Description of interference in the measurement of troponin T by a high-sensitivity method
INTRODUCTION: Measurement of high-sensitivity troponin T (hs-TnT) has become an essential step in the diagnosis of acute myocardial infarction. This high-sensitivity method allows quantifying the concentration of troponin T in blood of healthy subjects with a lower inaccuracy compared to previous re...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Croatian Society of Medical Biochemistry and Laboratory Medicine
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6559625/ https://www.ncbi.nlm.nih.gov/pubmed/31223267 http://dx.doi.org/10.11613/BM.2019.021001 |
_version_ | 1783425831536689152 |
---|---|
author | Aliste-Fernández, Miguel Sole-Enrech, Gemma Cano-Corres, Ruth Teodoro-Marin, Silvia Berlanga-Escalera, Eugenio |
author_facet | Aliste-Fernández, Miguel Sole-Enrech, Gemma Cano-Corres, Ruth Teodoro-Marin, Silvia Berlanga-Escalera, Eugenio |
author_sort | Aliste-Fernández, Miguel |
collection | PubMed |
description | INTRODUCTION: Measurement of high-sensitivity troponin T (hs-TnT) has become an essential step in the diagnosis of acute myocardial infarction. This high-sensitivity method allows quantifying the concentration of troponin T in blood of healthy subjects with a lower inaccuracy compared to previous reagent generations. However, the presence of certain compounds in the sample may interfere with the result. We present a patient who had repeatedly high concentrations of hs-TnT in the serum sample that did not agreed with the signs and symptoms. In addition, ultrasensitive troponin I concentration was undetectable. MATERIALS AND METHODS: To investigate the presence of an interfering compound, different analysis were carried out. In order to discard macro complexes in the sample, the serum was precipitated with polyethylene glycol. In addition, the serum was incubated with Scantibodies Heterophilic Blocking Tube, which can block heterophilic antibodies. Finally, a size exclusion chromatography of the sample was performed by the manufacturer. WHAT HAPPENED: The interfering substance was allocated into fractions with proteins of 150kDa, corresponding to high molecular weight proteins like immunoglobulin G (IgG). This compound was responsible for the falsely elevated hs-TnT results and it affected only the high-sensitivity methods. MAIN LESSON: The detected interfering compound was probably an IgG. This type of interference must be kept in mind in front of discordant results, even if they are extremely rare. Therefore, interdisciplinary cooperation between clinicians, laboratory and manufacturer is essential. |
format | Online Article Text |
id | pubmed-6559625 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Croatian Society of Medical Biochemistry and Laboratory Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-65596252019-06-20 Description of interference in the measurement of troponin T by a high-sensitivity method Aliste-Fernández, Miguel Sole-Enrech, Gemma Cano-Corres, Ruth Teodoro-Marin, Silvia Berlanga-Escalera, Eugenio Biochem Med (Zagreb) Preanalytical Mysteries INTRODUCTION: Measurement of high-sensitivity troponin T (hs-TnT) has become an essential step in the diagnosis of acute myocardial infarction. This high-sensitivity method allows quantifying the concentration of troponin T in blood of healthy subjects with a lower inaccuracy compared to previous reagent generations. However, the presence of certain compounds in the sample may interfere with the result. We present a patient who had repeatedly high concentrations of hs-TnT in the serum sample that did not agreed with the signs and symptoms. In addition, ultrasensitive troponin I concentration was undetectable. MATERIALS AND METHODS: To investigate the presence of an interfering compound, different analysis were carried out. In order to discard macro complexes in the sample, the serum was precipitated with polyethylene glycol. In addition, the serum was incubated with Scantibodies Heterophilic Blocking Tube, which can block heterophilic antibodies. Finally, a size exclusion chromatography of the sample was performed by the manufacturer. WHAT HAPPENED: The interfering substance was allocated into fractions with proteins of 150kDa, corresponding to high molecular weight proteins like immunoglobulin G (IgG). This compound was responsible for the falsely elevated hs-TnT results and it affected only the high-sensitivity methods. MAIN LESSON: The detected interfering compound was probably an IgG. This type of interference must be kept in mind in front of discordant results, even if they are extremely rare. Therefore, interdisciplinary cooperation between clinicians, laboratory and manufacturer is essential. Croatian Society of Medical Biochemistry and Laboratory Medicine 2019-06-15 2019-06-15 /pmc/articles/PMC6559625/ /pubmed/31223267 http://dx.doi.org/10.11613/BM.2019.021001 Text en ©Croatian Society of Medical Biochemistry and Laboratory Medicine. This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Preanalytical Mysteries Aliste-Fernández, Miguel Sole-Enrech, Gemma Cano-Corres, Ruth Teodoro-Marin, Silvia Berlanga-Escalera, Eugenio Description of interference in the measurement of troponin T by a high-sensitivity method |
title | Description of interference in the measurement of troponin T by a high-sensitivity method |
title_full | Description of interference in the measurement of troponin T by a high-sensitivity method |
title_fullStr | Description of interference in the measurement of troponin T by a high-sensitivity method |
title_full_unstemmed | Description of interference in the measurement of troponin T by a high-sensitivity method |
title_short | Description of interference in the measurement of troponin T by a high-sensitivity method |
title_sort | description of interference in the measurement of troponin t by a high-sensitivity method |
topic | Preanalytical Mysteries |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6559625/ https://www.ncbi.nlm.nih.gov/pubmed/31223267 http://dx.doi.org/10.11613/BM.2019.021001 |
work_keys_str_mv | AT alistefernandezmiguel descriptionofinterferenceinthemeasurementoftroponintbyahighsensitivitymethod AT soleenrechgemma descriptionofinterferenceinthemeasurementoftroponintbyahighsensitivitymethod AT canocorresruth descriptionofinterferenceinthemeasurementoftroponintbyahighsensitivitymethod AT teodoromarinsilvia descriptionofinterferenceinthemeasurementoftroponintbyahighsensitivitymethod AT berlangaescaleraeugenio descriptionofinterferenceinthemeasurementoftroponintbyahighsensitivitymethod |