Comparison of Performance of Digital Hemoglobinometer over Automated Hematology Analyzer for Hemoglobin Estimation and Its user-friendliness among the Pregnant Women in Selected District Hospitals of Madhya Pradesh
CONTEXT: There is a need for a simple screening method for the detection of anemia that can be used by public health workers in the field. AIMS: The aim of this study was to compare two methods for hemoglobin estimation, i.e., automated hematology analyzer and Digital Hemoglobinometer, and to find o...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6437795/ https://www.ncbi.nlm.nih.gov/pubmed/30983710 http://dx.doi.org/10.4103/ijcm.IJCM_216_18 |
_version_ | 1783406990969536512 |
---|---|
author | Toppo, Manju Pal, Dinesh Kumar Gour, Devendra Melwani, Veena Dubey, Manju Mishra, Archana |
author_facet | Toppo, Manju Pal, Dinesh Kumar Gour, Devendra Melwani, Veena Dubey, Manju Mishra, Archana |
author_sort | Toppo, Manju |
collection | PubMed |
description | CONTEXT: There is a need for a simple screening method for the detection of anemia that can be used by public health workers in the field. AIMS: The aim of this study was to compare two methods for hemoglobin estimation, i.e., automated hematology analyzer and Digital Hemoglobinometer, and to find out the sensitivity and specificity of Digital Hemoglobinometer for the estimation of hemoglobin. SUBJECTS AND METHODS: A hospital-based cross-sectional study was carried out for 6 months from April to September 2017 in a District Hospital of five High Priority Districts of Madhya Pradesh. Two hundred and sixty antenatal females per district were selected for the study. RESULTS: The mean hemoglobin by autoanalyzer is 10.19, and that by Digital Hemoglobinometer device is 9.89. Overall, sensitivity of Digital Hemoglobinometer for hemoglobin estimation was calculated to be 89.4% and specificity was calculated to be 63.6%. Positive predictive value was found to be 82.6% and negative predictive value was 75.8% compared against AutoAnalyser (gold standard). CONCLUSIONS: As the Digital Hemoglobinometer device has high sensitivity and specificity and good diagnostic accuracy, it must be used at the community level in resource-poor setting for hemoglobin estimation. In primary health-care conditions, Digital Hemoglobinometer can significantly reduce misdiagnosis of anemia compared with clinical assessment alone. |
format | Online Article Text |
id | pubmed-6437795 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Wolters Kluwer - Medknow |
record_format | MEDLINE/PubMed |
spelling | pubmed-64377952019-04-12 Comparison of Performance of Digital Hemoglobinometer over Automated Hematology Analyzer for Hemoglobin Estimation and Its user-friendliness among the Pregnant Women in Selected District Hospitals of Madhya Pradesh Toppo, Manju Pal, Dinesh Kumar Gour, Devendra Melwani, Veena Dubey, Manju Mishra, Archana Indian J Community Med Original Article CONTEXT: There is a need for a simple screening method for the detection of anemia that can be used by public health workers in the field. AIMS: The aim of this study was to compare two methods for hemoglobin estimation, i.e., automated hematology analyzer and Digital Hemoglobinometer, and to find out the sensitivity and specificity of Digital Hemoglobinometer for the estimation of hemoglobin. SUBJECTS AND METHODS: A hospital-based cross-sectional study was carried out for 6 months from April to September 2017 in a District Hospital of five High Priority Districts of Madhya Pradesh. Two hundred and sixty antenatal females per district were selected for the study. RESULTS: The mean hemoglobin by autoanalyzer is 10.19, and that by Digital Hemoglobinometer device is 9.89. Overall, sensitivity of Digital Hemoglobinometer for hemoglobin estimation was calculated to be 89.4% and specificity was calculated to be 63.6%. Positive predictive value was found to be 82.6% and negative predictive value was 75.8% compared against AutoAnalyser (gold standard). CONCLUSIONS: As the Digital Hemoglobinometer device has high sensitivity and specificity and good diagnostic accuracy, it must be used at the community level in resource-poor setting for hemoglobin estimation. In primary health-care conditions, Digital Hemoglobinometer can significantly reduce misdiagnosis of anemia compared with clinical assessment alone. Wolters Kluwer - Medknow 2019 /pmc/articles/PMC6437795/ /pubmed/30983710 http://dx.doi.org/10.4103/ijcm.IJCM_216_18 Text en Copyright: © 2019 Indian Journal of Community Medicine http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. |
spellingShingle | Original Article Toppo, Manju Pal, Dinesh Kumar Gour, Devendra Melwani, Veena Dubey, Manju Mishra, Archana Comparison of Performance of Digital Hemoglobinometer over Automated Hematology Analyzer for Hemoglobin Estimation and Its user-friendliness among the Pregnant Women in Selected District Hospitals of Madhya Pradesh |
title | Comparison of Performance of Digital Hemoglobinometer over Automated Hematology Analyzer for Hemoglobin Estimation and Its user-friendliness among the Pregnant Women in Selected District Hospitals of Madhya Pradesh |
title_full | Comparison of Performance of Digital Hemoglobinometer over Automated Hematology Analyzer for Hemoglobin Estimation and Its user-friendliness among the Pregnant Women in Selected District Hospitals of Madhya Pradesh |
title_fullStr | Comparison of Performance of Digital Hemoglobinometer over Automated Hematology Analyzer for Hemoglobin Estimation and Its user-friendliness among the Pregnant Women in Selected District Hospitals of Madhya Pradesh |
title_full_unstemmed | Comparison of Performance of Digital Hemoglobinometer over Automated Hematology Analyzer for Hemoglobin Estimation and Its user-friendliness among the Pregnant Women in Selected District Hospitals of Madhya Pradesh |
title_short | Comparison of Performance of Digital Hemoglobinometer over Automated Hematology Analyzer for Hemoglobin Estimation and Its user-friendliness among the Pregnant Women in Selected District Hospitals of Madhya Pradesh |
title_sort | comparison of performance of digital hemoglobinometer over automated hematology analyzer for hemoglobin estimation and its user-friendliness among the pregnant women in selected district hospitals of madhya pradesh |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6437795/ https://www.ncbi.nlm.nih.gov/pubmed/30983710 http://dx.doi.org/10.4103/ijcm.IJCM_216_18 |
work_keys_str_mv | AT toppomanju comparisonofperformanceofdigitalhemoglobinometeroverautomatedhematologyanalyzerforhemoglobinestimationanditsuserfriendlinessamongthepregnantwomeninselecteddistricthospitalsofmadhyapradesh AT paldineshkumar comparisonofperformanceofdigitalhemoglobinometeroverautomatedhematologyanalyzerforhemoglobinestimationanditsuserfriendlinessamongthepregnantwomeninselecteddistricthospitalsofmadhyapradesh AT gourdevendra comparisonofperformanceofdigitalhemoglobinometeroverautomatedhematologyanalyzerforhemoglobinestimationanditsuserfriendlinessamongthepregnantwomeninselecteddistricthospitalsofmadhyapradesh AT melwaniveena comparisonofperformanceofdigitalhemoglobinometeroverautomatedhematologyanalyzerforhemoglobinestimationanditsuserfriendlinessamongthepregnantwomeninselecteddistricthospitalsofmadhyapradesh AT dubeymanju comparisonofperformanceofdigitalhemoglobinometeroverautomatedhematologyanalyzerforhemoglobinestimationanditsuserfriendlinessamongthepregnantwomeninselecteddistricthospitalsofmadhyapradesh AT mishraarchana comparisonofperformanceofdigitalhemoglobinometeroverautomatedhematologyanalyzerforhemoglobinestimationanditsuserfriendlinessamongthepregnantwomeninselecteddistricthospitalsofmadhyapradesh |