Cargando…
Modular junction fractures in a modern rotating-platform knee arthroplasty system
In this series, we report the findings from four patients who presented with pain and mechanical symptoms after revision total knee arthroplasty with the DePuy Sigma TC3 RP prosthesis. Plain radiographs for each patient demonstrated failure of the femoral component at the modular junction of the fem...
Autores principales: | Ihekweazu, Ugonna, Courtney, P. Maxwell, Baral, Elexis C., Austin, Matthew S., McLawhorn, Alexander S. |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6470385/ https://www.ncbi.nlm.nih.gov/pubmed/31020020 http://dx.doi.org/10.1016/j.artd.2018.11.003 |
Ejemplares similares
-
Fracture of the femoral adapter bolt and taper adapter in a modern rotating platform knee arthroplasty
por: Baral, Elexis C., et al.
Publicado: (2017) -
Management of Resistant, Atypical and Culture-negative Periprosthetic Joint Infections after Hip and Knee Arthroplasty
por: McLawhorn, Alexander S., et al.
Publicado: (2016) -
Fracture of an S-ROM stem at the sleeve-stem junction
por: Rueckl, Kilian, et al.
Publicado: (2018) -
Nontraumatic Tibial Polyethylene Insert Cone Fracture in Rotating-Platform Total Knee Arthroplasty
por: Keogh, Cillian J., et al.
Publicado: (2021) -
Tribocorrosion is common but mild in modular humeral components in shoulder arthroplasty: an implant retrieval analysis
por: Brusalis, Christopher M., et al.
Publicado: (2022)