Description of interference in the measurement of troponin T by a high-sensitivity method

INTRODUCTION: Measurement of high-sensitivity troponin T (hs-TnT) has become an essential step in the diagnosis of acute myocardial infarction. This high-sensitivity method allows quantifying the concentration of troponin T in blood of healthy subjects with a lower inaccuracy compared to previous re...

Descripción completa

Detalles Bibliográficos
Autores principales: Aliste-Fernández, Miguel, Sole-Enrech, Gemma, Cano-Corres, Ruth, Teodoro-Marin, Silvia, Berlanga-Escalera, Eugenio
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Croatian Society of Medical Biochemistry and Laboratory Medicine 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6559625/
https://www.ncbi.nlm.nih.gov/pubmed/31223267
http://dx.doi.org/10.11613/BM.2019.021001
_version_ 1783425831536689152
author Aliste-Fernández, Miguel
Sole-Enrech, Gemma
Cano-Corres, Ruth
Teodoro-Marin, Silvia
Berlanga-Escalera, Eugenio
author_facet Aliste-Fernández, Miguel
Sole-Enrech, Gemma
Cano-Corres, Ruth
Teodoro-Marin, Silvia
Berlanga-Escalera, Eugenio
author_sort Aliste-Fernández, Miguel
collection PubMed
description INTRODUCTION: Measurement of high-sensitivity troponin T (hs-TnT) has become an essential step in the diagnosis of acute myocardial infarction. This high-sensitivity method allows quantifying the concentration of troponin T in blood of healthy subjects with a lower inaccuracy compared to previous reagent generations. However, the presence of certain compounds in the sample may interfere with the result. We present a patient who had repeatedly high concentrations of hs-TnT in the serum sample that did not agreed with the signs and symptoms. In addition, ultrasensitive troponin I concentration was undetectable. MATERIALS AND METHODS: To investigate the presence of an interfering compound, different analysis were carried out. In order to discard macro complexes in the sample, the serum was precipitated with polyethylene glycol. In addition, the serum was incubated with Scantibodies Heterophilic Blocking Tube, which can block heterophilic antibodies. Finally, a size exclusion chromatography of the sample was performed by the manufacturer. WHAT HAPPENED: The interfering substance was allocated into fractions with proteins of 150kDa, corresponding to high molecular weight proteins like immunoglobulin G (IgG). This compound was responsible for the falsely elevated hs-TnT results and it affected only the high-sensitivity methods. MAIN LESSON: The detected interfering compound was probably an IgG. This type of interference must be kept in mind in front of discordant results, even if they are extremely rare. Therefore, interdisciplinary cooperation between clinicians, laboratory and manufacturer is essential.
format Online
Article
Text
id pubmed-6559625
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Croatian Society of Medical Biochemistry and Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-65596252019-06-20 Description of interference in the measurement of troponin T by a high-sensitivity method Aliste-Fernández, Miguel Sole-Enrech, Gemma Cano-Corres, Ruth Teodoro-Marin, Silvia Berlanga-Escalera, Eugenio Biochem Med (Zagreb) Preanalytical Mysteries INTRODUCTION: Measurement of high-sensitivity troponin T (hs-TnT) has become an essential step in the diagnosis of acute myocardial infarction. This high-sensitivity method allows quantifying the concentration of troponin T in blood of healthy subjects with a lower inaccuracy compared to previous reagent generations. However, the presence of certain compounds in the sample may interfere with the result. We present a patient who had repeatedly high concentrations of hs-TnT in the serum sample that did not agreed with the signs and symptoms. In addition, ultrasensitive troponin I concentration was undetectable. MATERIALS AND METHODS: To investigate the presence of an interfering compound, different analysis were carried out. In order to discard macro complexes in the sample, the serum was precipitated with polyethylene glycol. In addition, the serum was incubated with Scantibodies Heterophilic Blocking Tube, which can block heterophilic antibodies. Finally, a size exclusion chromatography of the sample was performed by the manufacturer. WHAT HAPPENED: The interfering substance was allocated into fractions with proteins of 150kDa, corresponding to high molecular weight proteins like immunoglobulin G (IgG). This compound was responsible for the falsely elevated hs-TnT results and it affected only the high-sensitivity methods. MAIN LESSON: The detected interfering compound was probably an IgG. This type of interference must be kept in mind in front of discordant results, even if they are extremely rare. Therefore, interdisciplinary cooperation between clinicians, laboratory and manufacturer is essential. Croatian Society of Medical Biochemistry and Laboratory Medicine 2019-06-15 2019-06-15 /pmc/articles/PMC6559625/ /pubmed/31223267 http://dx.doi.org/10.11613/BM.2019.021001 Text en ©Croatian Society of Medical Biochemistry and Laboratory Medicine. This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Preanalytical Mysteries
Aliste-Fernández, Miguel
Sole-Enrech, Gemma
Cano-Corres, Ruth
Teodoro-Marin, Silvia
Berlanga-Escalera, Eugenio
Description of interference in the measurement of troponin T by a high-sensitivity method
title Description of interference in the measurement of troponin T by a high-sensitivity method
title_full Description of interference in the measurement of troponin T by a high-sensitivity method
title_fullStr Description of interference in the measurement of troponin T by a high-sensitivity method
title_full_unstemmed Description of interference in the measurement of troponin T by a high-sensitivity method
title_short Description of interference in the measurement of troponin T by a high-sensitivity method
title_sort description of interference in the measurement of troponin t by a high-sensitivity method
topic Preanalytical Mysteries
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6559625/
https://www.ncbi.nlm.nih.gov/pubmed/31223267
http://dx.doi.org/10.11613/BM.2019.021001
work_keys_str_mv AT alistefernandezmiguel descriptionofinterferenceinthemeasurementoftroponintbyahighsensitivitymethod
AT soleenrechgemma descriptionofinterferenceinthemeasurementoftroponintbyahighsensitivitymethod
AT canocorresruth descriptionofinterferenceinthemeasurementoftroponintbyahighsensitivitymethod
AT teodoromarinsilvia descriptionofinterferenceinthemeasurementoftroponintbyahighsensitivitymethod
AT berlangaescaleraeugenio descriptionofinterferenceinthemeasurementoftroponintbyahighsensitivitymethod