Assessment of the diagnostic performance of TrueHb® point-of-care hemometer compared with Sysmex i3 analyzer among patients at International Hospital Kampala, Uganda
Aim/Objective: To assess the diagnostic performance of TrueHb® point-of-care (POC) hemometer compared with Sysmex i3 analyzer at International Hospital Kampala, Uganda. Materials and methods: We analyzed ethylenediaminetetraacetic acid blood samples to estimate hemoglobin (Hb) levels using parallel...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Dove
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6497909/ https://www.ncbi.nlm.nih.gov/pubmed/31118851 http://dx.doi.org/10.2147/JBM.S188302 |
_version_ | 1783415559384203264 |
---|---|
author | Taremwa, Ivan Mugisha Ndeze, Ivan Mwambi, Bashir Atuhairwe, Christine Achieng, Diana Inda Natukunda, Bernard |
author_facet | Taremwa, Ivan Mugisha Ndeze, Ivan Mwambi, Bashir Atuhairwe, Christine Achieng, Diana Inda Natukunda, Bernard |
author_sort | Taremwa, Ivan Mugisha |
collection | PubMed |
description | Aim/Objective: To assess the diagnostic performance of TrueHb® point-of-care (POC) hemometer compared with Sysmex i3 analyzer at International Hospital Kampala, Uganda. Materials and methods: We analyzed ethylenediaminetetraacetic acid blood samples to estimate hemoglobin (Hb) levels using parallel testing with TrueHb® hemometer and Sysmex i3 analyzer. Data were analyzed to ascertain the diagnostic performance of the test assays using the Bland and Altman method. Sensitivity, specificity, positive and negative predictive values were calculated. Results: The study enrolled 402 patients; of these, 156 (38.8%) were males. The average Hb levels were 8.7±1.8 and 13.3±2.6 g/dL for the anemic and nonanemic patients, respectively. One hundred and fifty-five participants were anemic, giving anemia prevalence of 38.56% (95% CI: 35.17–40.38). The mean difference of the TrueHb® and Sysmex i3 assays was 2.2219 (SD 1.07915), and the two devices did not show a difference in their measurements (t=−2.407, p-value 0.017, 95% CI: −0.095–0.010). Further, they showed a significant level of agreement (t=41.281; 95% CI: 2.1161–2.3277) and intraclass correlation coefficients (ICC=0.793). The sensitivity, specificity, positive and negative predictive values were 100.00%, 51.01%, 55.16% and 100.00%, respectively. The average performance turnaround time (TAT) for the TrueHb® hemometer was 2.46 mins (95% CI: 2.37–2.55). Conclusion: TrueHb® POC hemometer is an accurate POC for Hb estimation with a good performance agreement with the Sysmex i3 analyzer. This, coupled with its utility aspects, makes it a good diagnostic tool in a high anemia burden and low-resource setting. |
format | Online Article Text |
id | pubmed-6497909 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Dove |
record_format | MEDLINE/PubMed |
spelling | pubmed-64979092019-05-22 Assessment of the diagnostic performance of TrueHb® point-of-care hemometer compared with Sysmex i3 analyzer among patients at International Hospital Kampala, Uganda Taremwa, Ivan Mugisha Ndeze, Ivan Mwambi, Bashir Atuhairwe, Christine Achieng, Diana Inda Natukunda, Bernard J Blood Med Original Research Aim/Objective: To assess the diagnostic performance of TrueHb® point-of-care (POC) hemometer compared with Sysmex i3 analyzer at International Hospital Kampala, Uganda. Materials and methods: We analyzed ethylenediaminetetraacetic acid blood samples to estimate hemoglobin (Hb) levels using parallel testing with TrueHb® hemometer and Sysmex i3 analyzer. Data were analyzed to ascertain the diagnostic performance of the test assays using the Bland and Altman method. Sensitivity, specificity, positive and negative predictive values were calculated. Results: The study enrolled 402 patients; of these, 156 (38.8%) were males. The average Hb levels were 8.7±1.8 and 13.3±2.6 g/dL for the anemic and nonanemic patients, respectively. One hundred and fifty-five participants were anemic, giving anemia prevalence of 38.56% (95% CI: 35.17–40.38). The mean difference of the TrueHb® and Sysmex i3 assays was 2.2219 (SD 1.07915), and the two devices did not show a difference in their measurements (t=−2.407, p-value 0.017, 95% CI: −0.095–0.010). Further, they showed a significant level of agreement (t=41.281; 95% CI: 2.1161–2.3277) and intraclass correlation coefficients (ICC=0.793). The sensitivity, specificity, positive and negative predictive values were 100.00%, 51.01%, 55.16% and 100.00%, respectively. The average performance turnaround time (TAT) for the TrueHb® hemometer was 2.46 mins (95% CI: 2.37–2.55). Conclusion: TrueHb® POC hemometer is an accurate POC for Hb estimation with a good performance agreement with the Sysmex i3 analyzer. This, coupled with its utility aspects, makes it a good diagnostic tool in a high anemia burden and low-resource setting. Dove 2019-04-01 /pmc/articles/PMC6497909/ /pubmed/31118851 http://dx.doi.org/10.2147/JBM.S188302 Text en © 2019 Taremwa et al. http://creativecommons.org/licenses/by-nc/3.0/ This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms (https://www.dovepress.com/terms.php). |
spellingShingle | Original Research Taremwa, Ivan Mugisha Ndeze, Ivan Mwambi, Bashir Atuhairwe, Christine Achieng, Diana Inda Natukunda, Bernard Assessment of the diagnostic performance of TrueHb® point-of-care hemometer compared with Sysmex i3 analyzer among patients at International Hospital Kampala, Uganda |
title | Assessment of the diagnostic performance of TrueHb® point-of-care hemometer compared with Sysmex i3 analyzer among patients at International Hospital Kampala, Uganda |
title_full | Assessment of the diagnostic performance of TrueHb® point-of-care hemometer compared with Sysmex i3 analyzer among patients at International Hospital Kampala, Uganda |
title_fullStr | Assessment of the diagnostic performance of TrueHb® point-of-care hemometer compared with Sysmex i3 analyzer among patients at International Hospital Kampala, Uganda |
title_full_unstemmed | Assessment of the diagnostic performance of TrueHb® point-of-care hemometer compared with Sysmex i3 analyzer among patients at International Hospital Kampala, Uganda |
title_short | Assessment of the diagnostic performance of TrueHb® point-of-care hemometer compared with Sysmex i3 analyzer among patients at International Hospital Kampala, Uganda |
title_sort | assessment of the diagnostic performance of truehb® point-of-care hemometer compared with sysmex i3 analyzer among patients at international hospital kampala, uganda |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6497909/ https://www.ncbi.nlm.nih.gov/pubmed/31118851 http://dx.doi.org/10.2147/JBM.S188302 |
work_keys_str_mv | AT taremwaivanmugisha assessmentofthediagnosticperformanceoftruehbpointofcarehemometercomparedwithsysmexi3analyzeramongpatientsatinternationalhospitalkampalauganda AT ndezeivan assessmentofthediagnosticperformanceoftruehbpointofcarehemometercomparedwithsysmexi3analyzeramongpatientsatinternationalhospitalkampalauganda AT mwambibashir assessmentofthediagnosticperformanceoftruehbpointofcarehemometercomparedwithsysmexi3analyzeramongpatientsatinternationalhospitalkampalauganda AT atuhairwechristine assessmentofthediagnosticperformanceoftruehbpointofcarehemometercomparedwithsysmexi3analyzeramongpatientsatinternationalhospitalkampalauganda AT achiengdianainda assessmentofthediagnosticperformanceoftruehbpointofcarehemometercomparedwithsysmexi3analyzeramongpatientsatinternationalhospitalkampalauganda AT natukundabernard assessmentofthediagnosticperformanceoftruehbpointofcarehemometercomparedwithsysmexi3analyzeramongpatientsatinternationalhospitalkampalauganda |