Inaccurate Blood Pressure Readings in the Intensive Care Unit: An Observational Study

Measuring and monitoring cerebral perfusion pressure (CPP) is important in the management of patients with certain neurological conditions. To accurately reflect blood pressure at the circle of Willis, the arterial line transducer should be leveled at the tragus. This study measured the relative dis...

Descripción completa

Detalles Bibliográficos
Autores principales: Tse, Brian, Burbridge, Mark A, Jaffe, Richard A, Brock-Utne, John
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6428362/
https://www.ncbi.nlm.nih.gov/pubmed/30906677
http://dx.doi.org/10.7759/cureus.3716
Descripción
Sumario:Measuring and monitoring cerebral perfusion pressure (CPP) is important in the management of patients with certain neurological conditions. To accurately reflect blood pressure at the circle of Willis, the arterial line transducer should be leveled at the tragus. This study measured the relative distance of the transducer to the tragus in 100 intensive care unit (ICU) patients in the mixed ICU at our institution, of which 44 patients had a pressure-sensitive neurological diagnosis. For neurological patients, the average distance was 10.9 cm and for non-neurological patients, the average distance was 11.4 cm (p-value: 0.60). This suggests that the arterial line transducer was leveled at approximately the same level regardless of pathology, potentially leading to falsely elevated CPP readings in patients with pressure-sensitive neurological pathology.