Cargando…

Do the Manual or Computer-Controlled Flowmeters Generate Similar Isoflurane Concentrations in Tafonius?

Introduction: Tafonius is an anesthesia machine with computer-controlled monitor and ventilator. We compared the isoflurane fluctuations in the circuit with manual (MF) or computer-driven (CF) flowmeters, investigated the origin of the differences and assessed whether isoflurane concentration time c...

Descripción completa

Detalles Bibliográficos
Autores principales: Raillard, Mathieu, Levionnois, Olivier, MacFarlane, Paul
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6549471/
https://www.ncbi.nlm.nih.gov/pubmed/31192237
http://dx.doi.org/10.3389/fvets.2019.00160
Descripción
Sumario:Introduction: Tafonius is an anesthesia machine with computer-controlled monitor and ventilator. We compared the isoflurane fluctuations in the circuit with manual (MF) or computer-driven (CF) flowmeters, investigated the origin of the differences and assessed whether isoflurane concentration time course followed a one-compartment model. Material and Methods: A calibrated TEC-3 isoflurane vaporizer was used. Gas composition and flows were measured using a multiparametric monitor and a digital flowmeter. Measurements included: (1) Effects of various F(i)O(2) with MF/CF on the isoflurane fraction changes in the breathing system during mechanical ventilation of a lung model; wash-in kinetic was fitted to a compartmental model; (2) Gas outflow at the common gas outlet (CGO) with MF/CF at different F(i)O(2); (3) Isoflurane output of the vaporizer at various dial settings with MF/CF set at different flows without and with reduction of the CGO diameter. Results: (1) The 3% targeted isoflurane concentration was not reached; additional time was required to reach specific concentrations with CF (lowest F(i)O(2), longer time). The exponential course fitted a two-compartment model; (2) Set and measured flows were identical with MF. With CF at 0.21 F(i)O(2), flow was intermittently 7.6 L min(−1) or zero (mean total: 38% of the set flow); with CF at 1.00 F(i)O(2), flow was 10.6 L min(−1) or zero (mean: 4–5.3 L min(−1)); with 0.21 < F(i)O(2) < 1.00, combined flow was intermittent (maximum output: 15.6 L min(−1)); (3) With MF, isoflurane output was matching dial setting at 5 L min(−1) but was lower at higher flows; with CF generating intermittent flows, isoflurane output was fluctuating. With the 4 mm diameter CGO, isoflurane concentration was close to dial setting with both MF and CF. With a 14 G CGO, isoflurane concentration was lower than dial setting with MF, higher with CF. Conclusions and Clinical Relevance: Using MF or CF led to different isoflurane fraction time course in Tafonius. Flows were lower than set with CF; the TEC-3 did not compensate for high/intermittent flows and pressures; the CGO diameter influenced isoflurane output.