Measurement of Circular Dichroism Spectra without Control of a Phase Modulator using Retardation Domain Analysis

This study investigated the measurement of circular dichroism (CD) spectra without controlling a phase modulator. In a conventional CD system, the peak retardation of the phase modulator must remain constant over the observed wavelength range. Thus, the phase modulator must be controlled to maintain...

Descripción completa

Detalles Bibliográficos
Autor principal: Satozono, Hiroshi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6480008/
https://www.ncbi.nlm.nih.gov/pubmed/30974911
http://dx.doi.org/10.3390/molecules24071418
Descripción
Sumario:This study investigated the measurement of circular dichroism (CD) spectra without controlling a phase modulator. In a conventional CD system, the peak retardation of the phase modulator must remain constant over the observed wavelength range. Thus, the phase modulator must be controlled to maintain an appropriate modulation degree at an observed wavelength. In contrast, CD obtained using retardation domain analysis is not affected by peak retardation. Consequently, CD spectra can be measured without control of the phase modulator, which was experimentally demonstrated in this study. Additionally, linear dichroism spectra were obtained using retardation domain analysis.