BACKGROUND: If the threshold at implant of leadless transcatheter pacemakers (LTPs) is less than 2.0 V, pacing thresholds reportedly decrease significantly by 1 month and maintain an optimal value of less than 1.0 V by 6 months. CASE SUMMARY: We report a case series of two patients with unstable pac...

Descripción completa

Detalles Bibliográficos
Autores principales: Yoh, Masue, Takagi, Masahiko, Takahashi, Hiroki, Yoshio, Takuro, Shiojima, Ichiro
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6439362/
https://www.ncbi.nlm.nih.gov/pubmed/31020236
http://dx.doi.org/10.1093/ehjcr/yty160
Descripción
Sumario:BACKGROUND: If the threshold at implant of leadless transcatheter pacemakers (LTPs) is less than 2.0 V, pacing thresholds reportedly decrease significantly by 1 month and maintain an optimal value of less than 1.0 V by 6 months. CASE SUMMARY: We report a case series of two patients with unstable pacing thresholds of the LTPs in the subacute phase after implant. The first patient (77-year-old man) was implanted an LTP for sick sinus syndrome. At that time of implant, the pacing threshold was 0.9 V at 0.24 ms. At 1 week and 1 month later, the threshold had increased to more than 2.0 V at 0.24 ms. We investigated the trend data for the week and found variations in the threshold. The second patient (81-year-old man) was implanted an LTP for bradycardia and atrial fibrillation. The pacing threshold at implantation was 0.63 V at 0.24 ms. One week later, the threshold had increased in supine position and decreased in sitting position. The trend data for the week were fluctuating greatly. DISCUSSION: The pacing threshold may increase to more than 2.0 V with significant fluctuation on assessment at 1 week and 1 month after implantation in association with changes in body position, even though we confirmed a stable threshold at implant. If an increased threshold is observed, it is necessary to check the trend data and threshold in each body position.