Assessment of the diagnostic performance of TrueHb® point-of-care hemometer compared with Sysmex i3 analyzer among patients at International Hospital Kampala, Uganda

Aim/Objective: To assess the diagnostic performance of TrueHb® point-of-care (POC) hemometer compared with Sysmex i3 analyzer at International Hospital Kampala, Uganda. Materials and methods: We analyzed ethylenediaminetetraacetic acid blood samples to estimate hemoglobin (Hb) levels using parallel...

Descripción completa

Detalles Bibliográficos
Autores principales: Taremwa, Ivan Mugisha, Ndeze, Ivan, Mwambi, Bashir, Atuhairwe, Christine, Achieng, Diana Inda, Natukunda, Bernard
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6497909/
https://www.ncbi.nlm.nih.gov/pubmed/31118851
http://dx.doi.org/10.2147/JBM.S188302
_version_ 1783415559384203264
author Taremwa, Ivan Mugisha
Ndeze, Ivan
Mwambi, Bashir
Atuhairwe, Christine
Achieng, Diana Inda
Natukunda, Bernard
author_facet Taremwa, Ivan Mugisha
Ndeze, Ivan
Mwambi, Bashir
Atuhairwe, Christine
Achieng, Diana Inda
Natukunda, Bernard
author_sort Taremwa, Ivan Mugisha
collection PubMed
description Aim/Objective: To assess the diagnostic performance of TrueHb® point-of-care (POC) hemometer compared with Sysmex i3 analyzer at International Hospital Kampala, Uganda. Materials and methods: We analyzed ethylenediaminetetraacetic acid blood samples to estimate hemoglobin (Hb) levels using parallel testing with TrueHb® hemometer and Sysmex i3 analyzer. Data were analyzed to ascertain the diagnostic performance of the test assays using the Bland and Altman method. Sensitivity, specificity, positive and negative predictive values were calculated. Results: The study enrolled 402 patients; of these, 156 (38.8%) were males. The average Hb levels were 8.7±1.8 and 13.3±2.6 g/dL for the anemic and nonanemic patients, respectively. One hundred and fifty-five participants were anemic, giving anemia prevalence of 38.56% (95% CI: 35.17–40.38). The mean difference of the TrueHb® and Sysmex i3 assays was 2.2219 (SD 1.07915), and the two devices did not show a difference in their measurements (t=−2.407, p-value 0.017, 95% CI: −0.095–0.010). Further, they showed a significant level of agreement (t=41.281; 95% CI: 2.1161–2.3277) and intraclass correlation coefficients (ICC=0.793). The sensitivity, specificity, positive and negative predictive values were 100.00%, 51.01%, 55.16% and 100.00%, respectively. The average performance turnaround time (TAT) for the TrueHb® hemometer was 2.46 mins (95% CI: 2.37–2.55). Conclusion: TrueHb® POC hemometer is an accurate POC for Hb estimation with a good performance agreement with the Sysmex i3 analyzer. This, coupled with its utility aspects, makes it a good diagnostic tool in a high anemia burden and low-resource setting.
format Online
Article
Text
id pubmed-6497909
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Dove
record_format MEDLINE/PubMed
spelling pubmed-64979092019-05-22 Assessment of the diagnostic performance of TrueHb® point-of-care hemometer compared with Sysmex i3 analyzer among patients at International Hospital Kampala, Uganda Taremwa, Ivan Mugisha Ndeze, Ivan Mwambi, Bashir Atuhairwe, Christine Achieng, Diana Inda Natukunda, Bernard J Blood Med Original Research Aim/Objective: To assess the diagnostic performance of TrueHb® point-of-care (POC) hemometer compared with Sysmex i3 analyzer at International Hospital Kampala, Uganda. Materials and methods: We analyzed ethylenediaminetetraacetic acid blood samples to estimate hemoglobin (Hb) levels using parallel testing with TrueHb® hemometer and Sysmex i3 analyzer. Data were analyzed to ascertain the diagnostic performance of the test assays using the Bland and Altman method. Sensitivity, specificity, positive and negative predictive values were calculated. Results: The study enrolled 402 patients; of these, 156 (38.8%) were males. The average Hb levels were 8.7±1.8 and 13.3±2.6 g/dL for the anemic and nonanemic patients, respectively. One hundred and fifty-five participants were anemic, giving anemia prevalence of 38.56% (95% CI: 35.17–40.38). The mean difference of the TrueHb® and Sysmex i3 assays was 2.2219 (SD 1.07915), and the two devices did not show a difference in their measurements (t=−2.407, p-value 0.017, 95% CI: −0.095–0.010). Further, they showed a significant level of agreement (t=41.281; 95% CI: 2.1161–2.3277) and intraclass correlation coefficients (ICC=0.793). The sensitivity, specificity, positive and negative predictive values were 100.00%, 51.01%, 55.16% and 100.00%, respectively. The average performance turnaround time (TAT) for the TrueHb® hemometer was 2.46 mins (95% CI: 2.37–2.55). Conclusion: TrueHb® POC hemometer is an accurate POC for Hb estimation with a good performance agreement with the Sysmex i3 analyzer. This, coupled with its utility aspects, makes it a good diagnostic tool in a high anemia burden and low-resource setting. Dove 2019-04-01 /pmc/articles/PMC6497909/ /pubmed/31118851 http://dx.doi.org/10.2147/JBM.S188302 Text en © 2019 Taremwa et al. http://creativecommons.org/licenses/by-nc/3.0/ This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms (https://www.dovepress.com/terms.php).
spellingShingle Original Research
Taremwa, Ivan Mugisha
Ndeze, Ivan
Mwambi, Bashir
Atuhairwe, Christine
Achieng, Diana Inda
Natukunda, Bernard
Assessment of the diagnostic performance of TrueHb® point-of-care hemometer compared with Sysmex i3 analyzer among patients at International Hospital Kampala, Uganda
title Assessment of the diagnostic performance of TrueHb® point-of-care hemometer compared with Sysmex i3 analyzer among patients at International Hospital Kampala, Uganda
title_full Assessment of the diagnostic performance of TrueHb® point-of-care hemometer compared with Sysmex i3 analyzer among patients at International Hospital Kampala, Uganda
title_fullStr Assessment of the diagnostic performance of TrueHb® point-of-care hemometer compared with Sysmex i3 analyzer among patients at International Hospital Kampala, Uganda
title_full_unstemmed Assessment of the diagnostic performance of TrueHb® point-of-care hemometer compared with Sysmex i3 analyzer among patients at International Hospital Kampala, Uganda
title_short Assessment of the diagnostic performance of TrueHb® point-of-care hemometer compared with Sysmex i3 analyzer among patients at International Hospital Kampala, Uganda
title_sort assessment of the diagnostic performance of truehb® point-of-care hemometer compared with sysmex i3 analyzer among patients at international hospital kampala, uganda
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6497909/
https://www.ncbi.nlm.nih.gov/pubmed/31118851
http://dx.doi.org/10.2147/JBM.S188302
work_keys_str_mv AT taremwaivanmugisha assessmentofthediagnosticperformanceoftruehbpointofcarehemometercomparedwithsysmexi3analyzeramongpatientsatinternationalhospitalkampalauganda
AT ndezeivan assessmentofthediagnosticperformanceoftruehbpointofcarehemometercomparedwithsysmexi3analyzeramongpatientsatinternationalhospitalkampalauganda
AT mwambibashir assessmentofthediagnosticperformanceoftruehbpointofcarehemometercomparedwithsysmexi3analyzeramongpatientsatinternationalhospitalkampalauganda
AT atuhairwechristine assessmentofthediagnosticperformanceoftruehbpointofcarehemometercomparedwithsysmexi3analyzeramongpatientsatinternationalhospitalkampalauganda
AT achiengdianainda assessmentofthediagnosticperformanceoftruehbpointofcarehemometercomparedwithsysmexi3analyzeramongpatientsatinternationalhospitalkampalauganda
AT natukundabernard assessmentofthediagnosticperformanceoftruehbpointofcarehemometercomparedwithsysmexi3analyzeramongpatientsatinternationalhospitalkampalauganda