Comparison of Performance of Digital Hemoglobinometer over Automated Hematology Analyzer for Hemoglobin Estimation and Its user-friendliness among the Pregnant Women in Selected District Hospitals of Madhya Pradesh

CONTEXT: There is a need for a simple screening method for the detection of anemia that can be used by public health workers in the field. AIMS: The aim of this study was to compare two methods for hemoglobin estimation, i.e., automated hematology analyzer and Digital Hemoglobinometer, and to find o...

Descripción completa

Detalles Bibliográficos
Autores principales: Toppo, Manju, Pal, Dinesh Kumar, Gour, Devendra, Melwani, Veena, Dubey, Manju, Mishra, Archana
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6437795/
https://www.ncbi.nlm.nih.gov/pubmed/30983710
http://dx.doi.org/10.4103/ijcm.IJCM_216_18
_version_ 1783406990969536512
author Toppo, Manju
Pal, Dinesh Kumar
Gour, Devendra
Melwani, Veena
Dubey, Manju
Mishra, Archana
author_facet Toppo, Manju
Pal, Dinesh Kumar
Gour, Devendra
Melwani, Veena
Dubey, Manju
Mishra, Archana
author_sort Toppo, Manju
collection PubMed
description CONTEXT: There is a need for a simple screening method for the detection of anemia that can be used by public health workers in the field. AIMS: The aim of this study was to compare two methods for hemoglobin estimation, i.e., automated hematology analyzer and Digital Hemoglobinometer, and to find out the sensitivity and specificity of Digital Hemoglobinometer for the estimation of hemoglobin. SUBJECTS AND METHODS: A hospital-based cross-sectional study was carried out for 6 months from April to September 2017 in a District Hospital of five High Priority Districts of Madhya Pradesh. Two hundred and sixty antenatal females per district were selected for the study. RESULTS: The mean hemoglobin by autoanalyzer is 10.19, and that by Digital Hemoglobinometer device is 9.89. Overall, sensitivity of Digital Hemoglobinometer for hemoglobin estimation was calculated to be 89.4% and specificity was calculated to be 63.6%. Positive predictive value was found to be 82.6% and negative predictive value was 75.8% compared against AutoAnalyser (gold standard). CONCLUSIONS: As the Digital Hemoglobinometer device has high sensitivity and specificity and good diagnostic accuracy, it must be used at the community level in resource-poor setting for hemoglobin estimation. In primary health-care conditions, Digital Hemoglobinometer can significantly reduce misdiagnosis of anemia compared with clinical assessment alone.
format Online
Article
Text
id pubmed-6437795
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-64377952019-04-12 Comparison of Performance of Digital Hemoglobinometer over Automated Hematology Analyzer for Hemoglobin Estimation and Its user-friendliness among the Pregnant Women in Selected District Hospitals of Madhya Pradesh Toppo, Manju Pal, Dinesh Kumar Gour, Devendra Melwani, Veena Dubey, Manju Mishra, Archana Indian J Community Med Original Article CONTEXT: There is a need for a simple screening method for the detection of anemia that can be used by public health workers in the field. AIMS: The aim of this study was to compare two methods for hemoglobin estimation, i.e., automated hematology analyzer and Digital Hemoglobinometer, and to find out the sensitivity and specificity of Digital Hemoglobinometer for the estimation of hemoglobin. SUBJECTS AND METHODS: A hospital-based cross-sectional study was carried out for 6 months from April to September 2017 in a District Hospital of five High Priority Districts of Madhya Pradesh. Two hundred and sixty antenatal females per district were selected for the study. RESULTS: The mean hemoglobin by autoanalyzer is 10.19, and that by Digital Hemoglobinometer device is 9.89. Overall, sensitivity of Digital Hemoglobinometer for hemoglobin estimation was calculated to be 89.4% and specificity was calculated to be 63.6%. Positive predictive value was found to be 82.6% and negative predictive value was 75.8% compared against AutoAnalyser (gold standard). CONCLUSIONS: As the Digital Hemoglobinometer device has high sensitivity and specificity and good diagnostic accuracy, it must be used at the community level in resource-poor setting for hemoglobin estimation. In primary health-care conditions, Digital Hemoglobinometer can significantly reduce misdiagnosis of anemia compared with clinical assessment alone. Wolters Kluwer - Medknow 2019 /pmc/articles/PMC6437795/ /pubmed/30983710 http://dx.doi.org/10.4103/ijcm.IJCM_216_18 Text en Copyright: © 2019 Indian Journal of Community Medicine http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Toppo, Manju
Pal, Dinesh Kumar
Gour, Devendra
Melwani, Veena
Dubey, Manju
Mishra, Archana
Comparison of Performance of Digital Hemoglobinometer over Automated Hematology Analyzer for Hemoglobin Estimation and Its user-friendliness among the Pregnant Women in Selected District Hospitals of Madhya Pradesh
title Comparison of Performance of Digital Hemoglobinometer over Automated Hematology Analyzer for Hemoglobin Estimation and Its user-friendliness among the Pregnant Women in Selected District Hospitals of Madhya Pradesh
title_full Comparison of Performance of Digital Hemoglobinometer over Automated Hematology Analyzer for Hemoglobin Estimation and Its user-friendliness among the Pregnant Women in Selected District Hospitals of Madhya Pradesh
title_fullStr Comparison of Performance of Digital Hemoglobinometer over Automated Hematology Analyzer for Hemoglobin Estimation and Its user-friendliness among the Pregnant Women in Selected District Hospitals of Madhya Pradesh
title_full_unstemmed Comparison of Performance of Digital Hemoglobinometer over Automated Hematology Analyzer for Hemoglobin Estimation and Its user-friendliness among the Pregnant Women in Selected District Hospitals of Madhya Pradesh
title_short Comparison of Performance of Digital Hemoglobinometer over Automated Hematology Analyzer for Hemoglobin Estimation and Its user-friendliness among the Pregnant Women in Selected District Hospitals of Madhya Pradesh
title_sort comparison of performance of digital hemoglobinometer over automated hematology analyzer for hemoglobin estimation and its user-friendliness among the pregnant women in selected district hospitals of madhya pradesh
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6437795/
https://www.ncbi.nlm.nih.gov/pubmed/30983710
http://dx.doi.org/10.4103/ijcm.IJCM_216_18
work_keys_str_mv AT toppomanju comparisonofperformanceofdigitalhemoglobinometeroverautomatedhematologyanalyzerforhemoglobinestimationanditsuserfriendlinessamongthepregnantwomeninselecteddistricthospitalsofmadhyapradesh
AT paldineshkumar comparisonofperformanceofdigitalhemoglobinometeroverautomatedhematologyanalyzerforhemoglobinestimationanditsuserfriendlinessamongthepregnantwomeninselecteddistricthospitalsofmadhyapradesh
AT gourdevendra comparisonofperformanceofdigitalhemoglobinometeroverautomatedhematologyanalyzerforhemoglobinestimationanditsuserfriendlinessamongthepregnantwomeninselecteddistricthospitalsofmadhyapradesh
AT melwaniveena comparisonofperformanceofdigitalhemoglobinometeroverautomatedhematologyanalyzerforhemoglobinestimationanditsuserfriendlinessamongthepregnantwomeninselecteddistricthospitalsofmadhyapradesh
AT dubeymanju comparisonofperformanceofdigitalhemoglobinometeroverautomatedhematologyanalyzerforhemoglobinestimationanditsuserfriendlinessamongthepregnantwomeninselecteddistricthospitalsofmadhyapradesh
AT mishraarchana comparisonofperformanceofdigitalhemoglobinometeroverautomatedhematologyanalyzerforhemoglobinestimationanditsuserfriendlinessamongthepregnantwomeninselecteddistricthospitalsofmadhyapradesh