Compare the accuracy and precision of Coulter LH780, Mindray BC-6000 Plus, and Sysmex XN-9000 with the international reference flow cytometric method in platelet counting

OBJECTIVE: The aim of this study is to evaluate the performance of different platelet counting methods (optical, impedance, fluorescence and hand counting) applied in different analysers by comparing with the international flow cytometric reference method (IRM). METHODS: A total of 333 blood samples...

Descripción completa

Detalles Bibliográficos
Autores principales: Sun, Yi, Hu, Zuojian, Huang, Zhili, Chen, Huaping, Qin, Shanzi, Jianing, Zhong, Chen, Siyuan, Qin, Xue, Ye, Yi, Wang, Chengbin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6534315/
https://www.ncbi.nlm.nih.gov/pubmed/31125378
http://dx.doi.org/10.1371/journal.pone.0217298
_version_ 1783421389077741568
author Sun, Yi
Hu, Zuojian
Huang, Zhili
Chen, Huaping
Qin, Shanzi
Jianing, Zhong
Chen, Siyuan
Qin, Xue
Ye, Yi
Wang, Chengbin
author_facet Sun, Yi
Hu, Zuojian
Huang, Zhili
Chen, Huaping
Qin, Shanzi
Jianing, Zhong
Chen, Siyuan
Qin, Xue
Ye, Yi
Wang, Chengbin
author_sort Sun, Yi
collection PubMed
description OBJECTIVE: The aim of this study is to evaluate the performance of different platelet counting methods (optical, impedance, fluorescence and hand counting) applied in different analysers by comparing with the international flow cytometric reference method (IRM). METHODS: A total of 333 blood samples from different subgroups (168 cases with thrombocytopenia, 136 cases with normal platelet counts and 29 cases with thrombocytosis) were tested. Regarding IRM as the gold standard, we compared the accuracy and precision of different platelet count methods; i.e. LH780 (impedance), BC-6000 Plus (optical (O) and impedance (I)), Sysmex XN-9000 (optical (O), impedance (I), fluorescence (F)), and hand counting. RESULTS: Sysmex XN-9000-F (r = 0.988) had the best correlation with IRM for thrombocytopenic samples; BC-6000 Plus-I (r = 0.966) was more relevant to IRM than any other method for samples with normal platelet counts. Correlation between Sysmex XN-9000-I (r = 0.960) and IRM was the highest among these methods for samples with thrombocytosis. For bias evaluation, the average bias of Sysmex XN-9000-F was -1.5 × 10(9)/L (95% LA = -9.4 to +6.4) for samples with thrombocytopenia, compared with IRM. BC-6000 Plus-I had a small mean difference with IRM for samples with normal platelet counts or thrombocytosis. Moreover, all evaluated methods had acceptable sensitivity, specificity, and concordance rates as compared with IRM in the diagnosis of thrombocytopenia and thrombocytosis. CONCLUSIONS: Platelet counting by Sysmex XN-9000-F is more accurate than other methods for thrombocytopenic samples. BC-6000 Plus-I has superior association and consistency for normal platelet counts. As for thrombocytosis patients, Sysmex XN-9000-I has the highest correlation with IRM while Sysmex XN-9000-O has the highest diagnosis efficacy.
format Online
Article
Text
id pubmed-6534315
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-65343152019-06-05 Compare the accuracy and precision of Coulter LH780, Mindray BC-6000 Plus, and Sysmex XN-9000 with the international reference flow cytometric method in platelet counting Sun, Yi Hu, Zuojian Huang, Zhili Chen, Huaping Qin, Shanzi Jianing, Zhong Chen, Siyuan Qin, Xue Ye, Yi Wang, Chengbin PLoS One Research Article OBJECTIVE: The aim of this study is to evaluate the performance of different platelet counting methods (optical, impedance, fluorescence and hand counting) applied in different analysers by comparing with the international flow cytometric reference method (IRM). METHODS: A total of 333 blood samples from different subgroups (168 cases with thrombocytopenia, 136 cases with normal platelet counts and 29 cases with thrombocytosis) were tested. Regarding IRM as the gold standard, we compared the accuracy and precision of different platelet count methods; i.e. LH780 (impedance), BC-6000 Plus (optical (O) and impedance (I)), Sysmex XN-9000 (optical (O), impedance (I), fluorescence (F)), and hand counting. RESULTS: Sysmex XN-9000-F (r = 0.988) had the best correlation with IRM for thrombocytopenic samples; BC-6000 Plus-I (r = 0.966) was more relevant to IRM than any other method for samples with normal platelet counts. Correlation between Sysmex XN-9000-I (r = 0.960) and IRM was the highest among these methods for samples with thrombocytosis. For bias evaluation, the average bias of Sysmex XN-9000-F was -1.5 × 10(9)/L (95% LA = -9.4 to +6.4) for samples with thrombocytopenia, compared with IRM. BC-6000 Plus-I had a small mean difference with IRM for samples with normal platelet counts or thrombocytosis. Moreover, all evaluated methods had acceptable sensitivity, specificity, and concordance rates as compared with IRM in the diagnosis of thrombocytopenia and thrombocytosis. CONCLUSIONS: Platelet counting by Sysmex XN-9000-F is more accurate than other methods for thrombocytopenic samples. BC-6000 Plus-I has superior association and consistency for normal platelet counts. As for thrombocytosis patients, Sysmex XN-9000-I has the highest correlation with IRM while Sysmex XN-9000-O has the highest diagnosis efficacy. Public Library of Science 2019-05-24 /pmc/articles/PMC6534315/ /pubmed/31125378 http://dx.doi.org/10.1371/journal.pone.0217298 Text en © 2019 Sun et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Sun, Yi
Hu, Zuojian
Huang, Zhili
Chen, Huaping
Qin, Shanzi
Jianing, Zhong
Chen, Siyuan
Qin, Xue
Ye, Yi
Wang, Chengbin
Compare the accuracy and precision of Coulter LH780, Mindray BC-6000 Plus, and Sysmex XN-9000 with the international reference flow cytometric method in platelet counting
title Compare the accuracy and precision of Coulter LH780, Mindray BC-6000 Plus, and Sysmex XN-9000 with the international reference flow cytometric method in platelet counting
title_full Compare the accuracy and precision of Coulter LH780, Mindray BC-6000 Plus, and Sysmex XN-9000 with the international reference flow cytometric method in platelet counting
title_fullStr Compare the accuracy and precision of Coulter LH780, Mindray BC-6000 Plus, and Sysmex XN-9000 with the international reference flow cytometric method in platelet counting
title_full_unstemmed Compare the accuracy and precision of Coulter LH780, Mindray BC-6000 Plus, and Sysmex XN-9000 with the international reference flow cytometric method in platelet counting
title_short Compare the accuracy and precision of Coulter LH780, Mindray BC-6000 Plus, and Sysmex XN-9000 with the international reference flow cytometric method in platelet counting
title_sort compare the accuracy and precision of coulter lh780, mindray bc-6000 plus, and sysmex xn-9000 with the international reference flow cytometric method in platelet counting
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6534315/
https://www.ncbi.nlm.nih.gov/pubmed/31125378
http://dx.doi.org/10.1371/journal.pone.0217298
work_keys_str_mv AT sunyi comparetheaccuracyandprecisionofcoulterlh780mindraybc6000plusandsysmexxn9000withtheinternationalreferenceflowcytometricmethodinplateletcounting
AT huzuojian comparetheaccuracyandprecisionofcoulterlh780mindraybc6000plusandsysmexxn9000withtheinternationalreferenceflowcytometricmethodinplateletcounting
AT huangzhili comparetheaccuracyandprecisionofcoulterlh780mindraybc6000plusandsysmexxn9000withtheinternationalreferenceflowcytometricmethodinplateletcounting
AT chenhuaping comparetheaccuracyandprecisionofcoulterlh780mindraybc6000plusandsysmexxn9000withtheinternationalreferenceflowcytometricmethodinplateletcounting
AT qinshanzi comparetheaccuracyandprecisionofcoulterlh780mindraybc6000plusandsysmexxn9000withtheinternationalreferenceflowcytometricmethodinplateletcounting
AT jianingzhong comparetheaccuracyandprecisionofcoulterlh780mindraybc6000plusandsysmexxn9000withtheinternationalreferenceflowcytometricmethodinplateletcounting
AT chensiyuan comparetheaccuracyandprecisionofcoulterlh780mindraybc6000plusandsysmexxn9000withtheinternationalreferenceflowcytometricmethodinplateletcounting
AT qinxue comparetheaccuracyandprecisionofcoulterlh780mindraybc6000plusandsysmexxn9000withtheinternationalreferenceflowcytometricmethodinplateletcounting
AT yeyi comparetheaccuracyandprecisionofcoulterlh780mindraybc6000plusandsysmexxn9000withtheinternationalreferenceflowcytometricmethodinplateletcounting
AT wangchengbin comparetheaccuracyandprecisionofcoulterlh780mindraybc6000plusandsysmexxn9000withtheinternationalreferenceflowcytometricmethodinplateletcounting