Evaluation of a commercial liquid-chromatography high-resolution mass-spectrometry method for the determination of hepcidin-25
INTRODUCTION: Reliable determination of hepcidin-25, a key regulator of iron metabolism, is important. This study aimed at evaluating the performance of the Hepcidin-25 Liquid Chromatography-Tandem Mass-Spectrometry (LC-MS/MS) Kit (Immundiagnostik AG, Bensheim, Germany) for quantification of the hep...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Croatian Society of Medical Biochemistry and Laboratory Medicine
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6457918/ https://www.ncbi.nlm.nih.gov/pubmed/31015783 http://dx.doi.org/10.11613/BM.2019.020701 |
_version_ | 1783409938966511616 |
---|---|
author | Enko, Dietmar Zelzer, Sieglinde Fauler, Günter Herrmann, Markus |
author_facet | Enko, Dietmar Zelzer, Sieglinde Fauler, Günter Herrmann, Markus |
author_sort | Enko, Dietmar |
collection | PubMed |
description | INTRODUCTION: Reliable determination of hepcidin-25, a key regulator of iron metabolism, is important. This study aimed at evaluating the performance of the Hepcidin-25 Liquid Chromatography-Tandem Mass-Spectrometry (LC-MS/MS) Kit (Immundiagnostik AG, Bensheim, Germany) for quantification of the hepcidin-25 protein. MATERIALS AND METHODS: Precision, accuracy, linearity, and preanalytical requirements of the liquid-chromatography high-resolution mass-spectrometry (LC-HR-MS) method were evaluated. The imprecision and bias acceptance criteria were defined ≤ 15%. We investigated sample stability at room temperature (RT) and after repeated freeze and thaw cycles. Additionally, we assessed serum hepcidin-25 concentrations of 165 healthy adults referred for a medical check-up. RESULTS: The hepcidin-25 LC-MS/MS assay was linear over the concentration range of 3 – 200 ng/mL. Within- and between-run precision ranged between 1.9 – 8.6% and 5.1 – 12.4%, respectively. The mean bias of the low and high control material was - 2.7% and 2.1%, respectively. At RT, serum samples were stable for 3 h (mean bias + 0.3%). After two and three freeze and thaw cycles, hepcidin-25 concentrations showed a bias of + 8.0 and + 20%, respectively. Of 165 healthy adults, 109 females had a significantly lower median of 8.42 (range: 1.00 – 60.10) ng/mL compared to 56 males with 15.76 (range: 1.50 – 60.50) ng/mL (P = 0.002). CONCLUSIONS: The hepcidin-25 LC-MS/MS kit shows a broad analytical range and meets the imprecision and bias acceptance criteria of ≤ 15%. Serum samples can be stored at RT for 3 h and resist up to two freeze and thaw cycles. |
format | Online Article Text |
id | pubmed-6457918 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Croatian Society of Medical Biochemistry and Laboratory Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-64579182019-04-23 Evaluation of a commercial liquid-chromatography high-resolution mass-spectrometry method for the determination of hepcidin-25 Enko, Dietmar Zelzer, Sieglinde Fauler, Günter Herrmann, Markus Biochem Med (Zagreb) Original Articles INTRODUCTION: Reliable determination of hepcidin-25, a key regulator of iron metabolism, is important. This study aimed at evaluating the performance of the Hepcidin-25 Liquid Chromatography-Tandem Mass-Spectrometry (LC-MS/MS) Kit (Immundiagnostik AG, Bensheim, Germany) for quantification of the hepcidin-25 protein. MATERIALS AND METHODS: Precision, accuracy, linearity, and preanalytical requirements of the liquid-chromatography high-resolution mass-spectrometry (LC-HR-MS) method were evaluated. The imprecision and bias acceptance criteria were defined ≤ 15%. We investigated sample stability at room temperature (RT) and after repeated freeze and thaw cycles. Additionally, we assessed serum hepcidin-25 concentrations of 165 healthy adults referred for a medical check-up. RESULTS: The hepcidin-25 LC-MS/MS assay was linear over the concentration range of 3 – 200 ng/mL. Within- and between-run precision ranged between 1.9 – 8.6% and 5.1 – 12.4%, respectively. The mean bias of the low and high control material was - 2.7% and 2.1%, respectively. At RT, serum samples were stable for 3 h (mean bias + 0.3%). After two and three freeze and thaw cycles, hepcidin-25 concentrations showed a bias of + 8.0 and + 20%, respectively. Of 165 healthy adults, 109 females had a significantly lower median of 8.42 (range: 1.00 – 60.10) ng/mL compared to 56 males with 15.76 (range: 1.50 – 60.50) ng/mL (P = 0.002). CONCLUSIONS: The hepcidin-25 LC-MS/MS kit shows a broad analytical range and meets the imprecision and bias acceptance criteria of ≤ 15%. Serum samples can be stored at RT for 3 h and resist up to two freeze and thaw cycles. Croatian Society of Medical Biochemistry and Laboratory Medicine 2019-04-15 2019-06-15 /pmc/articles/PMC6457918/ /pubmed/31015783 http://dx.doi.org/10.11613/BM.2019.020701 Text en ©Croatian Society of Medical Biochemistry and Laboratory Medicine. This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Articles Enko, Dietmar Zelzer, Sieglinde Fauler, Günter Herrmann, Markus Evaluation of a commercial liquid-chromatography high-resolution mass-spectrometry method for the determination of hepcidin-25 |
title | Evaluation of a commercial liquid-chromatography high-resolution mass-spectrometry method for the determination of hepcidin-25 |
title_full | Evaluation of a commercial liquid-chromatography high-resolution mass-spectrometry method for the determination of hepcidin-25 |
title_fullStr | Evaluation of a commercial liquid-chromatography high-resolution mass-spectrometry method for the determination of hepcidin-25 |
title_full_unstemmed | Evaluation of a commercial liquid-chromatography high-resolution mass-spectrometry method for the determination of hepcidin-25 |
title_short | Evaluation of a commercial liquid-chromatography high-resolution mass-spectrometry method for the determination of hepcidin-25 |
title_sort | evaluation of a commercial liquid-chromatography high-resolution mass-spectrometry method for the determination of hepcidin-25 |
topic | Original Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6457918/ https://www.ncbi.nlm.nih.gov/pubmed/31015783 http://dx.doi.org/10.11613/BM.2019.020701 |
work_keys_str_mv | AT enkodietmar evaluationofacommercialliquidchromatographyhighresolutionmassspectrometrymethodforthedeterminationofhepcidin25 AT zelzersieglinde evaluationofacommercialliquidchromatographyhighresolutionmassspectrometrymethodforthedeterminationofhepcidin25 AT faulergunter evaluationofacommercialliquidchromatographyhighresolutionmassspectrometrymethodforthedeterminationofhepcidin25 AT herrmannmarkus evaluationofacommercialliquidchromatographyhighresolutionmassspectrometrymethodforthedeterminationofhepcidin25 |