Clinical Evaluation of a Two-dimensional Liquid-Filled Ion chamber Detector Array for Verification of High Modulation Small Fields in Radiotherapy

INTRODUCTION: Clinical evaluation of a two-dimensional (2D) liquid-filled ion chamber detector array used in the verification of highly modulated small beams of stereotactic body radiation therapy (SBRT) has been conducted. MATERIALS AND METHODS: Measurements with the Octavius 1000 SRS (PTW, Freibur...

Descripción completa

Detalles Bibliográficos
Autores principales: Markovic, Miljenko, Narayanasamy, Ganesh, Stathakis, Sotirios, Mavroidis, Panayiotis, Jurkovic, Ines-Ana, Saenz, Daniel, Papanikolaou, Niko
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6580813/
https://www.ncbi.nlm.nih.gov/pubmed/31359926
http://dx.doi.org/10.4103/jmp.JMP_4_19
_version_ 1783428093314072576
author Markovic, Miljenko
Narayanasamy, Ganesh
Stathakis, Sotirios
Mavroidis, Panayiotis
Jurkovic, Ines-Ana
Saenz, Daniel
Papanikolaou, Niko
author_facet Markovic, Miljenko
Narayanasamy, Ganesh
Stathakis, Sotirios
Mavroidis, Panayiotis
Jurkovic, Ines-Ana
Saenz, Daniel
Papanikolaou, Niko
author_sort Markovic, Miljenko
collection PubMed
description INTRODUCTION: Clinical evaluation of a two-dimensional (2D) liquid-filled ion chamber detector array used in the verification of highly modulated small beams of stereotactic body radiation therapy (SBRT) has been conducted. MATERIALS AND METHODS: Measurements with the Octavius 1000 SRS (PTW, Freiburg, Germany) detector with 977 liquid-filled ion chambers were compared against EDR2 film and PTW Octavius Seven29. The performance of detector array has been evaluated on ten SBRT patient plans. Dose profiles of individual and composite fields' calculated using Pinnacle(3) treatment planning system were compared against measurements with Octavius 1000 SRS detector array, EDR2 film, and Octavius Seven29 detector. Gamma index and profile comparison were used in the evaluation and assessment of the detector's performance. RESULTS: The Gamma index measurements show agreement between Pinnacle(3) computations and Octavius 1000 SRS array, PTW Octavius Seven29, and EDR2 film for >90% of the points using 2%, 2 mm tolerance criteria. Profiles obtained with the Octavius 1000 SRS were in agreement with the EDR2 film profiles, demonstrating the detector's superior sampling rate. CONCLUSIONS: The Octavius 1000 SRS is a dosimetrically accurate device to perform quality assurance checks in SBRT treatments. The broad range of measurements performed in this study quantified the dosimetric accuracy of Octavius 1000 SRS detector in the clinical setup of the small fields in radiotherapy.
format Online
Article
Text
id pubmed-6580813
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-65808132019-07-29 Clinical Evaluation of a Two-dimensional Liquid-Filled Ion chamber Detector Array for Verification of High Modulation Small Fields in Radiotherapy Markovic, Miljenko Narayanasamy, Ganesh Stathakis, Sotirios Mavroidis, Panayiotis Jurkovic, Ines-Ana Saenz, Daniel Papanikolaou, Niko J Med Phys Original Article INTRODUCTION: Clinical evaluation of a two-dimensional (2D) liquid-filled ion chamber detector array used in the verification of highly modulated small beams of stereotactic body radiation therapy (SBRT) has been conducted. MATERIALS AND METHODS: Measurements with the Octavius 1000 SRS (PTW, Freiburg, Germany) detector with 977 liquid-filled ion chambers were compared against EDR2 film and PTW Octavius Seven29. The performance of detector array has been evaluated on ten SBRT patient plans. Dose profiles of individual and composite fields' calculated using Pinnacle(3) treatment planning system were compared against measurements with Octavius 1000 SRS detector array, EDR2 film, and Octavius Seven29 detector. Gamma index and profile comparison were used in the evaluation and assessment of the detector's performance. RESULTS: The Gamma index measurements show agreement between Pinnacle(3) computations and Octavius 1000 SRS array, PTW Octavius Seven29, and EDR2 film for >90% of the points using 2%, 2 mm tolerance criteria. Profiles obtained with the Octavius 1000 SRS were in agreement with the EDR2 film profiles, demonstrating the detector's superior sampling rate. CONCLUSIONS: The Octavius 1000 SRS is a dosimetrically accurate device to perform quality assurance checks in SBRT treatments. The broad range of measurements performed in this study quantified the dosimetric accuracy of Octavius 1000 SRS detector in the clinical setup of the small fields in radiotherapy. Wolters Kluwer - Medknow 2019 /pmc/articles/PMC6580813/ /pubmed/31359926 http://dx.doi.org/10.4103/jmp.JMP_4_19 Text en Copyright: © 2019 Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Markovic, Miljenko
Narayanasamy, Ganesh
Stathakis, Sotirios
Mavroidis, Panayiotis
Jurkovic, Ines-Ana
Saenz, Daniel
Papanikolaou, Niko
Clinical Evaluation of a Two-dimensional Liquid-Filled Ion chamber Detector Array for Verification of High Modulation Small Fields in Radiotherapy
title Clinical Evaluation of a Two-dimensional Liquid-Filled Ion chamber Detector Array for Verification of High Modulation Small Fields in Radiotherapy
title_full Clinical Evaluation of a Two-dimensional Liquid-Filled Ion chamber Detector Array for Verification of High Modulation Small Fields in Radiotherapy
title_fullStr Clinical Evaluation of a Two-dimensional Liquid-Filled Ion chamber Detector Array for Verification of High Modulation Small Fields in Radiotherapy
title_full_unstemmed Clinical Evaluation of a Two-dimensional Liquid-Filled Ion chamber Detector Array for Verification of High Modulation Small Fields in Radiotherapy
title_short Clinical Evaluation of a Two-dimensional Liquid-Filled Ion chamber Detector Array for Verification of High Modulation Small Fields in Radiotherapy
title_sort clinical evaluation of a two-dimensional liquid-filled ion chamber detector array for verification of high modulation small fields in radiotherapy
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6580813/
https://www.ncbi.nlm.nih.gov/pubmed/31359926
http://dx.doi.org/10.4103/jmp.JMP_4_19
work_keys_str_mv AT markovicmiljenko clinicalevaluationofatwodimensionalliquidfilledionchamberdetectorarrayforverificationofhighmodulationsmallfieldsinradiotherapy
AT narayanasamyganesh clinicalevaluationofatwodimensionalliquidfilledionchamberdetectorarrayforverificationofhighmodulationsmallfieldsinradiotherapy
AT stathakissotirios clinicalevaluationofatwodimensionalliquidfilledionchamberdetectorarrayforverificationofhighmodulationsmallfieldsinradiotherapy
AT mavroidispanayiotis clinicalevaluationofatwodimensionalliquidfilledionchamberdetectorarrayforverificationofhighmodulationsmallfieldsinradiotherapy
AT jurkovicinesana clinicalevaluationofatwodimensionalliquidfilledionchamberdetectorarrayforverificationofhighmodulationsmallfieldsinradiotherapy
AT saenzdaniel clinicalevaluationofatwodimensionalliquidfilledionchamberdetectorarrayforverificationofhighmodulationsmallfieldsinradiotherapy
AT papanikolaouniko clinicalevaluationofatwodimensionalliquidfilledionchamberdetectorarrayforverificationofhighmodulationsmallfieldsinradiotherapy