Equipment failure of intravenous syringe pump detected by increase in Narcotrend stage: A case report
RATIONALE: Awareness is the recovery of consciousness during general anesthesia. It occurs when patients under general anesthesia receive inadequate anesthetic medications to maintain unconsciousness during surgery. Equipment failure is a common cause of intraoperative awareness. PATIENT CONCERNS: A...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer Health
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6392940/ https://www.ncbi.nlm.nih.gov/pubmed/30461615 http://dx.doi.org/10.1097/MD.0000000000013174 |
_version_ | 1783398588122923008 |
---|---|
author | Niu, Kai Guo, Chunjie Han, Chunying Teng, Shiyong |
author_facet | Niu, Kai Guo, Chunjie Han, Chunying Teng, Shiyong |
author_sort | Niu, Kai |
collection | PubMed |
description | RATIONALE: Awareness is the recovery of consciousness during general anesthesia. It occurs when patients under general anesthesia receive inadequate anesthetic medications to maintain unconsciousness during surgery. Equipment failure is a common cause of intraoperative awareness. PATIENT CONCERNS: A 16-year-old boy, 85 kg in weight, was admitted to our hospital for thyroglossal cystectomy under general anesthesia. Six minutes after the intubation, we noted that the Narcotrend index indicated a condition of light anesthesia and the patient was observed to be in tears. DIAGNOSIS: Improper positioning of the syringe fixing clamp on the CP700TCI infusion pump caused equipment failure and light anesthesia. INTERVENTIONS: Bolus of 50 mg propofol and 2 mg midazolam were administered manually by syringe, and inhalation of 2% sevoflurane was supplemented. Infusion pump was replaced. OUTCOMES: The Narcotrend index of the patient returned to state of deep anesthesia following manual administration of the anesthetic medications. Following the surgery, the patient had an uneventful recovery, and did not present with evidence of awareness. LESSONS: Users of the CP700TCI syringe pump should pay attention to the position of the syringe fixing clamp. Anesthesiologists should check all the equipment according to a defined checklist prior to anesthesia. Narcotrend monitor could help to detect light anesthesia and prevent potential awareness. |
format | Online Article Text |
id | pubmed-6392940 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Wolters Kluwer Health |
record_format | MEDLINE/PubMed |
spelling | pubmed-63929402019-03-15 Equipment failure of intravenous syringe pump detected by increase in Narcotrend stage: A case report Niu, Kai Guo, Chunjie Han, Chunying Teng, Shiyong Medicine (Baltimore) Research Article RATIONALE: Awareness is the recovery of consciousness during general anesthesia. It occurs when patients under general anesthesia receive inadequate anesthetic medications to maintain unconsciousness during surgery. Equipment failure is a common cause of intraoperative awareness. PATIENT CONCERNS: A 16-year-old boy, 85 kg in weight, was admitted to our hospital for thyroglossal cystectomy under general anesthesia. Six minutes after the intubation, we noted that the Narcotrend index indicated a condition of light anesthesia and the patient was observed to be in tears. DIAGNOSIS: Improper positioning of the syringe fixing clamp on the CP700TCI infusion pump caused equipment failure and light anesthesia. INTERVENTIONS: Bolus of 50 mg propofol and 2 mg midazolam were administered manually by syringe, and inhalation of 2% sevoflurane was supplemented. Infusion pump was replaced. OUTCOMES: The Narcotrend index of the patient returned to state of deep anesthesia following manual administration of the anesthetic medications. Following the surgery, the patient had an uneventful recovery, and did not present with evidence of awareness. LESSONS: Users of the CP700TCI syringe pump should pay attention to the position of the syringe fixing clamp. Anesthesiologists should check all the equipment according to a defined checklist prior to anesthesia. Narcotrend monitor could help to detect light anesthesia and prevent potential awareness. Wolters Kluwer Health 2018-11-21 /pmc/articles/PMC6392940/ /pubmed/30461615 http://dx.doi.org/10.1097/MD.0000000000013174 Text en Copyright © 2018 the Author(s). Published by Wolters Kluwer Health, Inc. http://creativecommons.org/licenses/by-nc/4.0 This is an open access article distributed under the terms of the Creative Commons Attribution-Non Commercial License 4.0 (CCBY-NC), where it is permissible to download, share, remix, transform, and buildup the work provided it is properly cited. The work cannot be used commercially without permission from the journal. http://creativecommons.org/licenses/by-nc/4.0 |
spellingShingle | Research Article Niu, Kai Guo, Chunjie Han, Chunying Teng, Shiyong Equipment failure of intravenous syringe pump detected by increase in Narcotrend stage: A case report |
title | Equipment failure of intravenous syringe pump detected by increase in Narcotrend stage: A case report |
title_full | Equipment failure of intravenous syringe pump detected by increase in Narcotrend stage: A case report |
title_fullStr | Equipment failure of intravenous syringe pump detected by increase in Narcotrend stage: A case report |
title_full_unstemmed | Equipment failure of intravenous syringe pump detected by increase in Narcotrend stage: A case report |
title_short | Equipment failure of intravenous syringe pump detected by increase in Narcotrend stage: A case report |
title_sort | equipment failure of intravenous syringe pump detected by increase in narcotrend stage: a case report |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6392940/ https://www.ncbi.nlm.nih.gov/pubmed/30461615 http://dx.doi.org/10.1097/MD.0000000000013174 |
work_keys_str_mv | AT niukai equipmentfailureofintravenoussyringepumpdetectedbyincreaseinnarcotrendstageacasereport AT guochunjie equipmentfailureofintravenoussyringepumpdetectedbyincreaseinnarcotrendstageacasereport AT hanchunying equipmentfailureofintravenoussyringepumpdetectedbyincreaseinnarcotrendstageacasereport AT tengshiyong equipmentfailureofintravenoussyringepumpdetectedbyincreaseinnarcotrendstageacasereport |