Evaluation of percutaneous unilateral trans-femoral implantation of side-hole port-catheter system with coil only fixed-catheter-tip for hepatic arterial infusion chemotherapy
BACKGROUND: The technique for arterial infusion chemotherapy (HAIC) is not standardized which limits its widely application. The aim of this study was to evaluate the long-term functionality and complications of port-catheter system using percutaneous unilateral trans-femoral implantation with coil...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6423799/ https://www.ncbi.nlm.nih.gov/pubmed/30885278 http://dx.doi.org/10.1186/s40644-019-0202-z |
Sumario: | BACKGROUND: The technique for arterial infusion chemotherapy (HAIC) is not standardized which limits its widely application. The aim of this study was to evaluate the long-term functionality and complications of port-catheter system using percutaneous unilateral trans-femoral implantation with coil only fixed-catheter-tip method. METHODS: From January 2013 to January 2017, 205 consecutive patients (138 men; aged 28–88 years; mean, 59.1 ± 11.2 years) with unresectable malignant liver tumors underwent percutaneous implantation of side-hole infusion port-catheter into hepatic artery using coil only fixed-catheter-tip method via the unilateral femoral artery. Technical success, procedure time, duration of port functionality, and complications of port dysfunction were investigated. RESULTS: Implantation technical success was 98.5% and the procedure time was 59.1 ± 10.2 min. Predictable functionality of the port-catheter system at 6-, 12-, and 24 months were 97.5, 89.9, 70.5%, respectively. Complications of port irreversible dysfunction were hepatic artery obstruction (4.0%), catheter occlusion (3.5%), and catheter dislocation (0.5%). Median 5 HAIC cycles (range: 1–14 cycles) were received via port. CONCLUSION: Percutaneous unilateral trans-femoral implantation of a side-hole port-catheter with coils only fixed-catheter-tip method is a simple and feasible interventional technique for HAIC which offers long-term functionality. |
---|