PURPOSE: To evaluate the different IOP readings by Goldmann applanation tonometer (GAT), ICare rebound tonometer, and Tono-Pen in keratoconus patients after MyoRing implantation. To assess the influence of central corneal thickness (CCT) and thinnest corneal location (TCL) on IOP measurements by dif...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6532290/ https://www.ncbi.nlm.nih.gov/pubmed/31210982 http://dx.doi.org/10.1155/2019/1964107 |
_version_ | 1783420994741862400 |
---|---|
author | Rateb, Mahmoud Abdel-Radi, Mahmoud Eldaly, Zeiad Elmohamady, Mohamed Nagy Noor El Din, Asaad |
author_facet | Rateb, Mahmoud Abdel-Radi, Mahmoud Eldaly, Zeiad Elmohamady, Mohamed Nagy Noor El Din, Asaad |
author_sort | Rateb, Mahmoud |
collection | PubMed |
description | PURPOSE: To evaluate the different IOP readings by Goldmann applanation tonometer (GAT), ICare rebound tonometer, and Tono-Pen in keratoconus patients after MyoRing implantation. To assess the influence of central corneal thickness (CCT) and thinnest corneal location (TCL) on IOP measurements by different tonometers. SETTING: Prospective observational study was conducted in two private centers in Egypt from February 2015 to November 2016. METHODS: Seventeen eyes of 10 patients suffering from keratoconus and who underwent MyoRing implantation were recruited. All subjects underwent GAT, ICare, and Tono-Pen IOP measurements in random order. Central corneal thickness and thinnest corneal location were assessed by Pentacam. Difference in mean in IOP readings was assessed by T-test. Correlation between each pair of devices was evaluated by Pearson correlation coefficient. The Bland–Altman analysis was used to assess intertonometer agreement. RESULTS: Seventeen eyes (10 patients) were evaluated. The mean IOP reading was 13.9 ± 3.68, 12.41 ± 2.87, and 14.29 ± 1.31 mmHg in GAT, ICare, and Tono-Pen group, respectively. There was a significant difference between IOP readings by GAT/ICare and Tono-Pen/ICare (p value: 0.032 and 0.002, respectively) with no significant difference between GAT/Tono-Pen (p value: 0.554). Mean difference in IOP measurements between GAT/ICare was 1.49 ± 2.61 mmHg, Tono-Pen/ICare was 1.89 ± 2.15 mmHg, and GAT/Tono-Pen was −0.39 ± 2.59 mmHg. There was no significant correlation between the difference in IOP readings among any pair of devices and CCC or TCL. The Bland–Altman analysis showed a reasonable agreement between any pair of tonometers. |
format | Online Article Text |
id | pubmed-6532290 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Hindawi |
record_format | MEDLINE/PubMed |
spelling | pubmed-65322902019-06-17 Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation Rateb, Mahmoud Abdel-Radi, Mahmoud Eldaly, Zeiad Elmohamady, Mohamed Nagy Noor El Din, Asaad J Ophthalmol Research Article PURPOSE: To evaluate the different IOP readings by Goldmann applanation tonometer (GAT), ICare rebound tonometer, and Tono-Pen in keratoconus patients after MyoRing implantation. To assess the influence of central corneal thickness (CCT) and thinnest corneal location (TCL) on IOP measurements by different tonometers. SETTING: Prospective observational study was conducted in two private centers in Egypt from February 2015 to November 2016. METHODS: Seventeen eyes of 10 patients suffering from keratoconus and who underwent MyoRing implantation were recruited. All subjects underwent GAT, ICare, and Tono-Pen IOP measurements in random order. Central corneal thickness and thinnest corneal location were assessed by Pentacam. Difference in mean in IOP readings was assessed by T-test. Correlation between each pair of devices was evaluated by Pearson correlation coefficient. The Bland–Altman analysis was used to assess intertonometer agreement. RESULTS: Seventeen eyes (10 patients) were evaluated. The mean IOP reading was 13.9 ± 3.68, 12.41 ± 2.87, and 14.29 ± 1.31 mmHg in GAT, ICare, and Tono-Pen group, respectively. There was a significant difference between IOP readings by GAT/ICare and Tono-Pen/ICare (p value: 0.032 and 0.002, respectively) with no significant difference between GAT/Tono-Pen (p value: 0.554). Mean difference in IOP measurements between GAT/ICare was 1.49 ± 2.61 mmHg, Tono-Pen/ICare was 1.89 ± 2.15 mmHg, and GAT/Tono-Pen was −0.39 ± 2.59 mmHg. There was no significant correlation between the difference in IOP readings among any pair of devices and CCC or TCL. The Bland–Altman analysis showed a reasonable agreement between any pair of tonometers. Hindawi 2019-05-09 /pmc/articles/PMC6532290/ /pubmed/31210982 http://dx.doi.org/10.1155/2019/1964107 Text en Copyright © 2019 Mahmoud Rateb et al. http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Rateb, Mahmoud Abdel-Radi, Mahmoud Eldaly, Zeiad Elmohamady, Mohamed Nagy Noor El Din, Asaad Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation |
title | Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation |
title_full | Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation |
title_fullStr | Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation |
title_full_unstemmed | Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation |
title_short | Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation |
title_sort | comparison of iop measurement by goldmann applanation tonometer, icare rebound tonometer, and tono-pen in keratoconus patients after myoring implantation |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6532290/ https://www.ncbi.nlm.nih.gov/pubmed/31210982 http://dx.doi.org/10.1155/2019/1964107 |
work_keys_str_mv | AT ratebmahmoud comparisonofiopmeasurementbygoldmannapplanationtonometericarereboundtonometerandtonopeninkeratoconuspatientsaftermyoringimplantation AT abdelradimahmoud comparisonofiopmeasurementbygoldmannapplanationtonometericarereboundtonometerandtonopeninkeratoconuspatientsaftermyoringimplantation AT eldalyzeiad comparisonofiopmeasurementbygoldmannapplanationtonometericarereboundtonometerandtonopeninkeratoconuspatientsaftermyoringimplantation AT elmohamadymohamednagy comparisonofiopmeasurementbygoldmannapplanationtonometericarereboundtonometerandtonopeninkeratoconuspatientsaftermyoringimplantation AT nooreldinasaad comparisonofiopmeasurementbygoldmannapplanationtonometericarereboundtonometerandtonopeninkeratoconuspatientsaftermyoringimplantation |