Performance characteristics of silicon photomultiplier based 15-cm AFOV TOF PET/CT
BACKGROUND: This paper describes the National Electrical Manufacturers Association (NEMA) system performance of the Discovery MI 3-ring PET/CT (GE Healthcare) installed in Bruges, Belgium. This time-of-flight (TOF) PET camera is based on silicon photomultipliers instead of photomultiplier tubes. MET...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer International Publishing
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6510743/ https://www.ncbi.nlm.nih.gov/pubmed/31076884 http://dx.doi.org/10.1186/s40658-019-0244-0 |
_version_ | 1783417457319346176 |
---|---|
author | Vandendriessche, Delphine Uribe, Jorge Bertin, Hugo De Geeter, Frank |
author_facet | Vandendriessche, Delphine Uribe, Jorge Bertin, Hugo De Geeter, Frank |
author_sort | Vandendriessche, Delphine |
collection | PubMed |
description | BACKGROUND: This paper describes the National Electrical Manufacturers Association (NEMA) system performance of the Discovery MI 3-ring PET/CT (GE Healthcare) installed in Bruges, Belgium. This time-of-flight (TOF) PET camera is based on silicon photomultipliers instead of photomultiplier tubes. METHODS: The NEMA NU2-2012 standard was used to evaluate spatial resolution, sensitivity, image quality (IQ) and count rate curves of the system. Timing and energy resolution were determined. RESULTS: Full width at half maximum (FWHM) of spatial resolution in radial, tangential and axial direction was 4.69, 4.08 and 4.68 mm at 1 cm; 5.58, 4.64 and 5.83 mm at 10 cm; and 7.53, 5.08 and 5.47 mm at 20 cm from the centre of the field of view (FOV) for the filtered backprojection reconstruction. For non-TOF ordered subset expectation maximization (OSEM) reconstruction without point spread function (PSF) correction, FWHM was 3.87, 3.69 and 4.15 mm at 1 cm; 4.80, 3.81 and 4.87 mm at 10 cm; and 7.38, 4.16 and 3.98 mm at 20 cm. Sensitivity was 7.258 cps/kBq at the centre of the FOV and 7.117 cps/kBq at 10-cm radial offset. Contrast recovery (CR) using the IQ phantom for the TOF OSEM reconstruction without PSF correction was 47.4, 59.3, 67.0 and 77.0% for the 10-, 13-, 17- and 22-mm radioactive spheres and 82.5 and 85.1% for the 28- and 37-mm non-radioactive spheres. Background variability (BV) was 16.4, 12.1, 9.1, 6.6, 5.1 and 3.8% for the 10-, 13-, 17-, 22-, 28- and 37-mm spheres. Lung error was 8.5%. Peak noise equivalent count rate (NECR) was 102.3 kcps at 23.0 kBq/ml with a scatter fraction of 41.2%. Maximum accuracy error was 3.88%. Coincidence timing resolution was 375.6 ps FWHM. Energy resolution was 9.3% FWHM. Q.Clear reconstruction significantly improved CR and reduced BV compared with OSEM. CONCLUSION: System sensitivity and NECR are lower and IQ phantom’s BV is higher compared with larger axial FOV (AFOV) scanners like the 4-ring discovery MI, as expected from the smaller solid angle of the 3-ring system. The other NEMA performance parameters are all comparable with those of the larger AFOV scanners. |
format | Online Article Text |
id | pubmed-6510743 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Springer International Publishing |
record_format | MEDLINE/PubMed |
spelling | pubmed-65107432019-05-28 Performance characteristics of silicon photomultiplier based 15-cm AFOV TOF PET/CT Vandendriessche, Delphine Uribe, Jorge Bertin, Hugo De Geeter, Frank EJNMMI Phys Original Research BACKGROUND: This paper describes the National Electrical Manufacturers Association (NEMA) system performance of the Discovery MI 3-ring PET/CT (GE Healthcare) installed in Bruges, Belgium. This time-of-flight (TOF) PET camera is based on silicon photomultipliers instead of photomultiplier tubes. METHODS: The NEMA NU2-2012 standard was used to evaluate spatial resolution, sensitivity, image quality (IQ) and count rate curves of the system. Timing and energy resolution were determined. RESULTS: Full width at half maximum (FWHM) of spatial resolution in radial, tangential and axial direction was 4.69, 4.08 and 4.68 mm at 1 cm; 5.58, 4.64 and 5.83 mm at 10 cm; and 7.53, 5.08 and 5.47 mm at 20 cm from the centre of the field of view (FOV) for the filtered backprojection reconstruction. For non-TOF ordered subset expectation maximization (OSEM) reconstruction without point spread function (PSF) correction, FWHM was 3.87, 3.69 and 4.15 mm at 1 cm; 4.80, 3.81 and 4.87 mm at 10 cm; and 7.38, 4.16 and 3.98 mm at 20 cm. Sensitivity was 7.258 cps/kBq at the centre of the FOV and 7.117 cps/kBq at 10-cm radial offset. Contrast recovery (CR) using the IQ phantom for the TOF OSEM reconstruction without PSF correction was 47.4, 59.3, 67.0 and 77.0% for the 10-, 13-, 17- and 22-mm radioactive spheres and 82.5 and 85.1% for the 28- and 37-mm non-radioactive spheres. Background variability (BV) was 16.4, 12.1, 9.1, 6.6, 5.1 and 3.8% for the 10-, 13-, 17-, 22-, 28- and 37-mm spheres. Lung error was 8.5%. Peak noise equivalent count rate (NECR) was 102.3 kcps at 23.0 kBq/ml with a scatter fraction of 41.2%. Maximum accuracy error was 3.88%. Coincidence timing resolution was 375.6 ps FWHM. Energy resolution was 9.3% FWHM. Q.Clear reconstruction significantly improved CR and reduced BV compared with OSEM. CONCLUSION: System sensitivity and NECR are lower and IQ phantom’s BV is higher compared with larger axial FOV (AFOV) scanners like the 4-ring discovery MI, as expected from the smaller solid angle of the 3-ring system. The other NEMA performance parameters are all comparable with those of the larger AFOV scanners. Springer International Publishing 2019-05-10 /pmc/articles/PMC6510743/ /pubmed/31076884 http://dx.doi.org/10.1186/s40658-019-0244-0 Text en © The Author(s). 2019 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. |
spellingShingle | Original Research Vandendriessche, Delphine Uribe, Jorge Bertin, Hugo De Geeter, Frank Performance characteristics of silicon photomultiplier based 15-cm AFOV TOF PET/CT |
title | Performance characteristics of silicon photomultiplier based 15-cm AFOV TOF PET/CT |
title_full | Performance characteristics of silicon photomultiplier based 15-cm AFOV TOF PET/CT |
title_fullStr | Performance characteristics of silicon photomultiplier based 15-cm AFOV TOF PET/CT |
title_full_unstemmed | Performance characteristics of silicon photomultiplier based 15-cm AFOV TOF PET/CT |
title_short | Performance characteristics of silicon photomultiplier based 15-cm AFOV TOF PET/CT |
title_sort | performance characteristics of silicon photomultiplier based 15-cm afov tof pet/ct |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6510743/ https://www.ncbi.nlm.nih.gov/pubmed/31076884 http://dx.doi.org/10.1186/s40658-019-0244-0 |
work_keys_str_mv | AT vandendriesschedelphine performancecharacteristicsofsiliconphotomultiplierbased15cmafovtofpetct AT uribejorge performancecharacteristicsofsiliconphotomultiplierbased15cmafovtofpetct AT bertinhugo performancecharacteristicsofsiliconphotomultiplierbased15cmafovtofpetct AT degeeterfrank performancecharacteristicsofsiliconphotomultiplierbased15cmafovtofpetct |