Cargando…

Mechanical analysis of a novel biodegradable zinc alloy stent based on a degradation model

BACKGROUND: Biodegradable stents display insufficient scaffold performance due to their poor Young’s Modulus. In addition, the corresponding biodegradable materials harbor weakened structures during degradation processes. Consequently, such stents have not been extensively applied in clinical therap...

Descripción completa

Detalles Bibliográficos
Autores principales: Peng, Kun, Cui, Xinyang, Qiao, Aike, Mu, Yongliang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6444843/
https://www.ncbi.nlm.nih.gov/pubmed/30940146
http://dx.doi.org/10.1186/s12938-019-0661-2
Descripción
Sumario:BACKGROUND: Biodegradable stents display insufficient scaffold performance due to their poor Young’s Modulus. In addition, the corresponding biodegradable materials harbor weakened structures during degradation processes. Consequently, such stents have not been extensively applied in clinical therapy. In this study, the scaffold performance of a patented stent and its ability to reshape damaged vessels during degradation process were evaluated. METHODS: A common stent was chosen as a control to assess the mechanical behavior of the patented stent. Finite element analysis was used to simulate stent deployment into a 40% stenotic vessel. A material corrosion model involving uniform and stress corrosion was implemented within the finite element framework to update the stress state following degradation. RESULTS: The results showed that radial recoiling ratio and mass loss ratio of the patented stent is 7.19% and 3.1%, respectively, which are definitely lower than those of the common stent with the corresponding values of 22.6% and 14.1%, respectively. Moreover, the patented stent displayed stronger scaffold performance in a corrosive environment and the plaque treated with patented stents had a larger and flatter lumen. CONCLUSION: Owing to its improved mechanical performance, the novel biodegradable zinc alloy stent reported here has high potential as an alternative choice in surgery.