Repeatability, Reproducibility and Agreement of Central Corneal Thickness Measurements by Two Noncontact Pachymetry Devices
This study was designed to assess the repeatability, reproducibility, and agreement of Noncontact Pachymetry (NPC) (Topcan TRK-2P) and the non-contact optical low coherence reflectometer (Lenstar LS 900) devices for measuring Central Corneal Thickness (CCT) of healthy corneas. A total of 82 healthy...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medical Hypothesis, Discovery & Innovation Ophthalmology
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6433204/ https://www.ncbi.nlm.nih.gov/pubmed/30923721 |
_version_ | 1783406259230212096 |
---|---|
author | KOCAMIS, Ozkan KILIC, Rasit |
author_facet | KOCAMIS, Ozkan KILIC, Rasit |
author_sort | KOCAMIS, Ozkan |
collection | PubMed |
description | This study was designed to assess the repeatability, reproducibility, and agreement of Noncontact Pachymetry (NPC) (Topcan TRK-2P) and the non-contact optical low coherence reflectometer (Lenstar LS 900) devices for measuring Central Corneal Thickness (CCT) of healthy corneas. A total of 82 healthy volunteers were evaluated. The first observer used both the TopconTRK-2P and Lenstar 900 devices while the second observer only used the TopconTRK-2P for the measurements. The measurements with either device were repeated three times for each patient, consecutively. The central corneal thickness measurements with the Topcon TRK-2P revealed mean ± Standard Deviation (SD) values of 553.1 ± 36.1 micrometer (µm) for the first observer and 552.3 ± 35.9µm for the second observer and the mean ± SD of CCT was 537.3 ± 34.8µm with the Lenstar 900. The difference between the CCT measurements of the observers using the Topcon TRK-2P (P = 0.142) was insignificant. However, significantly lower measurements were found with the Lenstar 900 compared with the Topcon TRK-2P (P ˂ 0.001). The central corneal thickness measurements obtained by the Topcon TRK-2P were found to have high repeatability for both observers with a lower SD, less than 1% Coefficient of Variation (CV) and higher than 0.99 Intra-Class Correlation Coefficient (ICC) (Observer 1: 3.77 SD, 0.68 CV and 0.995 ICC; the second observer: 3.58 SD, 0.65 CV and 0.995 ICC). There was an excellent inter-observer reproducibility between the two observers for Topcon TRK-2P with 2.71 SD, 0.49 CV, and 0.994 ICC. The Bland-Altman plot showed high agreement between the two devices. These results suggest that the TopconTRK-2P is a reliable device for evaluating CCT in healthy corneas compared with Lenstar 900. |
format | Online Article Text |
id | pubmed-6433204 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Medical Hypothesis, Discovery & Innovation Ophthalmology |
record_format | MEDLINE/PubMed |
spelling | pubmed-64332042019-03-28 Repeatability, Reproducibility and Agreement of Central Corneal Thickness Measurements by Two Noncontact Pachymetry Devices KOCAMIS, Ozkan KILIC, Rasit Med Hypothesis Discov Innov Ophthalmol Original Article This study was designed to assess the repeatability, reproducibility, and agreement of Noncontact Pachymetry (NPC) (Topcan TRK-2P) and the non-contact optical low coherence reflectometer (Lenstar LS 900) devices for measuring Central Corneal Thickness (CCT) of healthy corneas. A total of 82 healthy volunteers were evaluated. The first observer used both the TopconTRK-2P and Lenstar 900 devices while the second observer only used the TopconTRK-2P for the measurements. The measurements with either device were repeated three times for each patient, consecutively. The central corneal thickness measurements with the Topcon TRK-2P revealed mean ± Standard Deviation (SD) values of 553.1 ± 36.1 micrometer (µm) for the first observer and 552.3 ± 35.9µm for the second observer and the mean ± SD of CCT was 537.3 ± 34.8µm with the Lenstar 900. The difference between the CCT measurements of the observers using the Topcon TRK-2P (P = 0.142) was insignificant. However, significantly lower measurements were found with the Lenstar 900 compared with the Topcon TRK-2P (P ˂ 0.001). The central corneal thickness measurements obtained by the Topcon TRK-2P were found to have high repeatability for both observers with a lower SD, less than 1% Coefficient of Variation (CV) and higher than 0.99 Intra-Class Correlation Coefficient (ICC) (Observer 1: 3.77 SD, 0.68 CV and 0.995 ICC; the second observer: 3.58 SD, 0.65 CV and 0.995 ICC). There was an excellent inter-observer reproducibility between the two observers for Topcon TRK-2P with 2.71 SD, 0.49 CV, and 0.994 ICC. The Bland-Altman plot showed high agreement between the two devices. These results suggest that the TopconTRK-2P is a reliable device for evaluating CCT in healthy corneas compared with Lenstar 900. Medical Hypothesis, Discovery & Innovation Ophthalmology 2019 /pmc/articles/PMC6433204/ /pubmed/30923721 Text en ©2019, Med Hypothesis Discov Innov Ophthalmol. This is an Open Access article distributed under the terms of the Creative Commons Attribution License, (http://creativecommons.org/licenses/by/3.0/) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article KOCAMIS, Ozkan KILIC, Rasit Repeatability, Reproducibility and Agreement of Central Corneal Thickness Measurements by Two Noncontact Pachymetry Devices |
title | Repeatability, Reproducibility and Agreement of Central Corneal Thickness Measurements by Two Noncontact Pachymetry Devices |
title_full | Repeatability, Reproducibility and Agreement of Central Corneal Thickness Measurements by Two Noncontact Pachymetry Devices |
title_fullStr | Repeatability, Reproducibility and Agreement of Central Corneal Thickness Measurements by Two Noncontact Pachymetry Devices |
title_full_unstemmed | Repeatability, Reproducibility and Agreement of Central Corneal Thickness Measurements by Two Noncontact Pachymetry Devices |
title_short | Repeatability, Reproducibility and Agreement of Central Corneal Thickness Measurements by Two Noncontact Pachymetry Devices |
title_sort | repeatability, reproducibility and agreement of central corneal thickness measurements by two noncontact pachymetry devices |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6433204/ https://www.ncbi.nlm.nih.gov/pubmed/30923721 |
work_keys_str_mv | AT kocamisozkan repeatabilityreproducibilityandagreementofcentralcornealthicknessmeasurementsbytwononcontactpachymetrydevices AT kilicrasit repeatabilityreproducibilityandagreementofcentralcornealthicknessmeasurementsbytwononcontactpachymetrydevices |