Analytical evaluation of the new Seal Autoanalyzer 3 High Resolution for urinary iodine determination

INTRODUCTION: The aim of the study was to evaluate the analytical performance of the new colorimetric, automatic analyser, Seal AutoAnalyzer 3 High Resolution (Seal AA3 HR) (Seal Analytical, Wisconsin, USA) for urinary iodine measurement. MATERIALS AND METHODS: This study included testing of several...

Descripción completa

Detalles Bibliográficos
Autores principales: Vidranski, Valentina, Franceschi, Maja, Krilić, Dražena, Jukić, Tomislav, Mihaljević, Ivan, Kusić, Zvonko
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Croatian Society of Medical Biochemistry and Laboratory Medicine 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6559621/
https://www.ncbi.nlm.nih.gov/pubmed/31223253
http://dx.doi.org/10.11613/BM.2019.020711
_version_ 1783425830618136576
author Vidranski, Valentina
Franceschi, Maja
Krilić, Dražena
Jukić, Tomislav
Mihaljević, Ivan
Kusić, Zvonko
author_facet Vidranski, Valentina
Franceschi, Maja
Krilić, Dražena
Jukić, Tomislav
Mihaljević, Ivan
Kusić, Zvonko
author_sort Vidranski, Valentina
collection PubMed
description INTRODUCTION: The aim of the study was to evaluate the analytical performance of the new colorimetric, automatic analyser, Seal AutoAnalyzer 3 High Resolution (Seal AA3 HR) (Seal Analytical, Wisconsin, USA) for urinary iodine measurement. MATERIALS AND METHODS: This study included testing of several analytical features of the method involving: imprecision (within-run %CVr, between-run %CVb and total laboratory precision %CVl), measurement uncertainty, carryover, linearity and method comparison, with 70 urine samples including the measuring range (20 - 700 µg/L). RESULTS: Within-run, %CVb and %CVl of two control levels were 2.03% and 3.04%, 0.51% and 2.61%, and 2.09% and 4.01%, respectively. Carryover effect was less than 1%. The linearity was good in the range of urinary iodine values between 60 and 500 µg/L (R(2) = 0.99). Good agreement of urinary iodine values was found between manual technique and Seal AA3 HR, using Passing-Bablok regression (y = 7.84 (- 3.00 to 15.29) + 0.95 (0.90 to 1.00) x) and Blant-Altman test. Cusum test for linearity indicates that there is no significant deviation from linearity (P > 0.1). CONCLUSIONS: The obtained results proved excellent precision, reproducibility and linearity, comparable to the already used, manual method. The New Seal AA3 HR automatic analyser is acceptable for urinary iodine measurement with very good analytical characteristics and can be used for urinary iodine epidemiological studies of the Croatian population.
format Online
Article
Text
id pubmed-6559621
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Croatian Society of Medical Biochemistry and Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-65596212019-06-20 Analytical evaluation of the new Seal Autoanalyzer 3 High Resolution for urinary iodine determination Vidranski, Valentina Franceschi, Maja Krilić, Dražena Jukić, Tomislav Mihaljević, Ivan Kusić, Zvonko Biochem Med (Zagreb) Original Articles INTRODUCTION: The aim of the study was to evaluate the analytical performance of the new colorimetric, automatic analyser, Seal AutoAnalyzer 3 High Resolution (Seal AA3 HR) (Seal Analytical, Wisconsin, USA) for urinary iodine measurement. MATERIALS AND METHODS: This study included testing of several analytical features of the method involving: imprecision (within-run %CVr, between-run %CVb and total laboratory precision %CVl), measurement uncertainty, carryover, linearity and method comparison, with 70 urine samples including the measuring range (20 - 700 µg/L). RESULTS: Within-run, %CVb and %CVl of two control levels were 2.03% and 3.04%, 0.51% and 2.61%, and 2.09% and 4.01%, respectively. Carryover effect was less than 1%. The linearity was good in the range of urinary iodine values between 60 and 500 µg/L (R(2) = 0.99). Good agreement of urinary iodine values was found between manual technique and Seal AA3 HR, using Passing-Bablok regression (y = 7.84 (- 3.00 to 15.29) + 0.95 (0.90 to 1.00) x) and Blant-Altman test. Cusum test for linearity indicates that there is no significant deviation from linearity (P > 0.1). CONCLUSIONS: The obtained results proved excellent precision, reproducibility and linearity, comparable to the already used, manual method. The New Seal AA3 HR automatic analyser is acceptable for urinary iodine measurement with very good analytical characteristics and can be used for urinary iodine epidemiological studies of the Croatian population. Croatian Society of Medical Biochemistry and Laboratory Medicine 2019-06-15 2019-06-15 /pmc/articles/PMC6559621/ /pubmed/31223253 http://dx.doi.org/10.11613/BM.2019.020711 Text en ©Croatian Society of Medical Biochemistry and Laboratory Medicine. This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Articles
Vidranski, Valentina
Franceschi, Maja
Krilić, Dražena
Jukić, Tomislav
Mihaljević, Ivan
Kusić, Zvonko
Analytical evaluation of the new Seal Autoanalyzer 3 High Resolution for urinary iodine determination
title Analytical evaluation of the new Seal Autoanalyzer 3 High Resolution for urinary iodine determination
title_full Analytical evaluation of the new Seal Autoanalyzer 3 High Resolution for urinary iodine determination
title_fullStr Analytical evaluation of the new Seal Autoanalyzer 3 High Resolution for urinary iodine determination
title_full_unstemmed Analytical evaluation of the new Seal Autoanalyzer 3 High Resolution for urinary iodine determination
title_short Analytical evaluation of the new Seal Autoanalyzer 3 High Resolution for urinary iodine determination
title_sort analytical evaluation of the new seal autoanalyzer 3 high resolution for urinary iodine determination
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6559621/
https://www.ncbi.nlm.nih.gov/pubmed/31223253
http://dx.doi.org/10.11613/BM.2019.020711
work_keys_str_mv AT vidranskivalentina analyticalevaluationofthenewsealautoanalyzer3highresolutionforurinaryiodinedetermination
AT franceschimaja analyticalevaluationofthenewsealautoanalyzer3highresolutionforurinaryiodinedetermination
AT krilicdrazena analyticalevaluationofthenewsealautoanalyzer3highresolutionforurinaryiodinedetermination
AT jukictomislav analyticalevaluationofthenewsealautoanalyzer3highresolutionforurinaryiodinedetermination
AT mihaljevicivan analyticalevaluationofthenewsealautoanalyzer3highresolutionforurinaryiodinedetermination
AT kusiczvonko analyticalevaluationofthenewsealautoanalyzer3highresolutionforurinaryiodinedetermination