Cargando…
Evaluation des performances du cytomètre « MUSE AUTO CD4/CD4% » vs « GUAVA AUTO CD4/CD4% » pour la mesure du taux de lymphocytes CD4 chez des patients infectés par le VIH au Cameroun
INTRODUCTION: This study aimed to evaluate the performances of the MUSE(®) flow cytometer compared with the reference GUAVA(®) flow cytometer. METHODS: We conducted an experimental study on HIV-infected patient samples. Venous blood samples, collected in a K3 EDTA tube, were analyzed within 24-48 ho...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The African Field Epidemiology Network
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6492305/ https://www.ncbi.nlm.nih.gov/pubmed/31068996 http://dx.doi.org/10.11604/pamj.2019.32.2.16990 |
Sumario: | INTRODUCTION: This study aimed to evaluate the performances of the MUSE(®) flow cytometer compared with the reference GUAVA(®) flow cytometer. METHODS: We conducted an experimental study on HIV-infected patient samples. Venous blood samples, collected in a K3 EDTA tube, were analyzed within 24-48 hours by MUSE(®) and GUAVA(®) cytometers at the International Center for medical diagnosis (Centre International de Diagnostic médical) in Yaoundé. RESULTS: In total, 227 samples were analyzed. There was a strong intraclass correlation (p<0.0001) between MUSE(®) and GUAVA(®) cytometers with a correlation coefficient 0.998 (95% CI: 0,998-0,999) for the absolute values and 0,992 (95% CI: 0,989-0,994) for the percentages. A strong positive linear correlation (p=0.0001) was found between MUSE(®) and GUAVA(®) cytometers with linear regression slope r2 = 0.98 (95% CI=0,97-0,99) for the absolute values and r2= 0.98 (95% CI= 0.96-1,00) for the percentages. The biases were -4,80 cells/µl (-101.31-91.71) for the absolute values and -0.89% (IC: -6,08-4.3) for the percentages. The percentage of data points outside the limits of agreement was 12/227 (5.29%) and 10/227 (4.41%) respectively for the absolute values and percentages. Cohen's kappa coefficient was 0.92 and the area under the curve was 0,9975 (CI 95%: 0.99-1). CONCLUSION: MUSE(®)AUTO CD4/CD4% cytometer is a powerful instrument because its results are consistent with those obtained by the reference cytometer. It can enable tracking of patients infected with HIV, in particular in the developing countries |
---|