Cargando…

Repeatability and reproducibility of FreeSurfer, FSL-SIENAX and SPM brain volumetric measurements and the effect of lesion filling in multiple sclerosis

OBJECTIVES: To compare the cross-sectional robustness of commonly used volumetric software and effects of lesion filling in multiple sclerosis (MS). METHODS: Nine MS patients (six females; age 38±13 years, disease duration 7.3±5.2 years) were scanned twice with repositioning on three MRI scanners (S...

Descripción completa

Detalles Bibliográficos
Autores principales: Guo, Chunjie, Ferreira, Daniel, Fink, Katarina, Westman, Eric, Granberg, Tobias
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6510869/
https://www.ncbi.nlm.nih.gov/pubmed/30242503
http://dx.doi.org/10.1007/s00330-018-5710-x
_version_ 1783417486843052032
author Guo, Chunjie
Ferreira, Daniel
Fink, Katarina
Westman, Eric
Granberg, Tobias
author_facet Guo, Chunjie
Ferreira, Daniel
Fink, Katarina
Westman, Eric
Granberg, Tobias
author_sort Guo, Chunjie
collection PubMed
description OBJECTIVES: To compare the cross-sectional robustness of commonly used volumetric software and effects of lesion filling in multiple sclerosis (MS). METHODS: Nine MS patients (six females; age 38±13 years, disease duration 7.3±5.2 years) were scanned twice with repositioning on three MRI scanners (Siemens Aera 1.5T, Avanto 1.5T, Trio 3.0T) the same day. Volumetric T(1)-weighted images were processed with FreeSurfer, FSL-SIENAX, SPM and SPM-CAT before and after 3D FLAIR lesion filling with LST. The whole-brain, grey matter (GM) and white matter (WM) volumes were calculated with and without normalisation to the intracranial volume or FSL-SIENAX scaling factor. Robustness was assessed using the coefficient of variation (CoV). RESULTS: Variability in volumetrics was lower within than between scanners (CoV 0.17–0.96% vs. 0.65–5.0%, p<0.001). All software provided similarly robust segmentations of the brain volume on the same scanner (CoV 0.17–0.28%, p=0.076). Normalisation improved inter-scanner reproducibility in FreeSurfer and SPM-based methods, but the FSL-SIENAX scaling factor did not improve robustness. Generally, SPM-based methods produced the most consistent volumetrics, while FreeSurfer was more robust for WM volumes on different scanners. FreeSurfer had more robust normalised brain and GM volumes on different scanners than FSL-SIENAX (p=0.004). MS lesion filling changed the output of FSL-SIENAX, SPM and SPM-CAT but not FreeSurfer. CONCLUSIONS: Consistent use of the same scanner is essential and normalisation to the intracranial volume is recommended for multiple scanners. Based on robustness, SPM-based methods are particularly suitable for cross-sectional volumetry. FreeSurfer poses a suitable alternative with WM segmentations less sensitive to MS lesions. KEY POINTS: • The same scanner should be used for brain volumetry. If different scanners are used, the intracranial volume normalisation improves the FreeSurfer and SPM robustness (but not the FSL scaling factor). • FreeSurfer, FSL and SPM all provide robust measures of the whole brain volume on the same MRI scanner. SPM-based methods overall provide the most robust segmentations (except white matter segmentations on different scanners where FreeSurfer is more robust). • MS lesion filling with Lesion Segmentation Toolbox changes the output of FSL-SIENAX and SPM. FreeSurfer output is not affected by MS lesion filling since it already takes white matter hypointensities into account and is therefore particularly suitable for MS brain volumetry. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (10.1007/s00330-018-5710-x) contains supplementary material, which is available to authorized users.
format Online
Article
Text
id pubmed-6510869
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-65108692019-05-28 Repeatability and reproducibility of FreeSurfer, FSL-SIENAX and SPM brain volumetric measurements and the effect of lesion filling in multiple sclerosis Guo, Chunjie Ferreira, Daniel Fink, Katarina Westman, Eric Granberg, Tobias Eur Radiol Neuro OBJECTIVES: To compare the cross-sectional robustness of commonly used volumetric software and effects of lesion filling in multiple sclerosis (MS). METHODS: Nine MS patients (six females; age 38±13 years, disease duration 7.3±5.2 years) were scanned twice with repositioning on three MRI scanners (Siemens Aera 1.5T, Avanto 1.5T, Trio 3.0T) the same day. Volumetric T(1)-weighted images were processed with FreeSurfer, FSL-SIENAX, SPM and SPM-CAT before and after 3D FLAIR lesion filling with LST. The whole-brain, grey matter (GM) and white matter (WM) volumes were calculated with and without normalisation to the intracranial volume or FSL-SIENAX scaling factor. Robustness was assessed using the coefficient of variation (CoV). RESULTS: Variability in volumetrics was lower within than between scanners (CoV 0.17–0.96% vs. 0.65–5.0%, p<0.001). All software provided similarly robust segmentations of the brain volume on the same scanner (CoV 0.17–0.28%, p=0.076). Normalisation improved inter-scanner reproducibility in FreeSurfer and SPM-based methods, but the FSL-SIENAX scaling factor did not improve robustness. Generally, SPM-based methods produced the most consistent volumetrics, while FreeSurfer was more robust for WM volumes on different scanners. FreeSurfer had more robust normalised brain and GM volumes on different scanners than FSL-SIENAX (p=0.004). MS lesion filling changed the output of FSL-SIENAX, SPM and SPM-CAT but not FreeSurfer. CONCLUSIONS: Consistent use of the same scanner is essential and normalisation to the intracranial volume is recommended for multiple scanners. Based on robustness, SPM-based methods are particularly suitable for cross-sectional volumetry. FreeSurfer poses a suitable alternative with WM segmentations less sensitive to MS lesions. KEY POINTS: • The same scanner should be used for brain volumetry. If different scanners are used, the intracranial volume normalisation improves the FreeSurfer and SPM robustness (but not the FSL scaling factor). • FreeSurfer, FSL and SPM all provide robust measures of the whole brain volume on the same MRI scanner. SPM-based methods overall provide the most robust segmentations (except white matter segmentations on different scanners where FreeSurfer is more robust). • MS lesion filling with Lesion Segmentation Toolbox changes the output of FSL-SIENAX and SPM. FreeSurfer output is not affected by MS lesion filling since it already takes white matter hypointensities into account and is therefore particularly suitable for MS brain volumetry. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (10.1007/s00330-018-5710-x) contains supplementary material, which is available to authorized users. Springer Berlin Heidelberg 2018-09-21 2019 /pmc/articles/PMC6510869/ /pubmed/30242503 http://dx.doi.org/10.1007/s00330-018-5710-x Text en © The Author(s) 2018 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Neuro
Guo, Chunjie
Ferreira, Daniel
Fink, Katarina
Westman, Eric
Granberg, Tobias
Repeatability and reproducibility of FreeSurfer, FSL-SIENAX and SPM brain volumetric measurements and the effect of lesion filling in multiple sclerosis
title Repeatability and reproducibility of FreeSurfer, FSL-SIENAX and SPM brain volumetric measurements and the effect of lesion filling in multiple sclerosis
title_full Repeatability and reproducibility of FreeSurfer, FSL-SIENAX and SPM brain volumetric measurements and the effect of lesion filling in multiple sclerosis
title_fullStr Repeatability and reproducibility of FreeSurfer, FSL-SIENAX and SPM brain volumetric measurements and the effect of lesion filling in multiple sclerosis
title_full_unstemmed Repeatability and reproducibility of FreeSurfer, FSL-SIENAX and SPM brain volumetric measurements and the effect of lesion filling in multiple sclerosis
title_short Repeatability and reproducibility of FreeSurfer, FSL-SIENAX and SPM brain volumetric measurements and the effect of lesion filling in multiple sclerosis
title_sort repeatability and reproducibility of freesurfer, fsl-sienax and spm brain volumetric measurements and the effect of lesion filling in multiple sclerosis
topic Neuro
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6510869/
https://www.ncbi.nlm.nih.gov/pubmed/30242503
http://dx.doi.org/10.1007/s00330-018-5710-x
work_keys_str_mv AT guochunjie repeatabilityandreproducibilityoffreesurferfslsienaxandspmbrainvolumetricmeasurementsandtheeffectoflesionfillinginmultiplesclerosis
AT ferreiradaniel repeatabilityandreproducibilityoffreesurferfslsienaxandspmbrainvolumetricmeasurementsandtheeffectoflesionfillinginmultiplesclerosis
AT finkkatarina repeatabilityandreproducibilityoffreesurferfslsienaxandspmbrainvolumetricmeasurementsandtheeffectoflesionfillinginmultiplesclerosis
AT westmaneric repeatabilityandreproducibilityoffreesurferfslsienaxandspmbrainvolumetricmeasurementsandtheeffectoflesionfillinginmultiplesclerosis
AT granbergtobias repeatabilityandreproducibilityoffreesurferfslsienaxandspmbrainvolumetricmeasurementsandtheeffectoflesionfillinginmultiplesclerosis