Evaluation of a commercial liquid-chromatography high-resolution mass-spectrometry method for the determination of hepcidin-25

INTRODUCTION: Reliable determination of hepcidin-25, a key regulator of iron metabolism, is important. This study aimed at evaluating the performance of the Hepcidin-25 Liquid Chromatography-Tandem Mass-Spectrometry (LC-MS/MS) Kit (Immundiagnostik AG, Bensheim, Germany) for quantification of the hep...

Descripción completa

Detalles Bibliográficos
Autores principales: Enko, Dietmar, Zelzer, Sieglinde, Fauler, Günter, Herrmann, Markus
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Croatian Society of Medical Biochemistry and Laboratory Medicine 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6457918/
https://www.ncbi.nlm.nih.gov/pubmed/31015783
http://dx.doi.org/10.11613/BM.2019.020701
_version_ 1783409938966511616
author Enko, Dietmar
Zelzer, Sieglinde
Fauler, Günter
Herrmann, Markus
author_facet Enko, Dietmar
Zelzer, Sieglinde
Fauler, Günter
Herrmann, Markus
author_sort Enko, Dietmar
collection PubMed
description INTRODUCTION: Reliable determination of hepcidin-25, a key regulator of iron metabolism, is important. This study aimed at evaluating the performance of the Hepcidin-25 Liquid Chromatography-Tandem Mass-Spectrometry (LC-MS/MS) Kit (Immundiagnostik AG, Bensheim, Germany) for quantification of the hepcidin-25 protein. MATERIALS AND METHODS: Precision, accuracy, linearity, and preanalytical requirements of the liquid-chromatography high-resolution mass-spectrometry (LC-HR-MS) method were evaluated. The imprecision and bias acceptance criteria were defined ≤ 15%. We investigated sample stability at room temperature (RT) and after repeated freeze and thaw cycles. Additionally, we assessed serum hepcidin-25 concentrations of 165 healthy adults referred for a medical check-up. RESULTS: The hepcidin-25 LC-MS/MS assay was linear over the concentration range of 3 – 200 ng/mL. Within- and between-run precision ranged between 1.9 – 8.6% and 5.1 – 12.4%, respectively. The mean bias of the low and high control material was - 2.7% and 2.1%, respectively. At RT, serum samples were stable for 3 h (mean bias + 0.3%). After two and three freeze and thaw cycles, hepcidin-25 concentrations showed a bias of + 8.0 and + 20%, respectively. Of 165 healthy adults, 109 females had a significantly lower median of 8.42 (range: 1.00 – 60.10) ng/mL compared to 56 males with 15.76 (range: 1.50 – 60.50) ng/mL (P = 0.002). CONCLUSIONS: The hepcidin-25 LC-MS/MS kit shows a broad analytical range and meets the imprecision and bias acceptance criteria of ≤ 15%. Serum samples can be stored at RT for 3 h and resist up to two freeze and thaw cycles.
format Online
Article
Text
id pubmed-6457918
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Croatian Society of Medical Biochemistry and Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-64579182019-04-23 Evaluation of a commercial liquid-chromatography high-resolution mass-spectrometry method for the determination of hepcidin-25 Enko, Dietmar Zelzer, Sieglinde Fauler, Günter Herrmann, Markus Biochem Med (Zagreb) Original Articles INTRODUCTION: Reliable determination of hepcidin-25, a key regulator of iron metabolism, is important. This study aimed at evaluating the performance of the Hepcidin-25 Liquid Chromatography-Tandem Mass-Spectrometry (LC-MS/MS) Kit (Immundiagnostik AG, Bensheim, Germany) for quantification of the hepcidin-25 protein. MATERIALS AND METHODS: Precision, accuracy, linearity, and preanalytical requirements of the liquid-chromatography high-resolution mass-spectrometry (LC-HR-MS) method were evaluated. The imprecision and bias acceptance criteria were defined ≤ 15%. We investigated sample stability at room temperature (RT) and after repeated freeze and thaw cycles. Additionally, we assessed serum hepcidin-25 concentrations of 165 healthy adults referred for a medical check-up. RESULTS: The hepcidin-25 LC-MS/MS assay was linear over the concentration range of 3 – 200 ng/mL. Within- and between-run precision ranged between 1.9 – 8.6% and 5.1 – 12.4%, respectively. The mean bias of the low and high control material was - 2.7% and 2.1%, respectively. At RT, serum samples were stable for 3 h (mean bias + 0.3%). After two and three freeze and thaw cycles, hepcidin-25 concentrations showed a bias of + 8.0 and + 20%, respectively. Of 165 healthy adults, 109 females had a significantly lower median of 8.42 (range: 1.00 – 60.10) ng/mL compared to 56 males with 15.76 (range: 1.50 – 60.50) ng/mL (P = 0.002). CONCLUSIONS: The hepcidin-25 LC-MS/MS kit shows a broad analytical range and meets the imprecision and bias acceptance criteria of ≤ 15%. Serum samples can be stored at RT for 3 h and resist up to two freeze and thaw cycles. Croatian Society of Medical Biochemistry and Laboratory Medicine 2019-04-15 2019-06-15 /pmc/articles/PMC6457918/ /pubmed/31015783 http://dx.doi.org/10.11613/BM.2019.020701 Text en ©Croatian Society of Medical Biochemistry and Laboratory Medicine. This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Articles
Enko, Dietmar
Zelzer, Sieglinde
Fauler, Günter
Herrmann, Markus
Evaluation of a commercial liquid-chromatography high-resolution mass-spectrometry method for the determination of hepcidin-25
title Evaluation of a commercial liquid-chromatography high-resolution mass-spectrometry method for the determination of hepcidin-25
title_full Evaluation of a commercial liquid-chromatography high-resolution mass-spectrometry method for the determination of hepcidin-25
title_fullStr Evaluation of a commercial liquid-chromatography high-resolution mass-spectrometry method for the determination of hepcidin-25
title_full_unstemmed Evaluation of a commercial liquid-chromatography high-resolution mass-spectrometry method for the determination of hepcidin-25
title_short Evaluation of a commercial liquid-chromatography high-resolution mass-spectrometry method for the determination of hepcidin-25
title_sort evaluation of a commercial liquid-chromatography high-resolution mass-spectrometry method for the determination of hepcidin-25
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6457918/
https://www.ncbi.nlm.nih.gov/pubmed/31015783
http://dx.doi.org/10.11613/BM.2019.020701
work_keys_str_mv AT enkodietmar evaluationofacommercialliquidchromatographyhighresolutionmassspectrometrymethodforthedeterminationofhepcidin25
AT zelzersieglinde evaluationofacommercialliquidchromatographyhighresolutionmassspectrometrymethodforthedeterminationofhepcidin25
AT faulergunter evaluationofacommercialliquidchromatographyhighresolutionmassspectrometrymethodforthedeterminationofhepcidin25
AT herrmannmarkus evaluationofacommercialliquidchromatographyhighresolutionmassspectrometrymethodforthedeterminationofhepcidin25