Fetal Heart Rate Monitoring Implemented by Dynamic Adaptation of Transmission Power of a Flexible Ultrasound Transducer Array

Fetal heart rate (fHR) monitoring using Doppler Ultrasound (US) is a standard method to assess fetal health before and during labor. Typically, an US transducer is positioned on the maternal abdomen and directed towards the fetal heart. Due to fetal movement or displacement of the transducer, the re...

Descripción completa

Detalles Bibliográficos
Autores principales: Hamelmann, Paul, Mischi, Massimo, Kolen, Alexander F., van Laar, Judith O. E. H., Vullings, Rik, Bergmans, Jan W. M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6427711/
https://www.ncbi.nlm.nih.gov/pubmed/30857218
http://dx.doi.org/10.3390/s19051195
_version_ 1783405273234276352
author Hamelmann, Paul
Mischi, Massimo
Kolen, Alexander F.
van Laar, Judith O. E. H.
Vullings, Rik
Bergmans, Jan W. M.
author_facet Hamelmann, Paul
Mischi, Massimo
Kolen, Alexander F.
van Laar, Judith O. E. H.
Vullings, Rik
Bergmans, Jan W. M.
author_sort Hamelmann, Paul
collection PubMed
description Fetal heart rate (fHR) monitoring using Doppler Ultrasound (US) is a standard method to assess fetal health before and during labor. Typically, an US transducer is positioned on the maternal abdomen and directed towards the fetal heart. Due to fetal movement or displacement of the transducer, the relative fetal heart location (fHL) with respect to the US transducer can change, leading to frequent periods of signal loss. Consequently, frequent repositioning of the US transducer is required, which is a cumbersome task affecting clinical workflow. In this research, a new flexible US transducer array is proposed which allows for measuring the fHR independently of the fHL. In addition, a method for dynamic adaptation of the transmission power of this array is introduced with the aim of reducing the total acoustic dose transmitted to the fetus and the associated power consumption, which is an important requirement for application in an ambulatory setting. The method is evaluated using an in-vitro setup of a beating chicken heart. We demonstrate that the signal quality of the Doppler signal acquired with the proposed method is comparable to that of a standard, clinical US transducer. At the same time, our transducer array is able to measure the fHR for varying fHL while only using 50% of the total transmission power of standard, clinical US transducers.
format Online
Article
Text
id pubmed-6427711
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-64277112019-04-15 Fetal Heart Rate Monitoring Implemented by Dynamic Adaptation of Transmission Power of a Flexible Ultrasound Transducer Array Hamelmann, Paul Mischi, Massimo Kolen, Alexander F. van Laar, Judith O. E. H. Vullings, Rik Bergmans, Jan W. M. Sensors (Basel) Article Fetal heart rate (fHR) monitoring using Doppler Ultrasound (US) is a standard method to assess fetal health before and during labor. Typically, an US transducer is positioned on the maternal abdomen and directed towards the fetal heart. Due to fetal movement or displacement of the transducer, the relative fetal heart location (fHL) with respect to the US transducer can change, leading to frequent periods of signal loss. Consequently, frequent repositioning of the US transducer is required, which is a cumbersome task affecting clinical workflow. In this research, a new flexible US transducer array is proposed which allows for measuring the fHR independently of the fHL. In addition, a method for dynamic adaptation of the transmission power of this array is introduced with the aim of reducing the total acoustic dose transmitted to the fetus and the associated power consumption, which is an important requirement for application in an ambulatory setting. The method is evaluated using an in-vitro setup of a beating chicken heart. We demonstrate that the signal quality of the Doppler signal acquired with the proposed method is comparable to that of a standard, clinical US transducer. At the same time, our transducer array is able to measure the fHR for varying fHL while only using 50% of the total transmission power of standard, clinical US transducers. MDPI 2019-03-08 /pmc/articles/PMC6427711/ /pubmed/30857218 http://dx.doi.org/10.3390/s19051195 Text en © 2019 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Hamelmann, Paul
Mischi, Massimo
Kolen, Alexander F.
van Laar, Judith O. E. H.
Vullings, Rik
Bergmans, Jan W. M.
Fetal Heart Rate Monitoring Implemented by Dynamic Adaptation of Transmission Power of a Flexible Ultrasound Transducer Array
title Fetal Heart Rate Monitoring Implemented by Dynamic Adaptation of Transmission Power of a Flexible Ultrasound Transducer Array
title_full Fetal Heart Rate Monitoring Implemented by Dynamic Adaptation of Transmission Power of a Flexible Ultrasound Transducer Array
title_fullStr Fetal Heart Rate Monitoring Implemented by Dynamic Adaptation of Transmission Power of a Flexible Ultrasound Transducer Array
title_full_unstemmed Fetal Heart Rate Monitoring Implemented by Dynamic Adaptation of Transmission Power of a Flexible Ultrasound Transducer Array
title_short Fetal Heart Rate Monitoring Implemented by Dynamic Adaptation of Transmission Power of a Flexible Ultrasound Transducer Array
title_sort fetal heart rate monitoring implemented by dynamic adaptation of transmission power of a flexible ultrasound transducer array
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6427711/
https://www.ncbi.nlm.nih.gov/pubmed/30857218
http://dx.doi.org/10.3390/s19051195
work_keys_str_mv AT hamelmannpaul fetalheartratemonitoringimplementedbydynamicadaptationoftransmissionpowerofaflexibleultrasoundtransducerarray
AT mischimassimo fetalheartratemonitoringimplementedbydynamicadaptationoftransmissionpowerofaflexibleultrasoundtransducerarray
AT kolenalexanderf fetalheartratemonitoringimplementedbydynamicadaptationoftransmissionpowerofaflexibleultrasoundtransducerarray
AT vanlaarjudithoeh fetalheartratemonitoringimplementedbydynamicadaptationoftransmissionpowerofaflexibleultrasoundtransducerarray
AT vullingsrik fetalheartratemonitoringimplementedbydynamicadaptationoftransmissionpowerofaflexibleultrasoundtransducerarray
AT bergmansjanwm fetalheartratemonitoringimplementedbydynamicadaptationoftransmissionpowerofaflexibleultrasoundtransducerarray