The effect of increasing power when grooving using phacoemulsification

PURPOSE: To determine optimal power settings on the Centurion Vision System during the grooving step in cataract surgery. METHODS: Intact porcine lenses hardened by formalin and placed in a chamber designed to simulate the anterior chamber of the eye were used to test longitudinal power at 40%, 70%,...

Descripción completa

Detalles Bibliográficos
Autores principales: Thomson, Rhett S, Bird, Brian A, Stutz, Lance A, Heczko, Joshua B, Bernhisel, Ashlie A, Barlow, William R, Zaugg, Brian, Olson, Randall J, Pettey, Jeff H
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove Medical Press 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6469737/
https://www.ncbi.nlm.nih.gov/pubmed/31043766
http://dx.doi.org/10.2147/OPTH.S194731
Descripción
Sumario:PURPOSE: To determine optimal power settings on the Centurion Vision System during the grooving step in cataract surgery. METHODS: Intact porcine lenses hardened by formalin and placed in a chamber designed to simulate the anterior chamber of the eye were used to test longitudinal power at 40%, 70%, and 100% and torsional power at 0%. Flow rate was set at 40 mL/min. Vacuum was set at 400 mmHg, intraocular pressure was set at 50 mmHg, and a balanced phacoemulsification tip with a 20 degree tip and a 30 degree bevel was used. Efficiency (time to groove the lens in half) was determined. RESULTS: Increasing longitudinal power from 40% to 70% increased efficiency by 28% (P<0.05), and by 32% (P<0.05) when increasing longitudinal power from 40% to 100%. There was no statistically significant increase in efficiency from 70% to 100%. CONCLUSION: For the tested variables, a longitudinal power of 70% was determined to be most efficient during the grooving step of cataract surgery for equivalent 3–4+ nuclei. Further increases in power demonstrated no statistically significant improvement in efficiency.