Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique

PURPOSE: The purpose of this study was to retrospectively compare the technical success and reliability of the measurements made using two 2-dimensional (2D) shear wave elastography (SWE) systems using the comb-push technique from the same manufacturer and to assess the intersystem reproducibility o...

Descripción completa

Detalles Bibliográficos
Autores principales: Ryu, Hwaseong, Ahn, Su Joa, Yoon, Jeong Hee, Lee, Jeong Min
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Society of Ultrasound in Medicine 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6595128/
https://www.ncbi.nlm.nih.gov/pubmed/30744303
http://dx.doi.org/10.14366/usg.18046
_version_ 1783430349445922816
author Ryu, Hwaseong
Ahn, Su Joa
Yoon, Jeong Hee
Lee, Jeong Min
author_facet Ryu, Hwaseong
Ahn, Su Joa
Yoon, Jeong Hee
Lee, Jeong Min
author_sort Ryu, Hwaseong
collection PubMed
description PURPOSE: The purpose of this study was to retrospectively compare the technical success and reliability of the measurements made using two 2-dimensional (2D) shear wave elastography (SWE) systems using the comb-push technique from the same manufacturer and to assess the intersystem reproducibility of the resultant liver stiffness (LS) measurements. METHODS: Ninety-four patients with suspected chronic liver diseases were included in this retrospective study. LS measurements were obtained using two 2D-SWE systems (LOGIQ E9 and LOGIQ S8) from the same manufacturer, with transient elastography (TE) serving as the reference standard, on the same day. The technical success rates and reliability of the measurements of the two 2D-SWE systems were compared. LS values measured using the two 2D-SWE systems and TE were correlated using Spearman correlation coefficients and 95% Bland-Altman limits of agreement. Thereafter, Bland-Altman limits of agreement and intraclass correlation coefficients (ICCs) were used to analyze the intersystem reproducibility of LS measurements. RESULTS: The two 2D-SWE systems showed similar technical success rates (98.9% for both) and reliability of LS measurements (92.3% for the LOGIQ E9, 91.2% for the LOGIQ S8; P=0.185). Despite the excellent correlation (ICC=0.92), the mean LS measurements obtained by the two 2D-SWE systems were significantly different (LOGIQ E9, 6.57±2.33 kPa; LOGIQ S8, 6.90±6.64 kPa; P=0.018). CONCLUSION: Significant intersystem variability was observed in the LS measurements made using the two 2D-SWE systems. Therefore, even 2D-SWE systems from the same manufacturer should not be used interchangeably in longitudinal follow-up.
format Online
Article
Text
id pubmed-6595128
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Korean Society of Ultrasound in Medicine
record_format MEDLINE/PubMed
spelling pubmed-65951282019-07-06 Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique Ryu, Hwaseong Ahn, Su Joa Yoon, Jeong Hee Lee, Jeong Min Ultrasonography Original Article PURPOSE: The purpose of this study was to retrospectively compare the technical success and reliability of the measurements made using two 2-dimensional (2D) shear wave elastography (SWE) systems using the comb-push technique from the same manufacturer and to assess the intersystem reproducibility of the resultant liver stiffness (LS) measurements. METHODS: Ninety-four patients with suspected chronic liver diseases were included in this retrospective study. LS measurements were obtained using two 2D-SWE systems (LOGIQ E9 and LOGIQ S8) from the same manufacturer, with transient elastography (TE) serving as the reference standard, on the same day. The technical success rates and reliability of the measurements of the two 2D-SWE systems were compared. LS values measured using the two 2D-SWE systems and TE were correlated using Spearman correlation coefficients and 95% Bland-Altman limits of agreement. Thereafter, Bland-Altman limits of agreement and intraclass correlation coefficients (ICCs) were used to analyze the intersystem reproducibility of LS measurements. RESULTS: The two 2D-SWE systems showed similar technical success rates (98.9% for both) and reliability of LS measurements (92.3% for the LOGIQ E9, 91.2% for the LOGIQ S8; P=0.185). Despite the excellent correlation (ICC=0.92), the mean LS measurements obtained by the two 2D-SWE systems were significantly different (LOGIQ E9, 6.57±2.33 kPa; LOGIQ S8, 6.90±6.64 kPa; P=0.018). CONCLUSION: Significant intersystem variability was observed in the LS measurements made using the two 2D-SWE systems. Therefore, even 2D-SWE systems from the same manufacturer should not be used interchangeably in longitudinal follow-up. Korean Society of Ultrasound in Medicine 2019-07 2018-11-23 /pmc/articles/PMC6595128/ /pubmed/30744303 http://dx.doi.org/10.14366/usg.18046 Text en Copyright © 2019 Korean Society of Ultrasound in Medicine (KSUM) This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted noncommercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Ryu, Hwaseong
Ahn, Su Joa
Yoon, Jeong Hee
Lee, Jeong Min
Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique
title Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique
title_full Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique
title_fullStr Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique
title_full_unstemmed Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique
title_short Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique
title_sort reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6595128/
https://www.ncbi.nlm.nih.gov/pubmed/30744303
http://dx.doi.org/10.14366/usg.18046
work_keys_str_mv AT ryuhwaseong reproducibilityofliverstiffnessmeasurementsmadewithtwodifferent2dimensionalshearwaveelastographysystemsusingthecombpushtechnique
AT ahnsujoa reproducibilityofliverstiffnessmeasurementsmadewithtwodifferent2dimensionalshearwaveelastographysystemsusingthecombpushtechnique
AT yoonjeonghee reproducibilityofliverstiffnessmeasurementsmadewithtwodifferent2dimensionalshearwaveelastographysystemsusingthecombpushtechnique
AT leejeongmin reproducibilityofliverstiffnessmeasurementsmadewithtwodifferent2dimensionalshearwaveelastographysystemsusingthecombpushtechnique