Determining efficient helical IMRT modulation factor from the MLC leaf‐open time distribution on precision treatment planning system

PURPOSE: Since the modulation factor (MF) impacts both plan quality and delivery efficiency in tomotherapy Intensity Modulated Radiation Therapy (IMRT) treatment planning, the purpose of this study was to demonstrate a technique in determining an efficient MF from the Multileaf Collimator (MLC) leaf...

Descripción completa

Detalles Bibliográficos
Autores principales: Boyd, Robert, Jeong, Kyoungkeun, Tomé, Wolfgang A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6523016/
https://www.ncbi.nlm.nih.gov/pubmed/30957967
http://dx.doi.org/10.1002/acm2.12581
_version_ 1783419236148838400
author Boyd, Robert
Jeong, Kyoungkeun
Tomé, Wolfgang A.
author_facet Boyd, Robert
Jeong, Kyoungkeun
Tomé, Wolfgang A.
author_sort Boyd, Robert
collection PubMed
description PURPOSE: Since the modulation factor (MF) impacts both plan quality and delivery efficiency in tomotherapy Intensity Modulated Radiation Therapy (IMRT) treatment planning, the purpose of this study was to demonstrate a technique in determining an efficient MF from the Multileaf Collimator (MLC) leaf‐open time (LOT) distribution of a tomotherapy treatment delivery plan. METHODS: Eight clinical plans of varying complexity were optimized with the highest allowed MF on the Accuracy Precision treatment planning system. Using a central limit theorem argument a range of reduced MFs were then determined from the first two moments of the LOT distribution. A step down approach was used to calculate the reduced‐MF plans and plan comparison tools available on the Precision treatment planning system were used to evaluate dose differences with the reference plan. RESULTS: A reduced‐MF plan that balanced delivery time and dosimetric quality was found from the set of five MFs determined from the LOT distribution of the reference plan. The reduced‐MF plan showed good agreement with the reference plan (target and critical organ dose‐volume region of interest dose differences were within 1% and 2% of prescription dose, respectively). DISCUSSION: Plan evaluation and acceptance criteria can vary depending on individual clinical expectations and dosimetric quality trade‐offs. With the scheme presented in this paper a planner should be able to efficiently generate a high‐quality plan with efficient delivery time without knowing a good MF beforehand. CONCLUSION: A methodology for deriving a reduced MF from the LOT distribution of a high MF treatment plan using the central limit theorem has been presented. A scheme for finding a reduced MF from a set of MFs that results in a plan balanced in both dosimetric quality and treatment delivery efficiency has also been presented.
format Online
Article
Text
id pubmed-6523016
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-65230162019-05-24 Determining efficient helical IMRT modulation factor from the MLC leaf‐open time distribution on precision treatment planning system Boyd, Robert Jeong, Kyoungkeun Tomé, Wolfgang A. J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: Since the modulation factor (MF) impacts both plan quality and delivery efficiency in tomotherapy Intensity Modulated Radiation Therapy (IMRT) treatment planning, the purpose of this study was to demonstrate a technique in determining an efficient MF from the Multileaf Collimator (MLC) leaf‐open time (LOT) distribution of a tomotherapy treatment delivery plan. METHODS: Eight clinical plans of varying complexity were optimized with the highest allowed MF on the Accuracy Precision treatment planning system. Using a central limit theorem argument a range of reduced MFs were then determined from the first two moments of the LOT distribution. A step down approach was used to calculate the reduced‐MF plans and plan comparison tools available on the Precision treatment planning system were used to evaluate dose differences with the reference plan. RESULTS: A reduced‐MF plan that balanced delivery time and dosimetric quality was found from the set of five MFs determined from the LOT distribution of the reference plan. The reduced‐MF plan showed good agreement with the reference plan (target and critical organ dose‐volume region of interest dose differences were within 1% and 2% of prescription dose, respectively). DISCUSSION: Plan evaluation and acceptance criteria can vary depending on individual clinical expectations and dosimetric quality trade‐offs. With the scheme presented in this paper a planner should be able to efficiently generate a high‐quality plan with efficient delivery time without knowing a good MF beforehand. CONCLUSION: A methodology for deriving a reduced MF from the LOT distribution of a high MF treatment plan using the central limit theorem has been presented. A scheme for finding a reduced MF from a set of MFs that results in a plan balanced in both dosimetric quality and treatment delivery efficiency has also been presented. John Wiley and Sons Inc. 2019-04-08 /pmc/articles/PMC6523016/ /pubmed/30957967 http://dx.doi.org/10.1002/acm2.12581 Text en © 2019 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Boyd, Robert
Jeong, Kyoungkeun
Tomé, Wolfgang A.
Determining efficient helical IMRT modulation factor from the MLC leaf‐open time distribution on precision treatment planning system
title Determining efficient helical IMRT modulation factor from the MLC leaf‐open time distribution on precision treatment planning system
title_full Determining efficient helical IMRT modulation factor from the MLC leaf‐open time distribution on precision treatment planning system
title_fullStr Determining efficient helical IMRT modulation factor from the MLC leaf‐open time distribution on precision treatment planning system
title_full_unstemmed Determining efficient helical IMRT modulation factor from the MLC leaf‐open time distribution on precision treatment planning system
title_short Determining efficient helical IMRT modulation factor from the MLC leaf‐open time distribution on precision treatment planning system
title_sort determining efficient helical imrt modulation factor from the mlc leaf‐open time distribution on precision treatment planning system
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6523016/
https://www.ncbi.nlm.nih.gov/pubmed/30957967
http://dx.doi.org/10.1002/acm2.12581
work_keys_str_mv AT boydrobert determiningefficienthelicalimrtmodulationfactorfromthemlcleafopentimedistributiononprecisiontreatmentplanningsystem
AT jeongkyoungkeun determiningefficienthelicalimrtmodulationfactorfromthemlcleafopentimedistributiononprecisiontreatmentplanningsystem
AT tomewolfganga determiningefficienthelicalimrtmodulationfactorfromthemlcleafopentimedistributiononprecisiontreatmentplanningsystem