A Comparative Study of Cyclic Fatigue of 10 Different Types of Endodontic Instruments: an in Vitro Study

OBJECTIVE: This study was designed to test the null hypothesis that there were no significant differences between size 25 files F360, F6 SkyTaper, Hyflex EDM, iRace, Neoniti, One Shape Protaper Next, Reciproc, Revo-S and Wave One Gold in terms of resistance to cyclic fatigue and length of broken fra...

Descripción completa

Detalles Bibliográficos
Autores principales: Rubio, Jorge, Zarzosa, José Ignacio, Pallarés, Antonio
Formato: Online Artículo Texto
Lenguaje:English
Publicado: University of Zagreb School of Dental Medicine, and Croatian Dental Society - Croatian Medical Association 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6508929/
https://www.ncbi.nlm.nih.gov/pubmed/31118530
http://dx.doi.org/10.15644/asc53/1/3
_version_ 1783417157569216512
author Rubio, Jorge
Zarzosa, José Ignacio
Pallarés, Antonio
author_facet Rubio, Jorge
Zarzosa, José Ignacio
Pallarés, Antonio
author_sort Rubio, Jorge
collection PubMed
description OBJECTIVE: This study was designed to test the null hypothesis that there were no significant differences between size 25 files F360, F6 SkyTaper, Hyflex EDM, iRace, Neoniti, One Shape Protaper Next, Reciproc, Revo-S and Wave One Gold in terms of resistance to cyclic fatigue and length of broken fragments. MATERIAL AND METHODS: 300 new size 25 files of the systems studied were selected (n=30). The instruments were mechanized with a X-Smart Plus endo motor at the speed and torque recommended by the manufacturer, holding the instruments steady with a clamping mechanism, with passive adjustment and without pressure in a stainless-steel block. The time was calculated in seconds until fracture. The number of fatigue cycles was calculated as (Resistance (s) x Speed)/60. The separated fragment lengths were measured with a digital Vernier calliper. A statistical analysis was performed with the SPSS 18 programme at a 95% confidence level, using the Levene´s Test to compare variances, the Welch’s Test to compare means, and the Games-Howell´s Test to reveal differences between groups. RESULTS: The Levene’s Test showed no equal variances (P<0.05). The Welch’s Test (P<0.05) showed significant differences in cyclic fatigue and separated fragment lengths. The Games-Howell test (P<0.05) exhibited significant differences in multiple comparisons, (P<0.05). CONCLUSIONS: The systems with CM-Wire (Hyflex EDM and Neoniti) were superior in resistance to the other systems for cyclic fatigue. For separated fragment lengths, F360 (conventional NiTi) and Reciproc (M-Wire) were better significantly better in terms of resistance.
format Online
Article
Text
id pubmed-6508929
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher University of Zagreb School of Dental Medicine, and Croatian Dental Society - Croatian Medical Association
record_format MEDLINE/PubMed
spelling pubmed-65089292019-05-22 A Comparative Study of Cyclic Fatigue of 10 Different Types of Endodontic Instruments: an in Vitro Study Rubio, Jorge Zarzosa, José Ignacio Pallarés, Antonio Acta Stomatol Croat Original Scientific Papers OBJECTIVE: This study was designed to test the null hypothesis that there were no significant differences between size 25 files F360, F6 SkyTaper, Hyflex EDM, iRace, Neoniti, One Shape Protaper Next, Reciproc, Revo-S and Wave One Gold in terms of resistance to cyclic fatigue and length of broken fragments. MATERIAL AND METHODS: 300 new size 25 files of the systems studied were selected (n=30). The instruments were mechanized with a X-Smart Plus endo motor at the speed and torque recommended by the manufacturer, holding the instruments steady with a clamping mechanism, with passive adjustment and without pressure in a stainless-steel block. The time was calculated in seconds until fracture. The number of fatigue cycles was calculated as (Resistance (s) x Speed)/60. The separated fragment lengths were measured with a digital Vernier calliper. A statistical analysis was performed with the SPSS 18 programme at a 95% confidence level, using the Levene´s Test to compare variances, the Welch’s Test to compare means, and the Games-Howell´s Test to reveal differences between groups. RESULTS: The Levene’s Test showed no equal variances (P<0.05). The Welch’s Test (P<0.05) showed significant differences in cyclic fatigue and separated fragment lengths. The Games-Howell test (P<0.05) exhibited significant differences in multiple comparisons, (P<0.05). CONCLUSIONS: The systems with CM-Wire (Hyflex EDM and Neoniti) were superior in resistance to the other systems for cyclic fatigue. For separated fragment lengths, F360 (conventional NiTi) and Reciproc (M-Wire) were better significantly better in terms of resistance. University of Zagreb School of Dental Medicine, and Croatian Dental Society - Croatian Medical Association 2019-03 /pmc/articles/PMC6508929/ /pubmed/31118530 http://dx.doi.org/10.15644/asc53/1/3 Text en http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial No Derivatives (CC BY-NC-ND) 4.0 License.
spellingShingle Original Scientific Papers
Rubio, Jorge
Zarzosa, José Ignacio
Pallarés, Antonio
A Comparative Study of Cyclic Fatigue of 10 Different Types of Endodontic Instruments: an in Vitro Study
title A Comparative Study of Cyclic Fatigue of 10 Different Types of Endodontic Instruments: an in Vitro Study
title_full A Comparative Study of Cyclic Fatigue of 10 Different Types of Endodontic Instruments: an in Vitro Study
title_fullStr A Comparative Study of Cyclic Fatigue of 10 Different Types of Endodontic Instruments: an in Vitro Study
title_full_unstemmed A Comparative Study of Cyclic Fatigue of 10 Different Types of Endodontic Instruments: an in Vitro Study
title_short A Comparative Study of Cyclic Fatigue of 10 Different Types of Endodontic Instruments: an in Vitro Study
title_sort comparative study of cyclic fatigue of 10 different types of endodontic instruments: an in vitro study
topic Original Scientific Papers
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6508929/
https://www.ncbi.nlm.nih.gov/pubmed/31118530
http://dx.doi.org/10.15644/asc53/1/3
work_keys_str_mv AT rubiojorge acomparativestudyofcyclicfatigueof10differenttypesofendodonticinstrumentsaninvitrostudy
AT zarzosajoseignacio acomparativestudyofcyclicfatigueof10differenttypesofendodonticinstrumentsaninvitrostudy
AT pallaresantonio acomparativestudyofcyclicfatigueof10differenttypesofendodonticinstrumentsaninvitrostudy
AT rubiojorge comparativestudyofcyclicfatigueof10differenttypesofendodonticinstrumentsaninvitrostudy
AT zarzosajoseignacio comparativestudyofcyclicfatigueof10differenttypesofendodonticinstrumentsaninvitrostudy
AT pallaresantonio comparativestudyofcyclicfatigueof10differenttypesofendodonticinstrumentsaninvitrostudy