A pilot study of improvised CPAP (iCPAP) via face mask for the treatment of adult respiratory distress in low-resource settings

BACKGROUND: Continuous positive airway pressure (CPAP) is a mode of non-invasive ventilation used to treat a variety of respiratory conditions in the emergency department and intensive care unit. In low-resource settings where ventilators are not available, the ability to improvise a CPAP system fro...

Descripción completa

Detalles Bibliográficos
Autores principales: Milliner, Brendan H., Bentley, Suzanne, DuCanto, James
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6399909/
https://www.ncbi.nlm.nih.gov/pubmed/31179948
http://dx.doi.org/10.1186/s12245-019-0224-0
Descripción
Sumario:BACKGROUND: Continuous positive airway pressure (CPAP) is a mode of non-invasive ventilation used to treat a variety of respiratory conditions in the emergency department and intensive care unit. In low-resource settings where ventilators are not available, the ability to improvise a CPAP system from locally available equipment would provide a previously unavailable means of respiratory support for patients in respiratory distress. This manuscript details the design of such a system and its performance in healthy volunteers. METHODS: An improvised CPAP system was assembled from standard emergency department equipment and tested in 10 healthy volunteers (6 male, 4 female; ages 29–33). The system utilizes a water seal and high-flow air to create airway pressure; it was set to provide a pressure of 5 cmH2O for the purposes of this pilot study. Subjects used the system in a monitored setting for 30 min. Airway pressure, heart rate, oxygen saturation, and end-tidal CO2 were monitored. Comfort with the device was assessed via questionnaire. RESULTS: The system maintained positive airway pressure for the full trial period in all subjects, with a mean expiratory pressure (EP) of 5.1 cmH2O (SD 0.7) and mean inspiratory pressure (IP) of 3.2 cmH2O (SD 0.8). There was a small decrease in average EP (5.28 vs 4.88 cmH2O, p = 0.03) and a trend toward decreasing IP (3.26 vs 3.07 cmH2O, p = 0.22) during the trial. No significant change in heart rate, O2 saturation, respiratory rate, or end-tidal CO2 was observed. The system was well tolerated, ranked an average of 4.0 on a 1–5 scale for comfort (with 5 = very comfortable). CONCLUSIONS: This improvised CPAP system maintained positive airway pressure for 30 min in healthy volunteers. Use did not cause tachycardia, hypoxia, or hypoventilation and was well tolerated. This system may be a useful adjunctive treatment for respiratory distress in low-resource settings. Further research should test this system in settings where other positive pressure modalities are not available. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (10.1186/s12245-019-0224-0) contains supplementary material, which is available to authorized users.