Impact of leaks and ventilation parameters on the efficacy of humidifiers during home ventilation for tracheostomized patients: a bench study
BACKGROUND: During invasive ventilation, the upper airway is bypassed and no longer participates in humidification of inspired gases, which is essential to avoid harmful consequences such as endotracheal tube occlusion. In the case of increased air flow, especially in the presence of leaks (intentio...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6379988/ https://www.ncbi.nlm.nih.gov/pubmed/30777036 http://dx.doi.org/10.1186/s12890-019-0812-z |
_version_ | 1783396227386179584 |
---|---|
author | Haziot, Noémie Ibrahim, Mohamed Zhu, Kaixian Thevenin, Charles-Philippe Hardy, Sebastien Gonzalez-Bermejo, Jésus |
author_facet | Haziot, Noémie Ibrahim, Mohamed Zhu, Kaixian Thevenin, Charles-Philippe Hardy, Sebastien Gonzalez-Bermejo, Jésus |
author_sort | Haziot, Noémie |
collection | PubMed |
description | BACKGROUND: During invasive ventilation, the upper airway is bypassed and no longer participates in humidification of inspired gases, which is essential to avoid harmful consequences such as endotracheal tube occlusion. In the case of increased air flow, especially in the presence of leaks (intentional or unintentional), the humidification provided by humidifiers may become ineffective. The objective of this bench study was to evaluate the quality of humidification provided by heated humidifiers under various home ventilation conditions. METHODS: Five heated humidifiers were tested in eight configurations combining circuit (expiratory valve or vented circuit), tidal volume (600 or 1000 mL) and presence of unintentional leak. Absolute humidity (AH) was measured at the upstream of the test lungs, which were placed in a 34 °C environmental chamber in order to simulate body temperature. RESULTS: The AH measured in the valve circuit ranged between 30 mg/L and 40 mg/L and three out of the five humidifiers achieved an AH higher than the recommended level (33 mg/L). With the vented circuit without unintentional leak, when tidal volume was set at 600 mL, all humidifiers reached an AH higher than 33 mg/L except one device; when the tidal volume was set at 1000 mL and unintentional leak was present, four out of the five humidifiers provided an AH lower than 33 mg/L. CONCLUSION: This study shows that, except under certain home ventilation conditions, such as high tidal volumes with unintentional leak in vented circuit, most heated humidifiers ensure sufficient humidification to avoid the risk of side effect in patients. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (10.1186/s12890-019-0812-z) contains supplementary material, which is available to authorized users. |
format | Online Article Text |
id | pubmed-6379988 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-63799882019-02-28 Impact of leaks and ventilation parameters on the efficacy of humidifiers during home ventilation for tracheostomized patients: a bench study Haziot, Noémie Ibrahim, Mohamed Zhu, Kaixian Thevenin, Charles-Philippe Hardy, Sebastien Gonzalez-Bermejo, Jésus BMC Pulm Med Research Article BACKGROUND: During invasive ventilation, the upper airway is bypassed and no longer participates in humidification of inspired gases, which is essential to avoid harmful consequences such as endotracheal tube occlusion. In the case of increased air flow, especially in the presence of leaks (intentional or unintentional), the humidification provided by humidifiers may become ineffective. The objective of this bench study was to evaluate the quality of humidification provided by heated humidifiers under various home ventilation conditions. METHODS: Five heated humidifiers were tested in eight configurations combining circuit (expiratory valve or vented circuit), tidal volume (600 or 1000 mL) and presence of unintentional leak. Absolute humidity (AH) was measured at the upstream of the test lungs, which were placed in a 34 °C environmental chamber in order to simulate body temperature. RESULTS: The AH measured in the valve circuit ranged between 30 mg/L and 40 mg/L and three out of the five humidifiers achieved an AH higher than the recommended level (33 mg/L). With the vented circuit without unintentional leak, when tidal volume was set at 600 mL, all humidifiers reached an AH higher than 33 mg/L except one device; when the tidal volume was set at 1000 mL and unintentional leak was present, four out of the five humidifiers provided an AH lower than 33 mg/L. CONCLUSION: This study shows that, except under certain home ventilation conditions, such as high tidal volumes with unintentional leak in vented circuit, most heated humidifiers ensure sufficient humidification to avoid the risk of side effect in patients. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (10.1186/s12890-019-0812-z) contains supplementary material, which is available to authorized users. BioMed Central 2019-02-18 /pmc/articles/PMC6379988/ /pubmed/30777036 http://dx.doi.org/10.1186/s12890-019-0812-z Text en © The Author(s). 2019 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Article Haziot, Noémie Ibrahim, Mohamed Zhu, Kaixian Thevenin, Charles-Philippe Hardy, Sebastien Gonzalez-Bermejo, Jésus Impact of leaks and ventilation parameters on the efficacy of humidifiers during home ventilation for tracheostomized patients: a bench study |
title | Impact of leaks and ventilation parameters on the efficacy of humidifiers during home ventilation for tracheostomized patients: a bench study |
title_full | Impact of leaks and ventilation parameters on the efficacy of humidifiers during home ventilation for tracheostomized patients: a bench study |
title_fullStr | Impact of leaks and ventilation parameters on the efficacy of humidifiers during home ventilation for tracheostomized patients: a bench study |
title_full_unstemmed | Impact of leaks and ventilation parameters on the efficacy of humidifiers during home ventilation for tracheostomized patients: a bench study |
title_short | Impact of leaks and ventilation parameters on the efficacy of humidifiers during home ventilation for tracheostomized patients: a bench study |
title_sort | impact of leaks and ventilation parameters on the efficacy of humidifiers during home ventilation for tracheostomized patients: a bench study |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6379988/ https://www.ncbi.nlm.nih.gov/pubmed/30777036 http://dx.doi.org/10.1186/s12890-019-0812-z |
work_keys_str_mv | AT haziotnoemie impactofleaksandventilationparametersontheefficacyofhumidifiersduringhomeventilationfortracheostomizedpatientsabenchstudy AT ibrahimmohamed impactofleaksandventilationparametersontheefficacyofhumidifiersduringhomeventilationfortracheostomizedpatientsabenchstudy AT zhukaixian impactofleaksandventilationparametersontheefficacyofhumidifiersduringhomeventilationfortracheostomizedpatientsabenchstudy AT thevenincharlesphilippe impactofleaksandventilationparametersontheefficacyofhumidifiersduringhomeventilationfortracheostomizedpatientsabenchstudy AT hardysebastien impactofleaksandventilationparametersontheefficacyofhumidifiersduringhomeventilationfortracheostomizedpatientsabenchstudy AT gonzalezbermejojesus impactofleaksandventilationparametersontheefficacyofhumidifiersduringhomeventilationfortracheostomizedpatientsabenchstudy |