Investigation of Internal Target Volumes Using Device and Deviceless Four-dimensional Respiratory Monitoring Systems for Moving Targets in Four-dimensional Computed Tomography Acquisition

AIMS AND OBJECTIVES: The influence of target motion on the reconstructed internal target volume (ITV) for device-based (DB) external surrogate system and Smart deviceless (DL) 4-dimensional (4D) system were compared in a controlled phantom experiment. The volumetric changes in reconstructed ITVs fro...

Descripción completa

Detalles Bibliográficos
Autores principales: Holla, Raghavendra, Khanna, D., Barsing, Shubhangi, Pillai, Bhaskaran K., Ganesh, Tharmarnadar
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6580822/
https://www.ncbi.nlm.nih.gov/pubmed/31359924
http://dx.doi.org/10.4103/jmp.JMP_101_18
_version_ 1783428095401787392
author Holla, Raghavendra
Khanna, D.
Barsing, Shubhangi
Pillai, Bhaskaran K.
Ganesh, Tharmarnadar
author_facet Holla, Raghavendra
Khanna, D.
Barsing, Shubhangi
Pillai, Bhaskaran K.
Ganesh, Tharmarnadar
author_sort Holla, Raghavendra
collection PubMed
description AIMS AND OBJECTIVES: The influence of target motion on the reconstructed internal target volume (ITV) for device-based (DB) external surrogate system and Smart deviceless (DL) 4-dimensional (4D) system were compared in a controlled phantom experiment. The volumetric changes in reconstructed ITVs from the average intensity projection (AveIP) images using DB method (Anzai Respiratory Gating System, ANZAI MEDICAL CO., LTD, Japan) and DL method (Smart deviceless 4D system by GE Medical Systems (Chicago, USA)) with the theoretical true volume (ITV(th)) for moving target with the increasing target motion in anterior-posterior (A-P), lateral (left-right [L-R]) and inferior-superior (S-I) directions were assessed. MATERIALS AND METHODS: 4D computed tomography (4DCT) of CIRS dynamic phantom (Computerized Imaging Reference Systems Inc., Norfolk, VA, USA) with 2.5 cm diameter spherical target of volume 8.2 cc programmed to move in a cos(4)(x) motion pattern placed in the lung volume were acquired for various target motion pattern using DB and DL method of gating. AveIP images of 10 phase binned image sets were generated and ITVs were delineated. RESULTS: The maximum absolute percent differences between ITV(ave) and ITV(th) for DL and DB methods were 15.91% and 4.94 % respectively for target motion of 5 mm in AP with 15 mm S-I direction. When the S-I motion was decreased to 10 mm, the observed % difference of the ITVs were also decreased to 12.5% and 0.3% for DL and DB method. When the lateral [L-R] motion was varied from 0 mm to 5 mm for S-I motion of 5 mm to 15 mm, the differences in the ITVs were significant (P = 0.004) with the maximum absolute percent difference of 18.61% and 4.94 % for DL and DB gating. With the simultaneous motion of the target in all the 3 directions, the difference in the reconstructed ITVs were statistically significant for DL method (P = 0.0002) and insignificant for DB method (P = 0.06) with an average increase of 10% in ITVDL against 2% in the ITVDB. The difference in ITVDL was significant for the target motion above 3 mm in A-P and L-R directions for S-I movement of above 10 mm (P = 0.0002). However, for low excursions of the target movement, no significant difference in the ITVs were observed (P > 0.06). In general, ITVDBs were closer to the ITVth (within 7.8%) than ITVDL (18.61%). CONCLUSION: The results showed that the DL method is an effective way of image sorting in 4D acquisition for smaller target excursion. When the target motion exceeds 3 mm in A-P and L-R directions with S-I more than 10 mm, DB method is the choice due to its accuracy in reproducing the absolute target volume.
format Online
Article
Text
id pubmed-6580822
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-65808222019-07-29 Investigation of Internal Target Volumes Using Device and Deviceless Four-dimensional Respiratory Monitoring Systems for Moving Targets in Four-dimensional Computed Tomography Acquisition Holla, Raghavendra Khanna, D. Barsing, Shubhangi Pillai, Bhaskaran K. Ganesh, Tharmarnadar J Med Phys Original Article AIMS AND OBJECTIVES: The influence of target motion on the reconstructed internal target volume (ITV) for device-based (DB) external surrogate system and Smart deviceless (DL) 4-dimensional (4D) system were compared in a controlled phantom experiment. The volumetric changes in reconstructed ITVs from the average intensity projection (AveIP) images using DB method (Anzai Respiratory Gating System, ANZAI MEDICAL CO., LTD, Japan) and DL method (Smart deviceless 4D system by GE Medical Systems (Chicago, USA)) with the theoretical true volume (ITV(th)) for moving target with the increasing target motion in anterior-posterior (A-P), lateral (left-right [L-R]) and inferior-superior (S-I) directions were assessed. MATERIALS AND METHODS: 4D computed tomography (4DCT) of CIRS dynamic phantom (Computerized Imaging Reference Systems Inc., Norfolk, VA, USA) with 2.5 cm diameter spherical target of volume 8.2 cc programmed to move in a cos(4)(x) motion pattern placed in the lung volume were acquired for various target motion pattern using DB and DL method of gating. AveIP images of 10 phase binned image sets were generated and ITVs were delineated. RESULTS: The maximum absolute percent differences between ITV(ave) and ITV(th) for DL and DB methods were 15.91% and 4.94 % respectively for target motion of 5 mm in AP with 15 mm S-I direction. When the S-I motion was decreased to 10 mm, the observed % difference of the ITVs were also decreased to 12.5% and 0.3% for DL and DB method. When the lateral [L-R] motion was varied from 0 mm to 5 mm for S-I motion of 5 mm to 15 mm, the differences in the ITVs were significant (P = 0.004) with the maximum absolute percent difference of 18.61% and 4.94 % for DL and DB gating. With the simultaneous motion of the target in all the 3 directions, the difference in the reconstructed ITVs were statistically significant for DL method (P = 0.0002) and insignificant for DB method (P = 0.06) with an average increase of 10% in ITVDL against 2% in the ITVDB. The difference in ITVDL was significant for the target motion above 3 mm in A-P and L-R directions for S-I movement of above 10 mm (P = 0.0002). However, for low excursions of the target movement, no significant difference in the ITVs were observed (P > 0.06). In general, ITVDBs were closer to the ITVth (within 7.8%) than ITVDL (18.61%). CONCLUSION: The results showed that the DL method is an effective way of image sorting in 4D acquisition for smaller target excursion. When the target motion exceeds 3 mm in A-P and L-R directions with S-I more than 10 mm, DB method is the choice due to its accuracy in reproducing the absolute target volume. Wolters Kluwer - Medknow 2019 /pmc/articles/PMC6580822/ /pubmed/31359924 http://dx.doi.org/10.4103/jmp.JMP_101_18 Text en Copyright: © 2019 Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Holla, Raghavendra
Khanna, D.
Barsing, Shubhangi
Pillai, Bhaskaran K.
Ganesh, Tharmarnadar
Investigation of Internal Target Volumes Using Device and Deviceless Four-dimensional Respiratory Monitoring Systems for Moving Targets in Four-dimensional Computed Tomography Acquisition
title Investigation of Internal Target Volumes Using Device and Deviceless Four-dimensional Respiratory Monitoring Systems for Moving Targets in Four-dimensional Computed Tomography Acquisition
title_full Investigation of Internal Target Volumes Using Device and Deviceless Four-dimensional Respiratory Monitoring Systems for Moving Targets in Four-dimensional Computed Tomography Acquisition
title_fullStr Investigation of Internal Target Volumes Using Device and Deviceless Four-dimensional Respiratory Monitoring Systems for Moving Targets in Four-dimensional Computed Tomography Acquisition
title_full_unstemmed Investigation of Internal Target Volumes Using Device and Deviceless Four-dimensional Respiratory Monitoring Systems for Moving Targets in Four-dimensional Computed Tomography Acquisition
title_short Investigation of Internal Target Volumes Using Device and Deviceless Four-dimensional Respiratory Monitoring Systems for Moving Targets in Four-dimensional Computed Tomography Acquisition
title_sort investigation of internal target volumes using device and deviceless four-dimensional respiratory monitoring systems for moving targets in four-dimensional computed tomography acquisition
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6580822/
https://www.ncbi.nlm.nih.gov/pubmed/31359924
http://dx.doi.org/10.4103/jmp.JMP_101_18
work_keys_str_mv AT hollaraghavendra investigationofinternaltargetvolumesusingdeviceanddevicelessfourdimensionalrespiratorymonitoringsystemsformovingtargetsinfourdimensionalcomputedtomographyacquisition
AT khannad investigationofinternaltargetvolumesusingdeviceanddevicelessfourdimensionalrespiratorymonitoringsystemsformovingtargetsinfourdimensionalcomputedtomographyacquisition
AT barsingshubhangi investigationofinternaltargetvolumesusingdeviceanddevicelessfourdimensionalrespiratorymonitoringsystemsformovingtargetsinfourdimensionalcomputedtomographyacquisition
AT pillaibhaskarank investigationofinternaltargetvolumesusingdeviceanddevicelessfourdimensionalrespiratorymonitoringsystemsformovingtargetsinfourdimensionalcomputedtomographyacquisition
AT ganeshtharmarnadar investigationofinternaltargetvolumesusingdeviceanddevicelessfourdimensionalrespiratorymonitoringsystemsformovingtargetsinfourdimensionalcomputedtomographyacquisition