The effect of increasing power when grooving using phacoemulsification
PURPOSE: To determine optimal power settings on the Centurion Vision System during the grooving step in cataract surgery. METHODS: Intact porcine lenses hardened by formalin and placed in a chamber designed to simulate the anterior chamber of the eye were used to test longitudinal power at 40%, 70%,...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Dove Medical Press
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6469737/ https://www.ncbi.nlm.nih.gov/pubmed/31043766 http://dx.doi.org/10.2147/OPTH.S194731 |
Sumario: | PURPOSE: To determine optimal power settings on the Centurion Vision System during the grooving step in cataract surgery. METHODS: Intact porcine lenses hardened by formalin and placed in a chamber designed to simulate the anterior chamber of the eye were used to test longitudinal power at 40%, 70%, and 100% and torsional power at 0%. Flow rate was set at 40 mL/min. Vacuum was set at 400 mmHg, intraocular pressure was set at 50 mmHg, and a balanced phacoemulsification tip with a 20 degree tip and a 30 degree bevel was used. Efficiency (time to groove the lens in half) was determined. RESULTS: Increasing longitudinal power from 40% to 70% increased efficiency by 28% (P<0.05), and by 32% (P<0.05) when increasing longitudinal power from 40% to 100%. There was no statistically significant increase in efficiency from 70% to 100%. CONCLUSION: For the tested variables, a longitudinal power of 70% was determined to be most efficient during the grooving step of cataract surgery for equivalent 3–4+ nuclei. Further increases in power demonstrated no statistically significant improvement in efficiency. |
---|