Optimization of Semiautomated Calibration Algorithm of Multichannel Electrotactile Feedback for Myoelectric Hand Prosthesis

The main drawback of the commercially available myoelectric hand prostheses is the absence of somatosensory feedback. We recently developed a feedback interface for multiple degrees of freedom myoelectric prosthesis that allows proprioceptive and sensory information (i.e., grasping force) to be tran...

Descripción completa

Detalles Bibliográficos
Autores principales: Isaković, Milica, Malešević, Jovana, Keller, Thierry, Kostić, Miloš, Štrbac, Matija
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6437744/
https://www.ncbi.nlm.nih.gov/pubmed/31001360
http://dx.doi.org/10.1155/2019/9298758
_version_ 1783406985364897792
author Isaković, Milica
Malešević, Jovana
Keller, Thierry
Kostić, Miloš
Štrbac, Matija
author_facet Isaković, Milica
Malešević, Jovana
Keller, Thierry
Kostić, Miloš
Štrbac, Matija
author_sort Isaković, Milica
collection PubMed
description The main drawback of the commercially available myoelectric hand prostheses is the absence of somatosensory feedback. We recently developed a feedback interface for multiple degrees of freedom myoelectric prosthesis that allows proprioceptive and sensory information (i.e., grasping force) to be transmitted to the wearer instantaneously. High information bandwidth is achieved through intelligent control of spatiotemporal distribution of electrical pulses over a custom-designed electrode array. As electrotactile sensations are location-dependent and the developed interface requires that electrical stimuli are perceived to be of the same intensity on all locations, a calibration procedure is of high importance. The aim of this study was to gain more insight into the calibration procedure and optimize this process by leveraging a priori knowledge. For this purpose, we conducted a study with 9 able-bodied subjects performing 10 sessions of the array electrode calibration. Based on the collected data, we optimized and simplified the calibration procedure by adapting the initial (baseline) amplitude values in the calibration algorithm. The results suggest there is an individual pattern of stimulation amplitudes across 16 electrode pads for each subject, which is not affected by the initial amplitudes. Moreover, the number of user actions performed and the time needed for the calibration procedure are significantly reduced by the proposed methodology.
format Online
Article
Text
id pubmed-6437744
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Hindawi
record_format MEDLINE/PubMed
spelling pubmed-64377442019-04-18 Optimization of Semiautomated Calibration Algorithm of Multichannel Electrotactile Feedback for Myoelectric Hand Prosthesis Isaković, Milica Malešević, Jovana Keller, Thierry Kostić, Miloš Štrbac, Matija Appl Bionics Biomech Research Article The main drawback of the commercially available myoelectric hand prostheses is the absence of somatosensory feedback. We recently developed a feedback interface for multiple degrees of freedom myoelectric prosthesis that allows proprioceptive and sensory information (i.e., grasping force) to be transmitted to the wearer instantaneously. High information bandwidth is achieved through intelligent control of spatiotemporal distribution of electrical pulses over a custom-designed electrode array. As electrotactile sensations are location-dependent and the developed interface requires that electrical stimuli are perceived to be of the same intensity on all locations, a calibration procedure is of high importance. The aim of this study was to gain more insight into the calibration procedure and optimize this process by leveraging a priori knowledge. For this purpose, we conducted a study with 9 able-bodied subjects performing 10 sessions of the array electrode calibration. Based on the collected data, we optimized and simplified the calibration procedure by adapting the initial (baseline) amplitude values in the calibration algorithm. The results suggest there is an individual pattern of stimulation amplitudes across 16 electrode pads for each subject, which is not affected by the initial amplitudes. Moreover, the number of user actions performed and the time needed for the calibration procedure are significantly reduced by the proposed methodology. Hindawi 2019-03-14 /pmc/articles/PMC6437744/ /pubmed/31001360 http://dx.doi.org/10.1155/2019/9298758 Text en Copyright © 2019 Milica Isaković et al. http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Isaković, Milica
Malešević, Jovana
Keller, Thierry
Kostić, Miloš
Štrbac, Matija
Optimization of Semiautomated Calibration Algorithm of Multichannel Electrotactile Feedback for Myoelectric Hand Prosthesis
title Optimization of Semiautomated Calibration Algorithm of Multichannel Electrotactile Feedback for Myoelectric Hand Prosthesis
title_full Optimization of Semiautomated Calibration Algorithm of Multichannel Electrotactile Feedback for Myoelectric Hand Prosthesis
title_fullStr Optimization of Semiautomated Calibration Algorithm of Multichannel Electrotactile Feedback for Myoelectric Hand Prosthesis
title_full_unstemmed Optimization of Semiautomated Calibration Algorithm of Multichannel Electrotactile Feedback for Myoelectric Hand Prosthesis
title_short Optimization of Semiautomated Calibration Algorithm of Multichannel Electrotactile Feedback for Myoelectric Hand Prosthesis
title_sort optimization of semiautomated calibration algorithm of multichannel electrotactile feedback for myoelectric hand prosthesis
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6437744/
https://www.ncbi.nlm.nih.gov/pubmed/31001360
http://dx.doi.org/10.1155/2019/9298758
work_keys_str_mv AT isakovicmilica optimizationofsemiautomatedcalibrationalgorithmofmultichannelelectrotactilefeedbackformyoelectrichandprosthesis
AT malesevicjovana optimizationofsemiautomatedcalibrationalgorithmofmultichannelelectrotactilefeedbackformyoelectrichandprosthesis
AT kellerthierry optimizationofsemiautomatedcalibrationalgorithmofmultichannelelectrotactilefeedbackformyoelectrichandprosthesis
AT kosticmilos optimizationofsemiautomatedcalibrationalgorithmofmultichannelelectrotactilefeedbackformyoelectrichandprosthesis
AT strbacmatija optimizationofsemiautomatedcalibrationalgorithmofmultichannelelectrotactilefeedbackformyoelectrichandprosthesis