Design, Fabrication and Mass-spectrometric Studies of a Micro Ion Source for High-Field Asymmetric Waveform Ion Mobility Spectrometry

A needle-to-cylinder electrode, adopted as an ion source for high-field asymmetric ion mobility spectrometry (FAIMS), is designed and fabricated by lithographie, galvanoformung and abformung (LIGA) technology. The needle, with a tip diameter of 20 μm and thickness of 20 μm, and a cylinder, with a di...

Descripción completa

Detalles Bibliográficos
Autores principales: Li, Hua, Yun, Hongmei, Du, Xiaoxia, Guo, Chaoqun, Zeng, Ruosheng, Jiang, Yongrong, Chen, Zhencheng
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6562519/
https://www.ncbi.nlm.nih.gov/pubmed/31035626
http://dx.doi.org/10.3390/mi10050286
Descripción
Sumario:A needle-to-cylinder electrode, adopted as an ion source for high-field asymmetric ion mobility spectrometry (FAIMS), is designed and fabricated by lithographie, galvanoformung and abformung (LIGA) technology. The needle, with a tip diameter of 20 μm and thickness of 20 μm, and a cylinder, with a diameter of 400 μm, were connected to the negative high voltage and ground, respectively. A negative corona and glow discharge were realized. For acetone with a density of 99.7 ppm, ethanol with a density of 300 ppm, and acetic ether with a density of 99.3 ppm, the sample gas was ionized by the needle-to-cylinder chip and the ions were detected by an LTQ XL™ (Thermo Scientific Corp.) mass spectrometer. The mass spectra show that the ions are mainly the protonated monomer, the proton bound dimer, and an ion-H(2)O molecule cluster. In tandem with a FAIMS system, the FAIMS spectra show that the resolving power increases with an increase in the RF voltage. The obtained experimental results showed that the micro needle-to-cylinder chip may serve as a miniature, low cost and non-radioactive ion source for FAIMS.