An abrupt rise of coagulation error messages on ACL TOP automated analysers
INTRODUCTION: Blood coagulation tests (BCT) are very important for clinicians to diagnose bleeding or thrombotic disorders and to monitor anticoagulant therapy. CASE DESCRIPTION: On a Saturday morning, a laboratory technician noted an abrupt rise in the number of coagulation error messages on our AL...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Croatian Society of Medical Biochemistry and Laboratory Medicine
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6559622/ https://www.ncbi.nlm.nih.gov/pubmed/31223268 http://dx.doi.org/10.11613/BM.2019.021002 |
_version_ | 1783425830850920448 |
---|---|
author | Calcoen, Bas Desmet, Koen Vermeersch, Pieter |
author_facet | Calcoen, Bas Desmet, Koen Vermeersch, Pieter |
author_sort | Calcoen, Bas |
collection | PubMed |
description | INTRODUCTION: Blood coagulation tests (BCT) are very important for clinicians to diagnose bleeding or thrombotic disorders and to monitor anticoagulant therapy. CASE DESCRIPTION: On a Saturday morning, a laboratory technician noted an abrupt rise in the number of coagulation error messages on our ALC TOP analysers. Visual inspection revealed the presence of partially and/or fully clotted citrate tubes and prompted the clinical biologist to further investigate a potential preanalytical cause. CONSIDERED CAUSES: Partially or fully clotted blood in citrate tubes can have multiple causes including improper mixing of the tube, under- or overfilling or combining blood samples from different tubes into one citrate tube. WHAT HAPPENED: The affected citrate tubes originated mostly from the same clinical departments. Moreover, all the affected tubes shared the same lot number (1 of 7 in use at the time). Visual inspection of 7 unopened boxes of 100 citrate tubes of this lot number revealed one box with nine completely empty and two partially filled tubes and one box with two partially filled tubes. No under-filled tubes were found in the other 5 boxes. DISCUSSION: The blood to additive ratio is crucial for BCT. A sudden rise in clot errors should trigger a thorough investigation to identify the cause. MAIN LESSON: Laboratories should regularly monitor and evaluate the percentage of clotted samples as a quality indicator at scheduled time points. A problem with the volume of additive in citrate tubes should be considered as a possible cause. |
format | Online Article Text |
id | pubmed-6559622 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Croatian Society of Medical Biochemistry and Laboratory Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-65596222019-06-20 An abrupt rise of coagulation error messages on ACL TOP automated analysers Calcoen, Bas Desmet, Koen Vermeersch, Pieter Biochem Med (Zagreb) Preanalytical Mysteries INTRODUCTION: Blood coagulation tests (BCT) are very important for clinicians to diagnose bleeding or thrombotic disorders and to monitor anticoagulant therapy. CASE DESCRIPTION: On a Saturday morning, a laboratory technician noted an abrupt rise in the number of coagulation error messages on our ALC TOP analysers. Visual inspection revealed the presence of partially and/or fully clotted citrate tubes and prompted the clinical biologist to further investigate a potential preanalytical cause. CONSIDERED CAUSES: Partially or fully clotted blood in citrate tubes can have multiple causes including improper mixing of the tube, under- or overfilling or combining blood samples from different tubes into one citrate tube. WHAT HAPPENED: The affected citrate tubes originated mostly from the same clinical departments. Moreover, all the affected tubes shared the same lot number (1 of 7 in use at the time). Visual inspection of 7 unopened boxes of 100 citrate tubes of this lot number revealed one box with nine completely empty and two partially filled tubes and one box with two partially filled tubes. No under-filled tubes were found in the other 5 boxes. DISCUSSION: The blood to additive ratio is crucial for BCT. A sudden rise in clot errors should trigger a thorough investigation to identify the cause. MAIN LESSON: Laboratories should regularly monitor and evaluate the percentage of clotted samples as a quality indicator at scheduled time points. A problem with the volume of additive in citrate tubes should be considered as a possible cause. Croatian Society of Medical Biochemistry and Laboratory Medicine 2019-06-15 2019-06-15 /pmc/articles/PMC6559622/ /pubmed/31223268 http://dx.doi.org/10.11613/BM.2019.021002 Text en ©Croatian Society of Medical Biochemistry and Laboratory Medicine. This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Preanalytical Mysteries Calcoen, Bas Desmet, Koen Vermeersch, Pieter An abrupt rise of coagulation error messages on ACL TOP automated analysers |
title | An abrupt rise of coagulation error messages on ACL TOP automated analysers |
title_full | An abrupt rise of coagulation error messages on ACL TOP automated analysers |
title_fullStr | An abrupt rise of coagulation error messages on ACL TOP automated analysers |
title_full_unstemmed | An abrupt rise of coagulation error messages on ACL TOP automated analysers |
title_short | An abrupt rise of coagulation error messages on ACL TOP automated analysers |
title_sort | abrupt rise of coagulation error messages on acl top automated analysers |
topic | Preanalytical Mysteries |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6559622/ https://www.ncbi.nlm.nih.gov/pubmed/31223268 http://dx.doi.org/10.11613/BM.2019.021002 |
work_keys_str_mv | AT calcoenbas anabruptriseofcoagulationerrormessagesonacltopautomatedanalysers AT desmetkoen anabruptriseofcoagulationerrormessagesonacltopautomatedanalysers AT vermeerschpieter anabruptriseofcoagulationerrormessagesonacltopautomatedanalysers AT calcoenbas abruptriseofcoagulationerrormessagesonacltopautomatedanalysers AT desmetkoen abruptriseofcoagulationerrormessagesonacltopautomatedanalysers AT vermeerschpieter abruptriseofcoagulationerrormessagesonacltopautomatedanalysers |