Electron streams in air during magnetic-resonance image-guided radiation therapy

To investigate the undesired irradiations outside of the treatment field by electron streams in air (air-electron-stream) during magnetic-resonance image-guided radiation therapy (MR-IGRT). A custom-made support phantom adjusting angles between the beam central axis (CAX) and the phantom surface (te...

Descripción completa

Detalles Bibliográficos
Autores principales: An, Hyun Joon, Kim, Jung-in, Park, Jong Min
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6519819/
https://www.ncbi.nlm.nih.gov/pubmed/31091270
http://dx.doi.org/10.1371/journal.pone.0216965
_version_ 1783418670519681024
author An, Hyun Joon
Kim, Jung-in
Park, Jong Min
author_facet An, Hyun Joon
Kim, Jung-in
Park, Jong Min
author_sort An, Hyun Joon
collection PubMed
description To investigate the undesired irradiations outside of the treatment field by electron streams in air (air-electron-stream) during magnetic-resonance image-guided radiation therapy (MR-IGRT). A custom-made support phantom adjusting angles between the beam central axis (CAX) and the phantom surface (termed phantom-angles), were used. Using the ViewRay system, a rectangular parallelepiped phantom placed on the support phantom, was irradiated with field sizes of 6.3 cm × 6.3 cm (FS6.3) and 12.6 cm × 12.6 cm (FS12.6) at gantry angles of 0°, 30°, and 330°, and phantom-angles of 10°, 20°, and 30°. For each beam delivery, the isocenter was located at the center of mass of the phantom and 3 Gy was delivered to the isocenter (prescription dose = 3 Gy). The doses given by the air-electron-streams were measured using the EBT3 films on the panels placed orthogonal to the direction of the magnetic field at distances of 10 and 17 cm from CAX. Two dose distributions per irradiation were measured on the panel facing the phantom surface of the incident beam (front panel) and on the panel facing the phantom surface of the beam exit (end panel). We investigated the doses by the air-electron-streams by calculating the average doses inside the circles drawn around a point of the maximum dose with radii of x cm (D(Rx)) from the dose distributions on the panels (x = 1–5 cm). The largest value of D(Rx) was D(R1) (1.64 Gy, 55% of the prescription dose) at 10 cm distance from CAX, with FS12.6, at 30° phantom-angle and 330° gantry angle. The average difference of the D(R1) at the end panels (FS12.6) between the calculations and measurements was 1.36 Gy. The average global gamma passing rate with 3%/3 mm on the dose distributions at the end panels (FS12.6) was 40.3%. The calculated dose distributions on both panels were not coincident with the measured dose distributions. The Spearman’s rank correlation coefficients between the projected areas and the D(Rx) values were always higher than 0.75 (all with p < 0.001). The doses by the air-electron-streams increased with the projected areas of the cross-sections of the treatment beams on the panels.
format Online
Article
Text
id pubmed-6519819
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-65198192019-05-31 Electron streams in air during magnetic-resonance image-guided radiation therapy An, Hyun Joon Kim, Jung-in Park, Jong Min PLoS One Research Article To investigate the undesired irradiations outside of the treatment field by electron streams in air (air-electron-stream) during magnetic-resonance image-guided radiation therapy (MR-IGRT). A custom-made support phantom adjusting angles between the beam central axis (CAX) and the phantom surface (termed phantom-angles), were used. Using the ViewRay system, a rectangular parallelepiped phantom placed on the support phantom, was irradiated with field sizes of 6.3 cm × 6.3 cm (FS6.3) and 12.6 cm × 12.6 cm (FS12.6) at gantry angles of 0°, 30°, and 330°, and phantom-angles of 10°, 20°, and 30°. For each beam delivery, the isocenter was located at the center of mass of the phantom and 3 Gy was delivered to the isocenter (prescription dose = 3 Gy). The doses given by the air-electron-streams were measured using the EBT3 films on the panels placed orthogonal to the direction of the magnetic field at distances of 10 and 17 cm from CAX. Two dose distributions per irradiation were measured on the panel facing the phantom surface of the incident beam (front panel) and on the panel facing the phantom surface of the beam exit (end panel). We investigated the doses by the air-electron-streams by calculating the average doses inside the circles drawn around a point of the maximum dose with radii of x cm (D(Rx)) from the dose distributions on the panels (x = 1–5 cm). The largest value of D(Rx) was D(R1) (1.64 Gy, 55% of the prescription dose) at 10 cm distance from CAX, with FS12.6, at 30° phantom-angle and 330° gantry angle. The average difference of the D(R1) at the end panels (FS12.6) between the calculations and measurements was 1.36 Gy. The average global gamma passing rate with 3%/3 mm on the dose distributions at the end panels (FS12.6) was 40.3%. The calculated dose distributions on both panels were not coincident with the measured dose distributions. The Spearman’s rank correlation coefficients between the projected areas and the D(Rx) values were always higher than 0.75 (all with p < 0.001). The doses by the air-electron-streams increased with the projected areas of the cross-sections of the treatment beams on the panels. Public Library of Science 2019-05-15 /pmc/articles/PMC6519819/ /pubmed/31091270 http://dx.doi.org/10.1371/journal.pone.0216965 Text en © 2019 An et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
An, Hyun Joon
Kim, Jung-in
Park, Jong Min
Electron streams in air during magnetic-resonance image-guided radiation therapy
title Electron streams in air during magnetic-resonance image-guided radiation therapy
title_full Electron streams in air during magnetic-resonance image-guided radiation therapy
title_fullStr Electron streams in air during magnetic-resonance image-guided radiation therapy
title_full_unstemmed Electron streams in air during magnetic-resonance image-guided radiation therapy
title_short Electron streams in air during magnetic-resonance image-guided radiation therapy
title_sort electron streams in air during magnetic-resonance image-guided radiation therapy
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6519819/
https://www.ncbi.nlm.nih.gov/pubmed/31091270
http://dx.doi.org/10.1371/journal.pone.0216965
work_keys_str_mv AT anhyunjoon electronstreamsinairduringmagneticresonanceimageguidedradiationtherapy
AT kimjungin electronstreamsinairduringmagneticresonanceimageguidedradiationtherapy
AT parkjongmin electronstreamsinairduringmagneticresonanceimageguidedradiationtherapy