Cargando…

Optimising quantitative (90)Y PET imaging: an investigation into the effects of scan length and Bayesian penalised likelihood reconstruction

BACKGROUND: Positron emission tomography (PET) imaging of (90)Y following selective internal radiation therapy (SIRT) is possible, but image quality is poor, and therefore, accurate quantification and dosimetry are challenging. This study aimed to quantitatively optimise (90)Y PET imaging using a ne...

Descripción completa

Detalles Bibliográficos
Autores principales: Scott, Nathaniel P., McGowan, Daniel R.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6510762/
https://www.ncbi.nlm.nih.gov/pubmed/31076913
http://dx.doi.org/10.1186/s13550-019-0512-y
_version_ 1783417461833465856
author Scott, Nathaniel P.
McGowan, Daniel R.
author_facet Scott, Nathaniel P.
McGowan, Daniel R.
author_sort Scott, Nathaniel P.
collection PubMed
description BACKGROUND: Positron emission tomography (PET) imaging of (90)Y following selective internal radiation therapy (SIRT) is possible, but image quality is poor, and therefore, accurate quantification and dosimetry are challenging. This study aimed to quantitatively optimise (90)Y PET imaging using a new Bayesian penalised likelihood (BPL) reconstruction algorithm (Q.Clear, GE Healthcare). The length of time per bed was also investigated to study its impact on quantification accuracy. METHODS: A NEMA IQ phantom with an 8:1 sphere-to-background ratio was scanned overnight on a GE Discovery 710 PET/CT scanner. Datasets were rebinned into varying lengths of time (5–60 min); the 15-min rebins were reconstructed using BPL reconstruction with a range of noise penalisation weighting factors (beta values). The metrics of contrast recovery (CR), background variability (BV), and recovered activity percentage (RAP) were calculated in order to identify the optimum beta value. Reconstructions were then carried out on the rest of the timing datasets using the optimised beta value; the same metrics were used to assess the quantification accuracy of the reconstructed images. RESULTS: A beta value of 1000 produced the highest CR and RAP (76% and 73%, 37 mm sphere) without overly accentuating the noise (BV) in the image. There was no statistically significant increase (p < 0.05) in either the CR or RAP for scan times of > 15 min. For the 5-min acquisitions, there was a statistically significant decrease in RAP (28 mm sphere, p < 0.01) when compared to the 15-min acquisition. CONCLUSION: Our results indicate that an acquisition length of 15 min and beta value of 1000 (when using Q.Clear reconstruction) are optimum for quantitative (90)Y PET imaging. Increasing the acquisition time to more than 15 min reduces the image noise but has no significant impact on image quantification. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (10.1186/s13550-019-0512-y) contains supplementary material, which is available to authorized users.
format Online
Article
Text
id pubmed-6510762
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-65107622019-05-28 Optimising quantitative (90)Y PET imaging: an investigation into the effects of scan length and Bayesian penalised likelihood reconstruction Scott, Nathaniel P. McGowan, Daniel R. EJNMMI Res Original Research BACKGROUND: Positron emission tomography (PET) imaging of (90)Y following selective internal radiation therapy (SIRT) is possible, but image quality is poor, and therefore, accurate quantification and dosimetry are challenging. This study aimed to quantitatively optimise (90)Y PET imaging using a new Bayesian penalised likelihood (BPL) reconstruction algorithm (Q.Clear, GE Healthcare). The length of time per bed was also investigated to study its impact on quantification accuracy. METHODS: A NEMA IQ phantom with an 8:1 sphere-to-background ratio was scanned overnight on a GE Discovery 710 PET/CT scanner. Datasets were rebinned into varying lengths of time (5–60 min); the 15-min rebins were reconstructed using BPL reconstruction with a range of noise penalisation weighting factors (beta values). The metrics of contrast recovery (CR), background variability (BV), and recovered activity percentage (RAP) were calculated in order to identify the optimum beta value. Reconstructions were then carried out on the rest of the timing datasets using the optimised beta value; the same metrics were used to assess the quantification accuracy of the reconstructed images. RESULTS: A beta value of 1000 produced the highest CR and RAP (76% and 73%, 37 mm sphere) without overly accentuating the noise (BV) in the image. There was no statistically significant increase (p < 0.05) in either the CR or RAP for scan times of > 15 min. For the 5-min acquisitions, there was a statistically significant decrease in RAP (28 mm sphere, p < 0.01) when compared to the 15-min acquisition. CONCLUSION: Our results indicate that an acquisition length of 15 min and beta value of 1000 (when using Q.Clear reconstruction) are optimum for quantitative (90)Y PET imaging. Increasing the acquisition time to more than 15 min reduces the image noise but has no significant impact on image quantification. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (10.1186/s13550-019-0512-y) contains supplementary material, which is available to authorized users. Springer Berlin Heidelberg 2019-05-10 /pmc/articles/PMC6510762/ /pubmed/31076913 http://dx.doi.org/10.1186/s13550-019-0512-y Text en © The Author(s). 2019 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Original Research
Scott, Nathaniel P.
McGowan, Daniel R.
Optimising quantitative (90)Y PET imaging: an investigation into the effects of scan length and Bayesian penalised likelihood reconstruction
title Optimising quantitative (90)Y PET imaging: an investigation into the effects of scan length and Bayesian penalised likelihood reconstruction
title_full Optimising quantitative (90)Y PET imaging: an investigation into the effects of scan length and Bayesian penalised likelihood reconstruction
title_fullStr Optimising quantitative (90)Y PET imaging: an investigation into the effects of scan length and Bayesian penalised likelihood reconstruction
title_full_unstemmed Optimising quantitative (90)Y PET imaging: an investigation into the effects of scan length and Bayesian penalised likelihood reconstruction
title_short Optimising quantitative (90)Y PET imaging: an investigation into the effects of scan length and Bayesian penalised likelihood reconstruction
title_sort optimising quantitative (90)y pet imaging: an investigation into the effects of scan length and bayesian penalised likelihood reconstruction
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6510762/
https://www.ncbi.nlm.nih.gov/pubmed/31076913
http://dx.doi.org/10.1186/s13550-019-0512-y
work_keys_str_mv AT scottnathanielp optimisingquantitative90ypetimaginganinvestigationintotheeffectsofscanlengthandbayesianpenalisedlikelihoodreconstruction
AT mcgowandanielr optimisingquantitative90ypetimaginganinvestigationintotheeffectsofscanlengthandbayesianpenalisedlikelihoodreconstruction