Pretreatment Dose Verification in Volumetric Modulated Arc Therapy Using Liquid Ionization Chamber

PURPOSE: The purpose of the present study was to evaluate the practicability of liquid ionization chamber (LIC) for pretreatment dose verification of the advanced radiotherapy techniques such as volumetric modulated arc therapy (VMAT). MATERIALS AND METHODS: The dosimetric characteristics of LIC suc...

Descripción completa

Detalles Bibliográficos
Autores principales: Kakade, Nitin R., Kumar, Rajesh, Sharma, Sunil Dutt, Mittal, Vikram, Datta, D.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6438046/
https://www.ncbi.nlm.nih.gov/pubmed/30983765
http://dx.doi.org/10.4103/jmp.JMP_108_18
Descripción
Sumario:PURPOSE: The purpose of the present study was to evaluate the practicability of liquid ionization chamber (LIC) for pretreatment dose verification of the advanced radiotherapy techniques such as volumetric modulated arc therapy (VMAT). MATERIALS AND METHODS: The dosimetric characteristics of LIC such as repeatability, sensitivity, monitor unit linearity, dose rate dependence, angular dependence, voltage-current response, and output factors were investigated in 6 MV therapeutic X-ray beams. The LIC was cross-calibrated against 0.125-cc air-filled thimble ionization chamber. A dedicated dosimetry insert made up of Perspex to incorporate the LIC at proper location in the intensity-modulated radiation therapy thorax phantom was locally fabricated. The collection efficiency and ion recombination correction factor was determined using the two-dose rate method. Pretreatment dose verification measurement of VMAT treatment plans were carried out using the liquid ionization chamber as well as small volume (0.125 cc) air-filled thimble ionization chamber. The measured dose values by the two dosimeters and TPS calculated dose at a given point were compared. RESULTS: The relative percentage differences between the TPS calculated and measured doses were within ± 1.57% for LIC and ± 2.21% for 0.125 cc ionization chamber, respectively. CONCLUSIONS: The measured dose values by the two dosimeters and TPS calculated dose at a given point were found comparable suggesting that the LIC could be a good choice of dosimeter for pretreatment dose verification in VMAT.