Performance of the cobas u 701 Analyzer in Urinary Tract Infection Screening
BACKGROUND: Negative urine cultures to rule out urinary tract infections (UTI) generate a considerable laboratory workload; thus, a rapid screening test is desirable. We evaluated the performance of a new automated microscopy analyzer, cobas u 701 (Roche Diagnostics International, Rotkreuz, Switzerl...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Society for Laboratory Medicine
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6502947/ https://www.ncbi.nlm.nih.gov/pubmed/31037865 http://dx.doi.org/10.3343/alm.2019.39.5.464 |
Sumario: | BACKGROUND: Negative urine cultures to rule out urinary tract infections (UTI) generate a considerable laboratory workload; thus, a rapid screening test is desirable. We evaluated the performance of a new automated microscopy analyzer, cobas u 701 (Roche Diagnostics International, Rotkreuz, Switzerland) for the screening of UTI, and developed a rule-out strategy to reduce the number of samples requiring culture. We also assessed squamous epithelial cell (SEC) count as a predictor of culture contamination. METHODS: In total, 1,604 urine samples from outpatients were analyzed with cobas u 701 and culture. Bacterial (BAC) and white blood cell (WBC) counts were used for sample interpretation. To determine a useful cut-off point to predict negative cultures, we selected the highest sensitivity and specificity values obtained from ROC curves. Diagnostic accuracy by age and gender was evaluated. RESULTS: Urine culture showed growth of ≥10(4) colony forming units (CFU)/mL in 256 samples (16.0%). The highest sensitivity (91.8%) and specificity (68.4%) were obtained for cut-off points of 119 BAC/µL and 22 WBC/µL. The combination of BAC and WBC improved the performance of the rule-out strategy with a low rate of false-negative results (1.5%) and a high negative predictive value (NPV, 97.3%). Fifty-seven percent of the samples would not have required culture. SEC count was a poor predictor of culture contamination. CONCLUSIONS: cobas u 701 can substantially reduce the number of urine samples requiring culture, with a low false-negative rate and a high NPV. |
---|