LUCAS Versus Manual Chest Compression During Ambulance Transport: A Hemodynamic Study in a Porcine Model of Cardiac Arrest
BACKGROUND: Mechanical chest compression (CC) is currently suggested to deliver sustained high‐quality CC in a moving ambulance. This study compared the hemodynamic support provided by a mechanical piston device or manual CC during ambulance transport in a porcine model of cardiopulmonary resuscitat...
Autores principales: | , , , , , , , , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6405722/ https://www.ncbi.nlm.nih.gov/pubmed/30590977 http://dx.doi.org/10.1161/JAHA.118.011189 |
Sumario: | BACKGROUND: Mechanical chest compression (CC) is currently suggested to deliver sustained high‐quality CC in a moving ambulance. This study compared the hemodynamic support provided by a mechanical piston device or manual CC during ambulance transport in a porcine model of cardiopulmonary resuscitation. METHODS AND RESULTS: In a simulated urban ambulance transport, 16 pigs in cardiac arrest were randomized to 18 minutes of mechanical CC with the LUCAS (n=8) or manual CC (n=8). ECG, arterial and right atrial pressure, together with end‐tidal CO(2) and transthoracic impedance curve were continuously recorded. Arterial lactate was assessed during cardiopulmonary resuscitation and after resuscitation. During the initial 3 minutes of cardiopulmonary resuscitation, the ambulance was stationary, while then proceeded along a predefined itinerary. When the ambulance was stationary, CC‐generated hemodynamics were equivalent in the 2 groups. However, during ambulance transport, arterial and coronary perfusion pressure, and end‐tidal CO(2) were significantly higher with mechanical CC compared with manual CC (coronary perfusion pressure: 43±4 versus 18±4 mmHg; end‐tidal CO(2): 31±2 versus 19±2 mmHg, P<0.01 at 18 minutes). During cardiopulmonary resuscitation, arterial lactate was lower with mechanical CC compared with manual CC (6.6±0.4 versus 8.2±0.5 mmol/L, P<0.01). During transport, mechanical CC showed greater constancy compared with the manual CC, as represented by a higher CC fraction and a lower transthoracic impedance curve variability (P<0.01). All animals in the mechanical CC group and 6 (75%) in the manual one were successfully resuscitated. CONCLUSIONS: This model adds evidence in favor of the use of mechanical devices to provide ongoing high‐quality CC and tissue perfusion during ambulance transport. |
---|