Diagnostic performance of the specific uptake size index for semi-quantitative analysis of I-123-FP-CIT SPECT: harmonized multi-center research setting versus typical clinical single-camera setting

INTRODUCTION: The specific uptake size index (SUSI) of striatal FP-CIT uptake is independent of spatial resolution in the SPECT image, in contrast to the specific binding ratio (SBR). This suggests that the SUSI is particularly appropriate for multi-site/multi-camera settings in which camera-specifi...

Descripción completa

Detalles Bibliográficos
Autores principales: Buchert, Ralph, Lange, Catharina, Spehl, Timo S., Apostolova, Ivayla, Frings, Lars, Jonsson, Cathrine, Meyer, Philipp T., Hellwig, Sabine
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6505020/
https://www.ncbi.nlm.nih.gov/pubmed/31065816
http://dx.doi.org/10.1186/s13550-019-0506-9
_version_ 1783416671580454912
author Buchert, Ralph
Lange, Catharina
Spehl, Timo S.
Apostolova, Ivayla
Frings, Lars
Jonsson, Cathrine
Meyer, Philipp T.
Hellwig, Sabine
author_facet Buchert, Ralph
Lange, Catharina
Spehl, Timo S.
Apostolova, Ivayla
Frings, Lars
Jonsson, Cathrine
Meyer, Philipp T.
Hellwig, Sabine
author_sort Buchert, Ralph
collection PubMed
description INTRODUCTION: The specific uptake size index (SUSI) of striatal FP-CIT uptake is independent of spatial resolution in the SPECT image, in contrast to the specific binding ratio (SBR). This suggests that the SUSI is particularly appropriate for multi-site/multi-camera settings in which camera-specific effects increase inter-subject variability of spatial resolution. However, the SUSI is sensitive to inter-subject variability of striatum size. Furthermore, it might be more sensitive to errors of the estimate of non-displaceable FP-CIT binding. This study compared SUSI and SBR in the multi-site/multi-camera (MULTI) setting of a prospective multi-center study and in a mono-site/mono-camera (MONO) setting representative of clinical routine. METHODS: The MULTI setting included patients with Parkinson’s disease (PD, n = 438) and healthy controls (n = 207) from the Parkinson Progression Marker Initiative. The MONO setting included 122 patients from routine clinical patient care in whom FP-CIT SPECT had been performed with the same double-head SPECT system according to the same acquisition and reconstruction protocol. Patients were categorized as “neurodegenerative” (n = 84) or “non-neurodegenerative” (n = 38) based on follow-up data. FP-CIT SPECTs were stereotactically normalized to MNI space. SUSI and SBR were computed for caudate, putamen, and whole striatum using unilateral ROIs predefined in MNI space. SUSI analysis was repeated in native patient space in the MONO setting. The area (AUC) under the ROC curve for identification of PD/“neurodegenerative” cases was used as performance measure. RESULTS: In both settings, the highest AUC was achieved by the putamen (minimum over both hemispheres), independent of the semi-quantitative method (SUSI or SBR). The putaminal SUSI provided slightly better performance with ROI analysis in MNI space compared to patient space (AUC = 0.969 vs. 0.961, p = 0.129). The SUSI (computed in MNI space) performed slightly better than the SBR in the MULTI setting (AUC = 0.993 vs. 0.991, p = 0.207) and slightly worse in the MONO setting (AUC = 0.969 vs. AUC = 0.976, p = 0.259). There was a trend toward larger AUC difference between SUSI and SBR in the MULTI setting compared to the MONO setting (p = 0.073). Variability of voxel intensity in the reference region was larger in misclassified cases compared to correctly classified cases for both SUSI and SBR (MULTI setting: p = 0.007 and p = 0.012, respectively). CONCLUSIONS: The SUSI is particularly useful in MULTI settings. SPECT images should be stereotactically normalized prior to SUSI analysis. The putaminal SUSI provides better diagnostic performance than the SUSI of the whole striatum. Errors of the estimate of non-displaceable count density in the reference region can cause misclassification by both SUSI and SBR, particularly in borderline cases. These cases might be identified by visual checking FP-CIT uptake in the reference region for particularly high variability.
format Online
Article
Text
id pubmed-6505020
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-65050202019-05-28 Diagnostic performance of the specific uptake size index for semi-quantitative analysis of I-123-FP-CIT SPECT: harmonized multi-center research setting versus typical clinical single-camera setting Buchert, Ralph Lange, Catharina Spehl, Timo S. Apostolova, Ivayla Frings, Lars Jonsson, Cathrine Meyer, Philipp T. Hellwig, Sabine EJNMMI Res Original Research INTRODUCTION: The specific uptake size index (SUSI) of striatal FP-CIT uptake is independent of spatial resolution in the SPECT image, in contrast to the specific binding ratio (SBR). This suggests that the SUSI is particularly appropriate for multi-site/multi-camera settings in which camera-specific effects increase inter-subject variability of spatial resolution. However, the SUSI is sensitive to inter-subject variability of striatum size. Furthermore, it might be more sensitive to errors of the estimate of non-displaceable FP-CIT binding. This study compared SUSI and SBR in the multi-site/multi-camera (MULTI) setting of a prospective multi-center study and in a mono-site/mono-camera (MONO) setting representative of clinical routine. METHODS: The MULTI setting included patients with Parkinson’s disease (PD, n = 438) and healthy controls (n = 207) from the Parkinson Progression Marker Initiative. The MONO setting included 122 patients from routine clinical patient care in whom FP-CIT SPECT had been performed with the same double-head SPECT system according to the same acquisition and reconstruction protocol. Patients were categorized as “neurodegenerative” (n = 84) or “non-neurodegenerative” (n = 38) based on follow-up data. FP-CIT SPECTs were stereotactically normalized to MNI space. SUSI and SBR were computed for caudate, putamen, and whole striatum using unilateral ROIs predefined in MNI space. SUSI analysis was repeated in native patient space in the MONO setting. The area (AUC) under the ROC curve for identification of PD/“neurodegenerative” cases was used as performance measure. RESULTS: In both settings, the highest AUC was achieved by the putamen (minimum over both hemispheres), independent of the semi-quantitative method (SUSI or SBR). The putaminal SUSI provided slightly better performance with ROI analysis in MNI space compared to patient space (AUC = 0.969 vs. 0.961, p = 0.129). The SUSI (computed in MNI space) performed slightly better than the SBR in the MULTI setting (AUC = 0.993 vs. 0.991, p = 0.207) and slightly worse in the MONO setting (AUC = 0.969 vs. AUC = 0.976, p = 0.259). There was a trend toward larger AUC difference between SUSI and SBR in the MULTI setting compared to the MONO setting (p = 0.073). Variability of voxel intensity in the reference region was larger in misclassified cases compared to correctly classified cases for both SUSI and SBR (MULTI setting: p = 0.007 and p = 0.012, respectively). CONCLUSIONS: The SUSI is particularly useful in MULTI settings. SPECT images should be stereotactically normalized prior to SUSI analysis. The putaminal SUSI provides better diagnostic performance than the SUSI of the whole striatum. Errors of the estimate of non-displaceable count density in the reference region can cause misclassification by both SUSI and SBR, particularly in borderline cases. These cases might be identified by visual checking FP-CIT uptake in the reference region for particularly high variability. Springer Berlin Heidelberg 2019-05-07 /pmc/articles/PMC6505020/ /pubmed/31065816 http://dx.doi.org/10.1186/s13550-019-0506-9 Text en © The Author(s). 2019 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Original Research
Buchert, Ralph
Lange, Catharina
Spehl, Timo S.
Apostolova, Ivayla
Frings, Lars
Jonsson, Cathrine
Meyer, Philipp T.
Hellwig, Sabine
Diagnostic performance of the specific uptake size index for semi-quantitative analysis of I-123-FP-CIT SPECT: harmonized multi-center research setting versus typical clinical single-camera setting
title Diagnostic performance of the specific uptake size index for semi-quantitative analysis of I-123-FP-CIT SPECT: harmonized multi-center research setting versus typical clinical single-camera setting
title_full Diagnostic performance of the specific uptake size index for semi-quantitative analysis of I-123-FP-CIT SPECT: harmonized multi-center research setting versus typical clinical single-camera setting
title_fullStr Diagnostic performance of the specific uptake size index for semi-quantitative analysis of I-123-FP-CIT SPECT: harmonized multi-center research setting versus typical clinical single-camera setting
title_full_unstemmed Diagnostic performance of the specific uptake size index for semi-quantitative analysis of I-123-FP-CIT SPECT: harmonized multi-center research setting versus typical clinical single-camera setting
title_short Diagnostic performance of the specific uptake size index for semi-quantitative analysis of I-123-FP-CIT SPECT: harmonized multi-center research setting versus typical clinical single-camera setting
title_sort diagnostic performance of the specific uptake size index for semi-quantitative analysis of i-123-fp-cit spect: harmonized multi-center research setting versus typical clinical single-camera setting
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6505020/
https://www.ncbi.nlm.nih.gov/pubmed/31065816
http://dx.doi.org/10.1186/s13550-019-0506-9
work_keys_str_mv AT buchertralph diagnosticperformanceofthespecificuptakesizeindexforsemiquantitativeanalysisofi123fpcitspectharmonizedmulticenterresearchsettingversustypicalclinicalsinglecamerasetting
AT langecatharina diagnosticperformanceofthespecificuptakesizeindexforsemiquantitativeanalysisofi123fpcitspectharmonizedmulticenterresearchsettingversustypicalclinicalsinglecamerasetting
AT spehltimos diagnosticperformanceofthespecificuptakesizeindexforsemiquantitativeanalysisofi123fpcitspectharmonizedmulticenterresearchsettingversustypicalclinicalsinglecamerasetting
AT apostolovaivayla diagnosticperformanceofthespecificuptakesizeindexforsemiquantitativeanalysisofi123fpcitspectharmonizedmulticenterresearchsettingversustypicalclinicalsinglecamerasetting
AT fringslars diagnosticperformanceofthespecificuptakesizeindexforsemiquantitativeanalysisofi123fpcitspectharmonizedmulticenterresearchsettingversustypicalclinicalsinglecamerasetting
AT jonssoncathrine diagnosticperformanceofthespecificuptakesizeindexforsemiquantitativeanalysisofi123fpcitspectharmonizedmulticenterresearchsettingversustypicalclinicalsinglecamerasetting
AT meyerphilippt diagnosticperformanceofthespecificuptakesizeindexforsemiquantitativeanalysisofi123fpcitspectharmonizedmulticenterresearchsettingversustypicalclinicalsinglecamerasetting
AT hellwigsabine diagnosticperformanceofthespecificuptakesizeindexforsemiquantitativeanalysisofi123fpcitspectharmonizedmulticenterresearchsettingversustypicalclinicalsinglecamerasetting