Speckle reduction in double-pass retinal images
The double pass (DP) technique quantifies the optical quality of the eye by measuring its point spread function. The low reflectivity of the retina requires the use of a high-brightness, point-like illumination source, and thus, DP systems use laser diodes (LDs). However, LDs light produces speckle,...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Nature Publishing Group UK
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6418212/ https://www.ncbi.nlm.nih.gov/pubmed/30872739 http://dx.doi.org/10.1038/s41598-019-41168-3 |
Sumario: | The double pass (DP) technique quantifies the optical quality of the eye by measuring its point spread function. The low reflectivity of the retina requires the use of a high-brightness, point-like illumination source, and thus, DP systems use laser diodes (LDs). However, LDs light produces speckle, and a low-cost solution to reduce speckle is to include a vibrating mirror in the beam path. With the goal of finding an all-optical solution, here we perform a comparative study of the amount of speckle produced by three semiconductor light sources: an LD, a light emitting diode (LED), and a superluminescent diode (SLED). We also compare the results with the speckle reduction that is obtained with a vibrating mirror. We find that the SLED is a good alternative to LD illumination, as the amount of speckle in the image is almost as low as that obtained with an LD and a vibrating mirror in the beam path. |
---|