Cargando…

Repeatability and Agreement of Macular Thickness Measurements Obtained with Two Different Scan Modes of the Optovue RTVue Optical Coherence Tomography Device

OBJECTIVES: To evaluate the repeatability and agreement of macular thickness measurements obtained with E-MM5 and MM6, two different scan modes, on the Optovue RTVue optic coherence tomography (OCT) device. MATERIALS AND METHODS: Three consecutive macular thickness measurements in 30 healthy volunte...

Descripción completa

Detalles Bibliográficos
Autores principales: Kola, Mehmet, Önal, Mehmet, Türk, Adem, Erdöl, Hidayet
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Galenos Publishing 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6517858/
https://www.ncbi.nlm.nih.gov/pubmed/31055892
http://dx.doi.org/10.4274/tjo.galenos.2018.88972
_version_ 1783418337913470976
author Kola, Mehmet
Önal, Mehmet
Türk, Adem
Erdöl, Hidayet
author_facet Kola, Mehmet
Önal, Mehmet
Türk, Adem
Erdöl, Hidayet
author_sort Kola, Mehmet
collection PubMed
description OBJECTIVES: To evaluate the repeatability and agreement of macular thickness measurements obtained with E-MM5 and MM6, two different scan modes, on the Optovue RTVue optic coherence tomography (OCT) device. MATERIALS AND METHODS: Three consecutive macular thickness measurements in 30 healthy volunteers were taken using the OCT device E-MM5 and MM6 scan modes. The repeatability and agreement of these measurements obtained from the two scan modes and divided into nine anatomical regions based on early treatment diabetic retinopathy study were subjected to statistical analysis. RESULTS: The mean age of the participants was 29.7±6.39 years. Intraclass correlation (all ICC values ≥0.86) and coefficient of variation (all coefficient of variation values ≤2%) analyses of consecutive OCT measurements in the nine regions of the macula obtained in both E-MM5 and MM6 scan modes gave high repeatability rates. Mean macular thickness values in the foveal region were 243.76±21.79 μm in E-MM5 mode and 247.04±19.83 μm in MM6 mode (p=0.543). Values for measurements obtained in E-MM5 and MM6 scan modes in parafoveal macular regions were also statistically similar (p>0.05 for all). However, a statistically significant difference was observed between the two modes in perifoveal macular measurements, except in the superior region. CONCLUSION: The Optovue RTVue OCT device gives highly repeatable measurement results for macular thicknesses in both E-MM5 and MM6 scan modes. However, it should be considered that measurements performed in E-MM5 and MM6 modes give different results in perifoveal regions.
format Online
Article
Text
id pubmed-6517858
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Galenos Publishing
record_format MEDLINE/PubMed
spelling pubmed-65178582019-05-28 Repeatability and Agreement of Macular Thickness Measurements Obtained with Two Different Scan Modes of the Optovue RTVue Optical Coherence Tomography Device Kola, Mehmet Önal, Mehmet Türk, Adem Erdöl, Hidayet Turk J Ophthalmol Original Article OBJECTIVES: To evaluate the repeatability and agreement of macular thickness measurements obtained with E-MM5 and MM6, two different scan modes, on the Optovue RTVue optic coherence tomography (OCT) device. MATERIALS AND METHODS: Three consecutive macular thickness measurements in 30 healthy volunteers were taken using the OCT device E-MM5 and MM6 scan modes. The repeatability and agreement of these measurements obtained from the two scan modes and divided into nine anatomical regions based on early treatment diabetic retinopathy study were subjected to statistical analysis. RESULTS: The mean age of the participants was 29.7±6.39 years. Intraclass correlation (all ICC values ≥0.86) and coefficient of variation (all coefficient of variation values ≤2%) analyses of consecutive OCT measurements in the nine regions of the macula obtained in both E-MM5 and MM6 scan modes gave high repeatability rates. Mean macular thickness values in the foveal region were 243.76±21.79 μm in E-MM5 mode and 247.04±19.83 μm in MM6 mode (p=0.543). Values for measurements obtained in E-MM5 and MM6 scan modes in parafoveal macular regions were also statistically similar (p>0.05 for all). However, a statistically significant difference was observed between the two modes in perifoveal macular measurements, except in the superior region. CONCLUSION: The Optovue RTVue OCT device gives highly repeatable measurement results for macular thicknesses in both E-MM5 and MM6 scan modes. However, it should be considered that measurements performed in E-MM5 and MM6 modes give different results in perifoveal regions. Galenos Publishing 2019-04 2019-04-30 /pmc/articles/PMC6517858/ /pubmed/31055892 http://dx.doi.org/10.4274/tjo.galenos.2018.88972 Text en © Copyright 2019 by Turkish Ophthalmological Association | Turkish Journal of Ophthalmology, published by Galenos Publishing House. http://creativecommons.org/licenses/by/2.5/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Kola, Mehmet
Önal, Mehmet
Türk, Adem
Erdöl, Hidayet
Repeatability and Agreement of Macular Thickness Measurements Obtained with Two Different Scan Modes of the Optovue RTVue Optical Coherence Tomography Device
title Repeatability and Agreement of Macular Thickness Measurements Obtained with Two Different Scan Modes of the Optovue RTVue Optical Coherence Tomography Device
title_full Repeatability and Agreement of Macular Thickness Measurements Obtained with Two Different Scan Modes of the Optovue RTVue Optical Coherence Tomography Device
title_fullStr Repeatability and Agreement of Macular Thickness Measurements Obtained with Two Different Scan Modes of the Optovue RTVue Optical Coherence Tomography Device
title_full_unstemmed Repeatability and Agreement of Macular Thickness Measurements Obtained with Two Different Scan Modes of the Optovue RTVue Optical Coherence Tomography Device
title_short Repeatability and Agreement of Macular Thickness Measurements Obtained with Two Different Scan Modes of the Optovue RTVue Optical Coherence Tomography Device
title_sort repeatability and agreement of macular thickness measurements obtained with two different scan modes of the optovue rtvue optical coherence tomography device
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6517858/
https://www.ncbi.nlm.nih.gov/pubmed/31055892
http://dx.doi.org/10.4274/tjo.galenos.2018.88972
work_keys_str_mv AT kolamehmet repeatabilityandagreementofmacularthicknessmeasurementsobtainedwithtwodifferentscanmodesoftheoptovuertvueopticalcoherencetomographydevice
AT onalmehmet repeatabilityandagreementofmacularthicknessmeasurementsobtainedwithtwodifferentscanmodesoftheoptovuertvueopticalcoherencetomographydevice
AT turkadem repeatabilityandagreementofmacularthicknessmeasurementsobtainedwithtwodifferentscanmodesoftheoptovuertvueopticalcoherencetomographydevice
AT erdolhidayet repeatabilityandagreementofmacularthicknessmeasurementsobtainedwithtwodifferentscanmodesoftheoptovuertvueopticalcoherencetomographydevice