Repeatability, Reproducibility and Agreement of Central Corneal Thickness Measurements by Two Noncontact Pachymetry Devices

This study was designed to assess the repeatability, reproducibility, and agreement of Noncontact Pachymetry (NPC) (Topcan TRK-2P) and the non-contact optical low coherence reflectometer (Lenstar LS 900) devices for measuring Central Corneal Thickness (CCT) of healthy corneas. A total of 82 healthy...

Descripción completa

Detalles Bibliográficos
Autores principales: KOCAMIS, Ozkan, KILIC, Rasit
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medical Hypothesis, Discovery & Innovation Ophthalmology 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6433204/
https://www.ncbi.nlm.nih.gov/pubmed/30923721
_version_ 1783406259230212096
author KOCAMIS, Ozkan
KILIC, Rasit
author_facet KOCAMIS, Ozkan
KILIC, Rasit
author_sort KOCAMIS, Ozkan
collection PubMed
description This study was designed to assess the repeatability, reproducibility, and agreement of Noncontact Pachymetry (NPC) (Topcan TRK-2P) and the non-contact optical low coherence reflectometer (Lenstar LS 900) devices for measuring Central Corneal Thickness (CCT) of healthy corneas. A total of 82 healthy volunteers were evaluated. The first observer used both the TopconTRK-2P and Lenstar 900 devices while the second observer only used the TopconTRK-2P for the measurements. The measurements with either device were repeated three times for each patient, consecutively. The central corneal thickness measurements with the Topcon TRK-2P revealed mean ± Standard Deviation (SD) values of 553.1 ± 36.1 micrometer (µm) for the first observer and 552.3 ± 35.9µm for the second observer and the mean ± SD of CCT was 537.3 ± 34.8µm with the Lenstar 900. The difference between the CCT measurements of the observers using the Topcon TRK-2P (P = 0.142) was insignificant. However, significantly lower measurements were found with the Lenstar 900 compared with the Topcon TRK-2P (P ˂ 0.001). The central corneal thickness measurements obtained by the Topcon TRK-2P were found to have high repeatability for both observers with a lower SD, less than 1% Coefficient of Variation (CV) and higher than 0.99 Intra-Class Correlation Coefficient (ICC) (Observer 1: 3.77 SD, 0.68 CV and 0.995 ICC; the second observer: 3.58 SD, 0.65 CV and 0.995 ICC). There was an excellent inter-observer reproducibility between the two observers for Topcon TRK-2P with 2.71 SD, 0.49 CV, and 0.994 ICC. The Bland-Altman plot showed high agreement between the two devices. These results suggest that the TopconTRK-2P is a reliable device for evaluating CCT in healthy corneas compared with Lenstar 900.
format Online
Article
Text
id pubmed-6433204
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Medical Hypothesis, Discovery & Innovation Ophthalmology
record_format MEDLINE/PubMed
spelling pubmed-64332042019-03-28 Repeatability, Reproducibility and Agreement of Central Corneal Thickness Measurements by Two Noncontact Pachymetry Devices KOCAMIS, Ozkan KILIC, Rasit Med Hypothesis Discov Innov Ophthalmol Original Article This study was designed to assess the repeatability, reproducibility, and agreement of Noncontact Pachymetry (NPC) (Topcan TRK-2P) and the non-contact optical low coherence reflectometer (Lenstar LS 900) devices for measuring Central Corneal Thickness (CCT) of healthy corneas. A total of 82 healthy volunteers were evaluated. The first observer used both the TopconTRK-2P and Lenstar 900 devices while the second observer only used the TopconTRK-2P for the measurements. The measurements with either device were repeated three times for each patient, consecutively. The central corneal thickness measurements with the Topcon TRK-2P revealed mean ± Standard Deviation (SD) values of 553.1 ± 36.1 micrometer (µm) for the first observer and 552.3 ± 35.9µm for the second observer and the mean ± SD of CCT was 537.3 ± 34.8µm with the Lenstar 900. The difference between the CCT measurements of the observers using the Topcon TRK-2P (P = 0.142) was insignificant. However, significantly lower measurements were found with the Lenstar 900 compared with the Topcon TRK-2P (P ˂ 0.001). The central corneal thickness measurements obtained by the Topcon TRK-2P were found to have high repeatability for both observers with a lower SD, less than 1% Coefficient of Variation (CV) and higher than 0.99 Intra-Class Correlation Coefficient (ICC) (Observer 1: 3.77 SD, 0.68 CV and 0.995 ICC; the second observer: 3.58 SD, 0.65 CV and 0.995 ICC). There was an excellent inter-observer reproducibility between the two observers for Topcon TRK-2P with 2.71 SD, 0.49 CV, and 0.994 ICC. The Bland-Altman plot showed high agreement between the two devices. These results suggest that the TopconTRK-2P is a reliable device for evaluating CCT in healthy corneas compared with Lenstar 900. Medical Hypothesis, Discovery & Innovation Ophthalmology 2019 /pmc/articles/PMC6433204/ /pubmed/30923721 Text en ©2019, Med Hypothesis Discov Innov Ophthalmol. This is an Open Access article distributed under the terms of the Creative Commons Attribution License, (http://creativecommons.org/licenses/by/3.0/) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
KOCAMIS, Ozkan
KILIC, Rasit
Repeatability, Reproducibility and Agreement of Central Corneal Thickness Measurements by Two Noncontact Pachymetry Devices
title Repeatability, Reproducibility and Agreement of Central Corneal Thickness Measurements by Two Noncontact Pachymetry Devices
title_full Repeatability, Reproducibility and Agreement of Central Corneal Thickness Measurements by Two Noncontact Pachymetry Devices
title_fullStr Repeatability, Reproducibility and Agreement of Central Corneal Thickness Measurements by Two Noncontact Pachymetry Devices
title_full_unstemmed Repeatability, Reproducibility and Agreement of Central Corneal Thickness Measurements by Two Noncontact Pachymetry Devices
title_short Repeatability, Reproducibility and Agreement of Central Corneal Thickness Measurements by Two Noncontact Pachymetry Devices
title_sort repeatability, reproducibility and agreement of central corneal thickness measurements by two noncontact pachymetry devices
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6433204/
https://www.ncbi.nlm.nih.gov/pubmed/30923721
work_keys_str_mv AT kocamisozkan repeatabilityreproducibilityandagreementofcentralcornealthicknessmeasurementsbytwononcontactpachymetrydevices
AT kilicrasit repeatabilityreproducibilityandagreementofcentralcornealthicknessmeasurementsbytwononcontactpachymetrydevices