Precontoured buttress plate vs reconstruction plate for acetabulum posterior wall fractures: A biomechanical study

BACKGROUND: The purpose of open reduction and internal fixation of acetabulum posterior wall fractures is to restore anatomical structure and stability of the hip joint, in order to start weight bearing as soon as possible and prevent hip arthrosis; restoration of the anatomy should preserve functio...

Descripción completa

Detalles Bibliográficos
Autores principales: Altun, Güray, Saka, Gürsel, Demir, Teyfik, Elibol, Fatma Kübra Erbay, Polat, Mehmet Orçun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Baishideng Publishing Group Inc 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6533420/
https://www.ncbi.nlm.nih.gov/pubmed/31149562
http://dx.doi.org/10.5312/wjo.v10.i5.219
_version_ 1783421205130248192
author Altun, Güray
Saka, Gürsel
Demir, Teyfik
Elibol, Fatma Kübra Erbay
Polat, Mehmet Orçun
author_facet Altun, Güray
Saka, Gürsel
Demir, Teyfik
Elibol, Fatma Kübra Erbay
Polat, Mehmet Orçun
author_sort Altun, Güray
collection PubMed
description BACKGROUND: The purpose of open reduction and internal fixation of acetabulum posterior wall fractures is to restore anatomical structure and stability of the hip joint, in order to start weight bearing as soon as possible and prevent hip arthrosis; restoration of the anatomy should preserve function of the joint as well. Although “special shaped precontoured plates” have been developed in recent years for surgical treatment of this region, studies comparing the traditional plates with the newly designed precontoured plates are lacking. AIM: To evaluate the biomechanical properties of precontoured anatomic buttress and conventional curved reconstruction plates (CCRPs) for posterior wall acetabulum fracture treatment. METHODS: Twelve pelvis models were created for testing plate treatment of fracture in the posterior wall of the acetabulum. These 12 pelvis models were used to create 24 hemipelvis models (experimental) by cutting from the sagittal plane and passing over the center of gravity, after which the posterior wall acetabular fractures (of similar type and size) were created. In these experimental models, the right acetabulum was fixed with a 5-hole CCRP, while the left was fixed with a precontoured anatomic buttress plate (PABP). Samples were placed through the test device and were subjected to static load testing, with a constant testing velocity of 2 mm/min until the load reached 2.3 kN or the acetabular fixation failed. Dynamic tests were also performed with sinusoidal wave load, with a maximal load of 2.3 kN and a load ratio of 0.1. RESULTS: The average stiffness values were 460.83 ± 95.47 N/mm for the PABP and 291.99 ± 118.58 N/mm for the 5-hole CCRP. The precontoured anatomic acetabulum buttress plates had significantly higher rigidity than the CCRPs (P = 0.022). There was a statistically significant difference between the unloaded and 2.3 kN-loaded values of AL (posterosuperior fracture line vertical to the ground surface) and CL (posteroinferior fracture line vertical to the ground surface) parameters for both the PABPs and the 5-hole CCRPs (P = 0.036 and P = 0.045, respectively). According to the static tests, the amount of total displacement was significantly less in the PABPs than in the CCRPs. Comparative analysis of the displacement in the BL (posterior wall fracture line horizontal to the ground) parameter yielded no statistically significant differences between the PABP and the 5-hole CCRPs (P = 0.261). CONCLUSION: PABP provides more stable fixation in acetabulum posterior wall fractures than 5-hole CCRP, allowing for proximal or distal fracture line screw application without reshaping.
format Online
Article
Text
id pubmed-6533420
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Baishideng Publishing Group Inc
record_format MEDLINE/PubMed
spelling pubmed-65334202019-05-30 Precontoured buttress plate vs reconstruction plate for acetabulum posterior wall fractures: A biomechanical study Altun, Güray Saka, Gürsel Demir, Teyfik Elibol, Fatma Kübra Erbay Polat, Mehmet Orçun World J Orthop Basic Study BACKGROUND: The purpose of open reduction and internal fixation of acetabulum posterior wall fractures is to restore anatomical structure and stability of the hip joint, in order to start weight bearing as soon as possible and prevent hip arthrosis; restoration of the anatomy should preserve function of the joint as well. Although “special shaped precontoured plates” have been developed in recent years for surgical treatment of this region, studies comparing the traditional plates with the newly designed precontoured plates are lacking. AIM: To evaluate the biomechanical properties of precontoured anatomic buttress and conventional curved reconstruction plates (CCRPs) for posterior wall acetabulum fracture treatment. METHODS: Twelve pelvis models were created for testing plate treatment of fracture in the posterior wall of the acetabulum. These 12 pelvis models were used to create 24 hemipelvis models (experimental) by cutting from the sagittal plane and passing over the center of gravity, after which the posterior wall acetabular fractures (of similar type and size) were created. In these experimental models, the right acetabulum was fixed with a 5-hole CCRP, while the left was fixed with a precontoured anatomic buttress plate (PABP). Samples were placed through the test device and were subjected to static load testing, with a constant testing velocity of 2 mm/min until the load reached 2.3 kN or the acetabular fixation failed. Dynamic tests were also performed with sinusoidal wave load, with a maximal load of 2.3 kN and a load ratio of 0.1. RESULTS: The average stiffness values were 460.83 ± 95.47 N/mm for the PABP and 291.99 ± 118.58 N/mm for the 5-hole CCRP. The precontoured anatomic acetabulum buttress plates had significantly higher rigidity than the CCRPs (P = 0.022). There was a statistically significant difference between the unloaded and 2.3 kN-loaded values of AL (posterosuperior fracture line vertical to the ground surface) and CL (posteroinferior fracture line vertical to the ground surface) parameters for both the PABPs and the 5-hole CCRPs (P = 0.036 and P = 0.045, respectively). According to the static tests, the amount of total displacement was significantly less in the PABPs than in the CCRPs. Comparative analysis of the displacement in the BL (posterior wall fracture line horizontal to the ground) parameter yielded no statistically significant differences between the PABP and the 5-hole CCRPs (P = 0.261). CONCLUSION: PABP provides more stable fixation in acetabulum posterior wall fractures than 5-hole CCRP, allowing for proximal or distal fracture line screw application without reshaping. Baishideng Publishing Group Inc 2019-05-18 /pmc/articles/PMC6533420/ /pubmed/31149562 http://dx.doi.org/10.5312/wjo.v10.i5.219 Text en ©The Author(s) 2019. Published by Baishideng Publishing Group Inc. All rights reserved. http://creativecommons.org/licenses/by-nc/4.0/ This article is an open-access article which was selected by an in-house editor and fully peer-reviewed by external reviewers. It is distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial.
spellingShingle Basic Study
Altun, Güray
Saka, Gürsel
Demir, Teyfik
Elibol, Fatma Kübra Erbay
Polat, Mehmet Orçun
Precontoured buttress plate vs reconstruction plate for acetabulum posterior wall fractures: A biomechanical study
title Precontoured buttress plate vs reconstruction plate for acetabulum posterior wall fractures: A biomechanical study
title_full Precontoured buttress plate vs reconstruction plate for acetabulum posterior wall fractures: A biomechanical study
title_fullStr Precontoured buttress plate vs reconstruction plate for acetabulum posterior wall fractures: A biomechanical study
title_full_unstemmed Precontoured buttress plate vs reconstruction plate for acetabulum posterior wall fractures: A biomechanical study
title_short Precontoured buttress plate vs reconstruction plate for acetabulum posterior wall fractures: A biomechanical study
title_sort precontoured buttress plate vs reconstruction plate for acetabulum posterior wall fractures: a biomechanical study
topic Basic Study
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6533420/
https://www.ncbi.nlm.nih.gov/pubmed/31149562
http://dx.doi.org/10.5312/wjo.v10.i5.219
work_keys_str_mv AT altunguray precontouredbuttressplatevsreconstructionplateforacetabulumposteriorwallfracturesabiomechanicalstudy
AT sakagursel precontouredbuttressplatevsreconstructionplateforacetabulumposteriorwallfracturesabiomechanicalstudy
AT demirteyfik precontouredbuttressplatevsreconstructionplateforacetabulumposteriorwallfracturesabiomechanicalstudy
AT elibolfatmakubraerbay precontouredbuttressplatevsreconstructionplateforacetabulumposteriorwallfracturesabiomechanicalstudy
AT polatmehmetorcun precontouredbuttressplatevsreconstructionplateforacetabulumposteriorwallfracturesabiomechanicalstudy