Cargando…

Comparison of analytical performance of i-Smart 300 and pHOx ultra for the accurate determination of pleural fluid pH

BACKGROUND: Pleural fluid pH is an essential test for diagnosing complicated parapneumonic effusion. We evaluated the performance of two blood gas analyzers in measuring pleural fluid pH. METHODS: The i-STAT G3+ (Abbott) was used as a reference analyzer to evaluate the pH values obtained from other...

Descripción completa

Detalles Bibliográficos
Autores principales: Cho, Jooyoung, Kim, Young Sam, Kim, Young Hwan, Lee, Jae-Yeon, Bae, In Cheol, Lee, Sang-Guk, Kim, Jeong-Ho
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6378778/
https://www.ncbi.nlm.nih.gov/pubmed/30815526
http://dx.doi.org/10.1016/j.plabm.2019.e00117
_version_ 1783395987110232064
author Cho, Jooyoung
Kim, Young Sam
Kim, Young Hwan
Lee, Jae-Yeon
Bae, In Cheol
Lee, Sang-Guk
Kim, Jeong-Ho
author_facet Cho, Jooyoung
Kim, Young Sam
Kim, Young Hwan
Lee, Jae-Yeon
Bae, In Cheol
Lee, Sang-Guk
Kim, Jeong-Ho
author_sort Cho, Jooyoung
collection PubMed
description BACKGROUND: Pleural fluid pH is an essential test for diagnosing complicated parapneumonic effusion. We evaluated the performance of two blood gas analyzers in measuring pleural fluid pH. METHODS: The i-STAT G3+ (Abbott) was used as a reference analyzer to evaluate the pH values obtained from other methods: the i-Smart 300 (i-SENS), the pHOx Ultra (Nova Biomedical), using a clot catcher to filter off microclot, and pH indicator paper. Within-device precision was performed using quality control materials. We compared pleural fluid pH (n = 86) by the above methods and analyzed the concordance rate at the level of the medical decision point, pH 7.2. RESULTS: The within-device coefficient of variations of pH were below 0.1% for all blood gas analyzers tested. The slopes of the regression equations for the i-Smart 300, pHOx Ultra, and pH indicator paper against the reference analyzer were 0.850 (95% confidence interval, CI, 0.800–0.896), 0.714 (95% CI, 0.671–0.766), and 1.105 (95% CI, 0.781–1.581), respectively. The kappa values for the i-Smart 300, pHOx Ultra, and pH indicator paper against the reference analyzer were 0.883 (95% CI, 0.656–1.110), 0.739 (95% CI, 0.393–1.084), and 0.464 (95% CI, 0.102–0.826), respectively. CONCLUSIONS: The i-Smart 300 and pHOx Ultra demonstrated good analytical performance and diagnostic accuracy when determining pleural fluid pH compared with that by the i-STAT G3+, whereas the pH indicator paper showed unsatisfactory results.
format Online
Article
Text
id pubmed-6378778
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-63787782019-02-27 Comparison of analytical performance of i-Smart 300 and pHOx ultra for the accurate determination of pleural fluid pH Cho, Jooyoung Kim, Young Sam Kim, Young Hwan Lee, Jae-Yeon Bae, In Cheol Lee, Sang-Guk Kim, Jeong-Ho Pract Lab Med Article BACKGROUND: Pleural fluid pH is an essential test for diagnosing complicated parapneumonic effusion. We evaluated the performance of two blood gas analyzers in measuring pleural fluid pH. METHODS: The i-STAT G3+ (Abbott) was used as a reference analyzer to evaluate the pH values obtained from other methods: the i-Smart 300 (i-SENS), the pHOx Ultra (Nova Biomedical), using a clot catcher to filter off microclot, and pH indicator paper. Within-device precision was performed using quality control materials. We compared pleural fluid pH (n = 86) by the above methods and analyzed the concordance rate at the level of the medical decision point, pH 7.2. RESULTS: The within-device coefficient of variations of pH were below 0.1% for all blood gas analyzers tested. The slopes of the regression equations for the i-Smart 300, pHOx Ultra, and pH indicator paper against the reference analyzer were 0.850 (95% confidence interval, CI, 0.800–0.896), 0.714 (95% CI, 0.671–0.766), and 1.105 (95% CI, 0.781–1.581), respectively. The kappa values for the i-Smart 300, pHOx Ultra, and pH indicator paper against the reference analyzer were 0.883 (95% CI, 0.656–1.110), 0.739 (95% CI, 0.393–1.084), and 0.464 (95% CI, 0.102–0.826), respectively. CONCLUSIONS: The i-Smart 300 and pHOx Ultra demonstrated good analytical performance and diagnostic accuracy when determining pleural fluid pH compared with that by the i-STAT G3+, whereas the pH indicator paper showed unsatisfactory results. Elsevier 2019-01-29 /pmc/articles/PMC6378778/ /pubmed/30815526 http://dx.doi.org/10.1016/j.plabm.2019.e00117 Text en © 2019 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Article
Cho, Jooyoung
Kim, Young Sam
Kim, Young Hwan
Lee, Jae-Yeon
Bae, In Cheol
Lee, Sang-Guk
Kim, Jeong-Ho
Comparison of analytical performance of i-Smart 300 and pHOx ultra for the accurate determination of pleural fluid pH
title Comparison of analytical performance of i-Smart 300 and pHOx ultra for the accurate determination of pleural fluid pH
title_full Comparison of analytical performance of i-Smart 300 and pHOx ultra for the accurate determination of pleural fluid pH
title_fullStr Comparison of analytical performance of i-Smart 300 and pHOx ultra for the accurate determination of pleural fluid pH
title_full_unstemmed Comparison of analytical performance of i-Smart 300 and pHOx ultra for the accurate determination of pleural fluid pH
title_short Comparison of analytical performance of i-Smart 300 and pHOx ultra for the accurate determination of pleural fluid pH
title_sort comparison of analytical performance of i-smart 300 and phox ultra for the accurate determination of pleural fluid ph
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6378778/
https://www.ncbi.nlm.nih.gov/pubmed/30815526
http://dx.doi.org/10.1016/j.plabm.2019.e00117
work_keys_str_mv AT chojooyoung comparisonofanalyticalperformanceofismart300andphoxultrafortheaccuratedeterminationofpleuralfluidph
AT kimyoungsam comparisonofanalyticalperformanceofismart300andphoxultrafortheaccuratedeterminationofpleuralfluidph
AT kimyounghwan comparisonofanalyticalperformanceofismart300andphoxultrafortheaccuratedeterminationofpleuralfluidph
AT leejaeyeon comparisonofanalyticalperformanceofismart300andphoxultrafortheaccuratedeterminationofpleuralfluidph
AT baeincheol comparisonofanalyticalperformanceofismart300andphoxultrafortheaccuratedeterminationofpleuralfluidph
AT leesangguk comparisonofanalyticalperformanceofismart300andphoxultrafortheaccuratedeterminationofpleuralfluidph
AT kimjeongho comparisonofanalyticalperformanceofismart300andphoxultrafortheaccuratedeterminationofpleuralfluidph