Determination of Optimal Measurement Points for Calibration Equations—Examples by RH Sensors

The calibration points for sensors must be selected carefully. This study uses accuracy and precision as the criteria to evaluate the required numbers of calibration points required. Two types of electric relative humidity (RH) sensors were used to illustrate the method and the standard RH environme...

Descripción completa

Detalles Bibliográficos
Autores principales: Chen, Hsuan-Yu, Chen, Chiachung
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6427136/
https://www.ncbi.nlm.nih.gov/pubmed/30857332
http://dx.doi.org/10.3390/s19051213
_version_ 1783405142263988224
author Chen, Hsuan-Yu
Chen, Chiachung
author_facet Chen, Hsuan-Yu
Chen, Chiachung
author_sort Chen, Hsuan-Yu
collection PubMed
description The calibration points for sensors must be selected carefully. This study uses accuracy and precision as the criteria to evaluate the required numbers of calibration points required. Two types of electric relative humidity (RH) sensors were used to illustrate the method and the standard RH environments were maintained using different saturated salt solutions. The best calibration equation is determined according to the t-value for the highest-order parameter and using the residual plots. Then, the estimated standard errors for the regression equation are used to determine the accuracy of the sensors. The combined uncertainties from the calibration equations for different calibration points for the different saturated salt solutions were then used to evaluate the precision of the sensors. The accuracy of the calibration equations is 0.8% RH for a resistive humidity sensor using 7 calibration points and 0.7% RH for a capacitance humidity sensor using 5 calibration points. The precision is less than 1.0% RH for a resistive sensor and less than 0.9% RH for a capacitive sensor. The method that this study proposed for the selection of calibration points can be applied to other sensors.
format Online
Article
Text
id pubmed-6427136
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-64271362019-04-15 Determination of Optimal Measurement Points for Calibration Equations—Examples by RH Sensors Chen, Hsuan-Yu Chen, Chiachung Sensors (Basel) Article The calibration points for sensors must be selected carefully. This study uses accuracy and precision as the criteria to evaluate the required numbers of calibration points required. Two types of electric relative humidity (RH) sensors were used to illustrate the method and the standard RH environments were maintained using different saturated salt solutions. The best calibration equation is determined according to the t-value for the highest-order parameter and using the residual plots. Then, the estimated standard errors for the regression equation are used to determine the accuracy of the sensors. The combined uncertainties from the calibration equations for different calibration points for the different saturated salt solutions were then used to evaluate the precision of the sensors. The accuracy of the calibration equations is 0.8% RH for a resistive humidity sensor using 7 calibration points and 0.7% RH for a capacitance humidity sensor using 5 calibration points. The precision is less than 1.0% RH for a resistive sensor and less than 0.9% RH for a capacitive sensor. The method that this study proposed for the selection of calibration points can be applied to other sensors. MDPI 2019-03-09 /pmc/articles/PMC6427136/ /pubmed/30857332 http://dx.doi.org/10.3390/s19051213 Text en © 2019 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Chen, Hsuan-Yu
Chen, Chiachung
Determination of Optimal Measurement Points for Calibration Equations—Examples by RH Sensors
title Determination of Optimal Measurement Points for Calibration Equations—Examples by RH Sensors
title_full Determination of Optimal Measurement Points for Calibration Equations—Examples by RH Sensors
title_fullStr Determination of Optimal Measurement Points for Calibration Equations—Examples by RH Sensors
title_full_unstemmed Determination of Optimal Measurement Points for Calibration Equations—Examples by RH Sensors
title_short Determination of Optimal Measurement Points for Calibration Equations—Examples by RH Sensors
title_sort determination of optimal measurement points for calibration equations—examples by rh sensors
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6427136/
https://www.ncbi.nlm.nih.gov/pubmed/30857332
http://dx.doi.org/10.3390/s19051213
work_keys_str_mv AT chenhsuanyu determinationofoptimalmeasurementpointsforcalibrationequationsexamplesbyrhsensors
AT chenchiachung determinationofoptimalmeasurementpointsforcalibrationequationsexamplesbyrhsensors