Thieme Medical Publishers 2019
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6377362/
https://www.ncbi.nlm.nih.gov/pubmed/30769366
http://dx.doi.org/10.1055/s-0039-1677810
_version_ 1783395720157462528
author Manunga, Jesse
author_facet Manunga, Jesse
author_sort Manunga, Jesse
collection PubMed
description The authors describe a technique of treating patients with aortic arch aneurysm using surgeon-modified fenestrated stent graft (SMFSG). The technique is demonstrated in a 80-year-old patient whose aneurysm was successfully excluded with a SMFSG using Cook Alpha thoracic stent graft. The device was deployed, removed from its delivery system, and a fenestration created before being mounted back on the delivery system and constrained. It was transitioned through a series of sheaths before being introduced into its original sheath. The device was implanted via a common femoral artery access site; fenestration cannulated from the left brachial artery and bridged with a stent graft.
format Online
Article
Text
id pubmed-6377362
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Thieme Medical Publishers
record_format MEDLINE/PubMed
spelling pubmed-63773622019-02-20 Endovascular Repair of Aortic Arch Aneurysm with Surgeon-Modified Fenestrated Stent Graft Manunga, Jesse Aorta (Stamford) The authors describe a technique of treating patients with aortic arch aneurysm using surgeon-modified fenestrated stent graft (SMFSG). The technique is demonstrated in a 80-year-old patient whose aneurysm was successfully excluded with a SMFSG using Cook Alpha thoracic stent graft. The device was deployed, removed from its delivery system, and a fenestration created before being mounted back on the delivery system and constrained. It was transitioned through a series of sheaths before being introduced into its original sheath. The device was implanted via a common femoral artery access site; fenestration cannulated from the left brachial artery and bridged with a stent graft. Thieme Medical Publishers 2019-02-15 /pmc/articles/PMC6377362/ /pubmed/30769366 http://dx.doi.org/10.1055/s-0039-1677810 Text en https://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Manunga, Jesse
Endovascular Repair of Aortic Arch Aneurysm with Surgeon-Modified Fenestrated Stent Graft
title Endovascular Repair of Aortic Arch Aneurysm with Surgeon-Modified Fenestrated Stent Graft
title_full Endovascular Repair of Aortic Arch Aneurysm with Surgeon-Modified Fenestrated Stent Graft
title_fullStr Endovascular Repair of Aortic Arch Aneurysm with Surgeon-Modified Fenestrated Stent Graft
title_full_unstemmed Endovascular Repair of Aortic Arch Aneurysm with Surgeon-Modified Fenestrated Stent Graft
title_short Endovascular Repair of Aortic Arch Aneurysm with Surgeon-Modified Fenestrated Stent Graft
title_sort endovascular repair of aortic arch aneurysm with surgeon-modified fenestrated stent graft
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6377362/
https://www.ncbi.nlm.nih.gov/pubmed/30769366
http://dx.doi.org/10.1055/s-0039-1677810
work_keys_str_mv AT manungajesse endovascularrepairofaorticarchaneurysmwithsurgeonmodifiedfenestratedstentgraft