Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation

PURPOSE: To evaluate the different IOP readings by Goldmann applanation tonometer (GAT), ICare rebound tonometer, and Tono-Pen in keratoconus patients after MyoRing implantation. To assess the influence of central corneal thickness (CCT) and thinnest corneal location (TCL) on IOP measurements by dif...

Descripción completa

Detalles Bibliográficos
Autores principales: Rateb, Mahmoud, Abdel-Radi, Mahmoud, Eldaly, Zeiad, Elmohamady, Mohamed Nagy, Noor El Din, Asaad
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6532290/
https://www.ncbi.nlm.nih.gov/pubmed/31210982
http://dx.doi.org/10.1155/2019/1964107
_version_ 1783420994741862400
author Rateb, Mahmoud
Abdel-Radi, Mahmoud
Eldaly, Zeiad
Elmohamady, Mohamed Nagy
Noor El Din, Asaad
author_facet Rateb, Mahmoud
Abdel-Radi, Mahmoud
Eldaly, Zeiad
Elmohamady, Mohamed Nagy
Noor El Din, Asaad
author_sort Rateb, Mahmoud
collection PubMed
description PURPOSE: To evaluate the different IOP readings by Goldmann applanation tonometer (GAT), ICare rebound tonometer, and Tono-Pen in keratoconus patients after MyoRing implantation. To assess the influence of central corneal thickness (CCT) and thinnest corneal location (TCL) on IOP measurements by different tonometers. SETTING: Prospective observational study was conducted in two private centers in Egypt from February 2015 to November 2016. METHODS: Seventeen eyes of 10 patients suffering from keratoconus and who underwent MyoRing implantation were recruited. All subjects underwent GAT, ICare, and Tono-Pen IOP measurements in random order. Central corneal thickness and thinnest corneal location were assessed by Pentacam. Difference in mean in IOP readings was assessed by T-test. Correlation between each pair of devices was evaluated by Pearson correlation coefficient. The Bland–Altman analysis was used to assess intertonometer agreement. RESULTS: Seventeen eyes (10 patients) were evaluated. The mean IOP reading was 13.9 ± 3.68, 12.41 ± 2.87, and 14.29 ± 1.31 mmHg in GAT, ICare, and Tono-Pen group, respectively. There was a significant difference between IOP readings by GAT/ICare and Tono-Pen/ICare (p value: 0.032 and 0.002, respectively) with no significant difference between GAT/Tono-Pen (p value: 0.554). Mean difference in IOP measurements between GAT/ICare was 1.49 ± 2.61 mmHg, Tono-Pen/ICare was 1.89 ± 2.15 mmHg, and GAT/Tono-Pen was −0.39 ± 2.59 mmHg. There was no significant correlation between the difference in IOP readings among any pair of devices and CCC or TCL. The Bland–Altman analysis showed a reasonable agreement between any pair of tonometers.
format Online
Article
Text
id pubmed-6532290
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Hindawi
record_format MEDLINE/PubMed
spelling pubmed-65322902019-06-17 Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation Rateb, Mahmoud Abdel-Radi, Mahmoud Eldaly, Zeiad Elmohamady, Mohamed Nagy Noor El Din, Asaad J Ophthalmol Research Article PURPOSE: To evaluate the different IOP readings by Goldmann applanation tonometer (GAT), ICare rebound tonometer, and Tono-Pen in keratoconus patients after MyoRing implantation. To assess the influence of central corneal thickness (CCT) and thinnest corneal location (TCL) on IOP measurements by different tonometers. SETTING: Prospective observational study was conducted in two private centers in Egypt from February 2015 to November 2016. METHODS: Seventeen eyes of 10 patients suffering from keratoconus and who underwent MyoRing implantation were recruited. All subjects underwent GAT, ICare, and Tono-Pen IOP measurements in random order. Central corneal thickness and thinnest corneal location were assessed by Pentacam. Difference in mean in IOP readings was assessed by T-test. Correlation between each pair of devices was evaluated by Pearson correlation coefficient. The Bland–Altman analysis was used to assess intertonometer agreement. RESULTS: Seventeen eyes (10 patients) were evaluated. The mean IOP reading was 13.9 ± 3.68, 12.41 ± 2.87, and 14.29 ± 1.31 mmHg in GAT, ICare, and Tono-Pen group, respectively. There was a significant difference between IOP readings by GAT/ICare and Tono-Pen/ICare (p value: 0.032 and 0.002, respectively) with no significant difference between GAT/Tono-Pen (p value: 0.554). Mean difference in IOP measurements between GAT/ICare was 1.49 ± 2.61 mmHg, Tono-Pen/ICare was 1.89 ± 2.15 mmHg, and GAT/Tono-Pen was −0.39 ± 2.59 mmHg. There was no significant correlation between the difference in IOP readings among any pair of devices and CCC or TCL. The Bland–Altman analysis showed a reasonable agreement between any pair of tonometers. Hindawi 2019-05-09 /pmc/articles/PMC6532290/ /pubmed/31210982 http://dx.doi.org/10.1155/2019/1964107 Text en Copyright © 2019 Mahmoud Rateb et al. http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Rateb, Mahmoud
Abdel-Radi, Mahmoud
Eldaly, Zeiad
Elmohamady, Mohamed Nagy
Noor El Din, Asaad
Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation
title Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation
title_full Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation
title_fullStr Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation
title_full_unstemmed Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation
title_short Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation
title_sort comparison of iop measurement by goldmann applanation tonometer, icare rebound tonometer, and tono-pen in keratoconus patients after myoring implantation
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6532290/
https://www.ncbi.nlm.nih.gov/pubmed/31210982
http://dx.doi.org/10.1155/2019/1964107
work_keys_str_mv AT ratebmahmoud comparisonofiopmeasurementbygoldmannapplanationtonometericarereboundtonometerandtonopeninkeratoconuspatientsaftermyoringimplantation
AT abdelradimahmoud comparisonofiopmeasurementbygoldmannapplanationtonometericarereboundtonometerandtonopeninkeratoconuspatientsaftermyoringimplantation
AT eldalyzeiad comparisonofiopmeasurementbygoldmannapplanationtonometericarereboundtonometerandtonopeninkeratoconuspatientsaftermyoringimplantation
AT elmohamadymohamednagy comparisonofiopmeasurementbygoldmannapplanationtonometericarereboundtonometerandtonopeninkeratoconuspatientsaftermyoringimplantation
AT nooreldinasaad comparisonofiopmeasurementbygoldmannapplanationtonometericarereboundtonometerandtonopeninkeratoconuspatientsaftermyoringimplantation