Accuracy, Utilization, and Effectiveness Comparisons of Different Continuous Glucose Monitoring Systems

Background: Accuracy and feature sets of continuous glucose monitoring (CGM) systems may influence device utilization and outcomes. We compared clinical trial accuracy and real-world utilization and effectiveness of two different CGM systems. Materials and Methods: Separately conducted accuracy stud...

Descripción completa

Detalles Bibliográficos
Autores principales: Welsh, John B., Gao, Peggy, Derdzinski, Mark, Puhr, Sarah, Johnson, Terri Kang, Walker, Tomas C., Graham, Claudia
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Mary Ann Liebert, Inc., publishers 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6434583/
https://www.ncbi.nlm.nih.gov/pubmed/30681379
http://dx.doi.org/10.1089/dia.2018.0374
Descripción
Sumario:Background: Accuracy and feature sets of continuous glucose monitoring (CGM) systems may influence device utilization and outcomes. We compared clinical trial accuracy and real-world utilization and effectiveness of two different CGM systems. Materials and Methods: Separately conducted accuracy studies of a fifth-generation and a sixth-generation CGM system involved 50 and 159 adults, respectively. For between-system performance comparisons, propensity score methods were utilized to balance cohort characteristics. Real-world outcomes were assessed in 10,000 anonymized patients who had switched from the fifth-generation to the sixth-generation system and had used connected mobile devices to upload data from both systems, allowing pairwise comparisons of device utilization and glucose concentration distributions. Results: Propensity score-adjusted mean absolute relative differences for the fifth- and sixth-generation systems were 9.0% and 9.9%, and the percentages of values within ±20%/20 mg/dL were 93.1% and 92.5%, respectively. The sixth-generation system, but not the fifth-generation system, met accuracy criteria for interoperable CGM systems. Both systems had high real-world utilization rates (93.8% and 95.3% in the fifth- and sixth-generation systems, respectively). Use of the sixth-generation system was associated with fewer glucose values <55 mg/dL (<3.1 mmol/L) (0.7% vs. 1.1%, P < 0.001) and more values 70–180 mg/dL (3.9–10.0 mmol/L) (57.3% vs. 56.0%, P < 0.001) than the fifth-generation system. Conclusions: CGM performance outcomes can be compared through the propensity score analysis of clinical trial data and pairwise comparisons of real-world data. The systems compared here had nearly equivalent accuracy and utilization rates. Longer term biochemical and psychosocial benefits observed with the fifth-generation system are also expected with the sixth-generation system.