Differentiating False Loss of Resistance from True Loss of Resistance While Performing the Epidural Block with the CompuFlo® Epidural Instrument
BACKGROUND: The occurrence of false losses of resistance may be one of the reasons for inadequate or failed epidural block. A CompuFlo® epidural instrument has been introduced to measure the pressure of human tissues in real time at the orifice of a needle and has been used as a tool to identify the...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6378074/ https://www.ncbi.nlm.nih.gov/pubmed/30863447 http://dx.doi.org/10.1155/2019/5185901 |
_version_ | 1783395861440495616 |
---|---|
author | Vaira, Pasquale Camorcia, Michela Palladino, Tiziana Velardo, Matteo Capogna, Giorgio |
author_facet | Vaira, Pasquale Camorcia, Michela Palladino, Tiziana Velardo, Matteo Capogna, Giorgio |
author_sort | Vaira, Pasquale |
collection | PubMed |
description | BACKGROUND: The occurrence of false losses of resistance may be one of the reasons for inadequate or failed epidural block. A CompuFlo® epidural instrument has been introduced to measure the pressure of human tissues in real time at the orifice of a needle and has been used as a tool to identify the epidural space. The aim of this study was to investigate the sensitivity and the specificity of the ability of CompuFlo® to differentiate the false loss of resistance from the true loss of resistance encountered during the epidural space identification procedure. METHOD: We performed epidural block with the CompuFlo® epidural instrument in 120 healthy women who requested labor epidural analgesia. The epidural needle was considered to have reached the epidural space when an increase in pressure (accompanied by an increase in the pitch of the audible tone) was followed by a sudden and sustained drop in pressure for more than 5 seconds accompanied by a sudden decrease in the pitch of the audible tone, resulting in the formation of a low and stable pressure plateau. We evaluate the sensitivity, specificity, and positive and negative predictive values of the ability of CompuFlo® recordings to correctly identify the true LOR from the false LOR. RESULTS: The drop in pressure associated with the epidural space identification was significantly greater than that recorded after the false loss of resistance (73% vs 33%) (P=0.000001). The sensitivity was 0.83, and the AUC was 0.82. DISCUSSION: We have confirmed the ability of CompuFlo® to differentiate the false loss of resistance from the true loss of resistance and established its specificity and sensitivity. CONCLUSION: An easier identification of dubious losses of resistance during the epidural procedure is essential to reduce the number of epidural attempts and/or needle reinsertions with the potential of a reduced risk of accidental dural puncture especially in difficult cases or when the procedure is performed by trainees. |
format | Online Article Text |
id | pubmed-6378074 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Hindawi |
record_format | MEDLINE/PubMed |
spelling | pubmed-63780742019-03-12 Differentiating False Loss of Resistance from True Loss of Resistance While Performing the Epidural Block with the CompuFlo® Epidural Instrument Vaira, Pasquale Camorcia, Michela Palladino, Tiziana Velardo, Matteo Capogna, Giorgio Anesthesiol Res Pract Research Article BACKGROUND: The occurrence of false losses of resistance may be one of the reasons for inadequate or failed epidural block. A CompuFlo® epidural instrument has been introduced to measure the pressure of human tissues in real time at the orifice of a needle and has been used as a tool to identify the epidural space. The aim of this study was to investigate the sensitivity and the specificity of the ability of CompuFlo® to differentiate the false loss of resistance from the true loss of resistance encountered during the epidural space identification procedure. METHOD: We performed epidural block with the CompuFlo® epidural instrument in 120 healthy women who requested labor epidural analgesia. The epidural needle was considered to have reached the epidural space when an increase in pressure (accompanied by an increase in the pitch of the audible tone) was followed by a sudden and sustained drop in pressure for more than 5 seconds accompanied by a sudden decrease in the pitch of the audible tone, resulting in the formation of a low and stable pressure plateau. We evaluate the sensitivity, specificity, and positive and negative predictive values of the ability of CompuFlo® recordings to correctly identify the true LOR from the false LOR. RESULTS: The drop in pressure associated with the epidural space identification was significantly greater than that recorded after the false loss of resistance (73% vs 33%) (P=0.000001). The sensitivity was 0.83, and the AUC was 0.82. DISCUSSION: We have confirmed the ability of CompuFlo® to differentiate the false loss of resistance from the true loss of resistance and established its specificity and sensitivity. CONCLUSION: An easier identification of dubious losses of resistance during the epidural procedure is essential to reduce the number of epidural attempts and/or needle reinsertions with the potential of a reduced risk of accidental dural puncture especially in difficult cases or when the procedure is performed by trainees. Hindawi 2019-02-03 /pmc/articles/PMC6378074/ /pubmed/30863447 http://dx.doi.org/10.1155/2019/5185901 Text en Copyright © 2019 Pasquale Vaira et al. http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Vaira, Pasquale Camorcia, Michela Palladino, Tiziana Velardo, Matteo Capogna, Giorgio Differentiating False Loss of Resistance from True Loss of Resistance While Performing the Epidural Block with the CompuFlo® Epidural Instrument |
title | Differentiating False Loss of Resistance from True Loss of Resistance While Performing the Epidural Block with the CompuFlo® Epidural Instrument |
title_full | Differentiating False Loss of Resistance from True Loss of Resistance While Performing the Epidural Block with the CompuFlo® Epidural Instrument |
title_fullStr | Differentiating False Loss of Resistance from True Loss of Resistance While Performing the Epidural Block with the CompuFlo® Epidural Instrument |
title_full_unstemmed | Differentiating False Loss of Resistance from True Loss of Resistance While Performing the Epidural Block with the CompuFlo® Epidural Instrument |
title_short | Differentiating False Loss of Resistance from True Loss of Resistance While Performing the Epidural Block with the CompuFlo® Epidural Instrument |
title_sort | differentiating false loss of resistance from true loss of resistance while performing the epidural block with the compuflo® epidural instrument |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6378074/ https://www.ncbi.nlm.nih.gov/pubmed/30863447 http://dx.doi.org/10.1155/2019/5185901 |
work_keys_str_mv | AT vairapasquale differentiatingfalselossofresistancefromtruelossofresistancewhileperformingtheepiduralblockwiththecompufloepiduralinstrument AT camorciamichela differentiatingfalselossofresistancefromtruelossofresistancewhileperformingtheepiduralblockwiththecompufloepiduralinstrument AT palladinotiziana differentiatingfalselossofresistancefromtruelossofresistancewhileperformingtheepiduralblockwiththecompufloepiduralinstrument AT velardomatteo differentiatingfalselossofresistancefromtruelossofresistancewhileperformingtheepiduralblockwiththecompufloepiduralinstrument AT capognagiorgio differentiatingfalselossofresistancefromtruelossofresistancewhileperformingtheepiduralblockwiththecompufloepiduralinstrument |