A multi‐institutional evaluation of machine performance check system on treatment beam output and symmetry using statistical process control

BACKGROUND: The automated and integrated machine performance check (MPC) tool was verified against independent detectors to evaluate its beam uniformity and output detection abilities to consider it suitable for daily quality assurance (QA). METHODS: Measurements were carried out on six linear accel...

Descripción completa

Detalles Bibliográficos
Autores principales: Binny, Diana, Aland, Trent, Archibald‐Heeren, Ben R., Trapp, Jamie V., Kairn, Tanya, Crowe, Scott B.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6414149/
https://www.ncbi.nlm.nih.gov/pubmed/30786139
http://dx.doi.org/10.1002/acm2.12547
_version_ 1783402936183816192
author Binny, Diana
Aland, Trent
Archibald‐Heeren, Ben R.
Trapp, Jamie V.
Kairn, Tanya
Crowe, Scott B.
author_facet Binny, Diana
Aland, Trent
Archibald‐Heeren, Ben R.
Trapp, Jamie V.
Kairn, Tanya
Crowe, Scott B.
author_sort Binny, Diana
collection PubMed
description BACKGROUND: The automated and integrated machine performance check (MPC) tool was verified against independent detectors to evaluate its beam uniformity and output detection abilities to consider it suitable for daily quality assurance (QA). METHODS: Measurements were carried out on six linear accelerators (each located at six individual sites) using clinically available photon and electron energies for a period up to 12 months (n = 350). Daily constancy checks on beam symmetry and output were compared against independent devices such as the SNC Daily QA 3, PTW Farmer ionization chamber, and SNC field size QA phantom. MPC uniformity detection of beam symmetry adjustments was also assessed. Sensitivity of symmetry and output measurements were assessed using statistical process control (SPC) methods to derive tolerances for daily machine QA and baseline resets to account for drifts in output readings. I‐charts were used to evaluate systematic and nonsystematic trends to improve error detection capabilities based on calculated upper and lower control levels (UCL/LCL) derived using standard deviations from the mean dataset. RESULTS: This study investigated the vendor's method of uniformity detection. Calculated mean uniformity variations were within ± 0.5% of Daily QA 3 vertical symmetry measurements. Mean MPC output variations were within ± 1.5% of Daily QA 3 and ±0.5% of Farmer ionization chamber detected variations. SPC calculated UCL values were a measure of change observed in the output detected for both MPC and Daily QA 3. CONCLUSIONS: Machine performance check was verified as a daily quality assurance tool to check machine output and symmetry while assessing against an independent detector on a weekly basis. MPC output detection can be improved by regular SPC‐based trend analysis to measure drifts in the inherent device and control systematic and random variations thereby increasing confidence in its capabilities as a QA device. A 3‐monthly MPC calibration assessment was recommended based on SPC capability and acceptability calculations.
format Online
Article
Text
id pubmed-6414149
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-64141492019-03-22 A multi‐institutional evaluation of machine performance check system on treatment beam output and symmetry using statistical process control Binny, Diana Aland, Trent Archibald‐Heeren, Ben R. Trapp, Jamie V. Kairn, Tanya Crowe, Scott B. J Appl Clin Med Phys Radiation Oncology Physics BACKGROUND: The automated and integrated machine performance check (MPC) tool was verified against independent detectors to evaluate its beam uniformity and output detection abilities to consider it suitable for daily quality assurance (QA). METHODS: Measurements were carried out on six linear accelerators (each located at six individual sites) using clinically available photon and electron energies for a period up to 12 months (n = 350). Daily constancy checks on beam symmetry and output were compared against independent devices such as the SNC Daily QA 3, PTW Farmer ionization chamber, and SNC field size QA phantom. MPC uniformity detection of beam symmetry adjustments was also assessed. Sensitivity of symmetry and output measurements were assessed using statistical process control (SPC) methods to derive tolerances for daily machine QA and baseline resets to account for drifts in output readings. I‐charts were used to evaluate systematic and nonsystematic trends to improve error detection capabilities based on calculated upper and lower control levels (UCL/LCL) derived using standard deviations from the mean dataset. RESULTS: This study investigated the vendor's method of uniformity detection. Calculated mean uniformity variations were within ± 0.5% of Daily QA 3 vertical symmetry measurements. Mean MPC output variations were within ± 1.5% of Daily QA 3 and ±0.5% of Farmer ionization chamber detected variations. SPC calculated UCL values were a measure of change observed in the output detected for both MPC and Daily QA 3. CONCLUSIONS: Machine performance check was verified as a daily quality assurance tool to check machine output and symmetry while assessing against an independent detector on a weekly basis. MPC output detection can be improved by regular SPC‐based trend analysis to measure drifts in the inherent device and control systematic and random variations thereby increasing confidence in its capabilities as a QA device. A 3‐monthly MPC calibration assessment was recommended based on SPC capability and acceptability calculations. John Wiley and Sons Inc. 2019-02-20 /pmc/articles/PMC6414149/ /pubmed/30786139 http://dx.doi.org/10.1002/acm2.12547 Text en © 2019 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Binny, Diana
Aland, Trent
Archibald‐Heeren, Ben R.
Trapp, Jamie V.
Kairn, Tanya
Crowe, Scott B.
A multi‐institutional evaluation of machine performance check system on treatment beam output and symmetry using statistical process control
title A multi‐institutional evaluation of machine performance check system on treatment beam output and symmetry using statistical process control
title_full A multi‐institutional evaluation of machine performance check system on treatment beam output and symmetry using statistical process control
title_fullStr A multi‐institutional evaluation of machine performance check system on treatment beam output and symmetry using statistical process control
title_full_unstemmed A multi‐institutional evaluation of machine performance check system on treatment beam output and symmetry using statistical process control
title_short A multi‐institutional evaluation of machine performance check system on treatment beam output and symmetry using statistical process control
title_sort multi‐institutional evaluation of machine performance check system on treatment beam output and symmetry using statistical process control
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6414149/
https://www.ncbi.nlm.nih.gov/pubmed/30786139
http://dx.doi.org/10.1002/acm2.12547
work_keys_str_mv AT binnydiana amultiinstitutionalevaluationofmachineperformancechecksystemontreatmentbeamoutputandsymmetryusingstatisticalprocesscontrol
AT alandtrent amultiinstitutionalevaluationofmachineperformancechecksystemontreatmentbeamoutputandsymmetryusingstatisticalprocesscontrol
AT archibaldheerenbenr amultiinstitutionalevaluationofmachineperformancechecksystemontreatmentbeamoutputandsymmetryusingstatisticalprocesscontrol
AT trappjamiev amultiinstitutionalevaluationofmachineperformancechecksystemontreatmentbeamoutputandsymmetryusingstatisticalprocesscontrol
AT kairntanya amultiinstitutionalevaluationofmachineperformancechecksystemontreatmentbeamoutputandsymmetryusingstatisticalprocesscontrol
AT crowescottb amultiinstitutionalevaluationofmachineperformancechecksystemontreatmentbeamoutputandsymmetryusingstatisticalprocesscontrol
AT binnydiana multiinstitutionalevaluationofmachineperformancechecksystemontreatmentbeamoutputandsymmetryusingstatisticalprocesscontrol
AT alandtrent multiinstitutionalevaluationofmachineperformancechecksystemontreatmentbeamoutputandsymmetryusingstatisticalprocesscontrol
AT archibaldheerenbenr multiinstitutionalevaluationofmachineperformancechecksystemontreatmentbeamoutputandsymmetryusingstatisticalprocesscontrol
AT trappjamiev multiinstitutionalevaluationofmachineperformancechecksystemontreatmentbeamoutputandsymmetryusingstatisticalprocesscontrol
AT kairntanya multiinstitutionalevaluationofmachineperformancechecksystemontreatmentbeamoutputandsymmetryusingstatisticalprocesscontrol
AT crowescottb multiinstitutionalevaluationofmachineperformancechecksystemontreatmentbeamoutputandsymmetryusingstatisticalprocesscontrol