Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique
PURPOSE: The purpose of this study was to retrospectively compare the technical success and reliability of the measurements made using two 2-dimensional (2D) shear wave elastography (SWE) systems using the comb-push technique from the same manufacturer and to assess the intersystem reproducibility o...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Korean Society of Ultrasound in Medicine
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6595128/ https://www.ncbi.nlm.nih.gov/pubmed/30744303 http://dx.doi.org/10.14366/usg.18046 |
_version_ | 1783430349445922816 |
---|---|
author | Ryu, Hwaseong Ahn, Su Joa Yoon, Jeong Hee Lee, Jeong Min |
author_facet | Ryu, Hwaseong Ahn, Su Joa Yoon, Jeong Hee Lee, Jeong Min |
author_sort | Ryu, Hwaseong |
collection | PubMed |
description | PURPOSE: The purpose of this study was to retrospectively compare the technical success and reliability of the measurements made using two 2-dimensional (2D) shear wave elastography (SWE) systems using the comb-push technique from the same manufacturer and to assess the intersystem reproducibility of the resultant liver stiffness (LS) measurements. METHODS: Ninety-four patients with suspected chronic liver diseases were included in this retrospective study. LS measurements were obtained using two 2D-SWE systems (LOGIQ E9 and LOGIQ S8) from the same manufacturer, with transient elastography (TE) serving as the reference standard, on the same day. The technical success rates and reliability of the measurements of the two 2D-SWE systems were compared. LS values measured using the two 2D-SWE systems and TE were correlated using Spearman correlation coefficients and 95% Bland-Altman limits of agreement. Thereafter, Bland-Altman limits of agreement and intraclass correlation coefficients (ICCs) were used to analyze the intersystem reproducibility of LS measurements. RESULTS: The two 2D-SWE systems showed similar technical success rates (98.9% for both) and reliability of LS measurements (92.3% for the LOGIQ E9, 91.2% for the LOGIQ S8; P=0.185). Despite the excellent correlation (ICC=0.92), the mean LS measurements obtained by the two 2D-SWE systems were significantly different (LOGIQ E9, 6.57±2.33 kPa; LOGIQ S8, 6.90±6.64 kPa; P=0.018). CONCLUSION: Significant intersystem variability was observed in the LS measurements made using the two 2D-SWE systems. Therefore, even 2D-SWE systems from the same manufacturer should not be used interchangeably in longitudinal follow-up. |
format | Online Article Text |
id | pubmed-6595128 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Korean Society of Ultrasound in Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-65951282019-07-06 Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique Ryu, Hwaseong Ahn, Su Joa Yoon, Jeong Hee Lee, Jeong Min Ultrasonography Original Article PURPOSE: The purpose of this study was to retrospectively compare the technical success and reliability of the measurements made using two 2-dimensional (2D) shear wave elastography (SWE) systems using the comb-push technique from the same manufacturer and to assess the intersystem reproducibility of the resultant liver stiffness (LS) measurements. METHODS: Ninety-four patients with suspected chronic liver diseases were included in this retrospective study. LS measurements were obtained using two 2D-SWE systems (LOGIQ E9 and LOGIQ S8) from the same manufacturer, with transient elastography (TE) serving as the reference standard, on the same day. The technical success rates and reliability of the measurements of the two 2D-SWE systems were compared. LS values measured using the two 2D-SWE systems and TE were correlated using Spearman correlation coefficients and 95% Bland-Altman limits of agreement. Thereafter, Bland-Altman limits of agreement and intraclass correlation coefficients (ICCs) were used to analyze the intersystem reproducibility of LS measurements. RESULTS: The two 2D-SWE systems showed similar technical success rates (98.9% for both) and reliability of LS measurements (92.3% for the LOGIQ E9, 91.2% for the LOGIQ S8; P=0.185). Despite the excellent correlation (ICC=0.92), the mean LS measurements obtained by the two 2D-SWE systems were significantly different (LOGIQ E9, 6.57±2.33 kPa; LOGIQ S8, 6.90±6.64 kPa; P=0.018). CONCLUSION: Significant intersystem variability was observed in the LS measurements made using the two 2D-SWE systems. Therefore, even 2D-SWE systems from the same manufacturer should not be used interchangeably in longitudinal follow-up. Korean Society of Ultrasound in Medicine 2019-07 2018-11-23 /pmc/articles/PMC6595128/ /pubmed/30744303 http://dx.doi.org/10.14366/usg.18046 Text en Copyright © 2019 Korean Society of Ultrasound in Medicine (KSUM) This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted noncommercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Ryu, Hwaseong Ahn, Su Joa Yoon, Jeong Hee Lee, Jeong Min Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique |
title | Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique |
title_full | Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique |
title_fullStr | Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique |
title_full_unstemmed | Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique |
title_short | Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique |
title_sort | reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6595128/ https://www.ncbi.nlm.nih.gov/pubmed/30744303 http://dx.doi.org/10.14366/usg.18046 |
work_keys_str_mv | AT ryuhwaseong reproducibilityofliverstiffnessmeasurementsmadewithtwodifferent2dimensionalshearwaveelastographysystemsusingthecombpushtechnique AT ahnsujoa reproducibilityofliverstiffnessmeasurementsmadewithtwodifferent2dimensionalshearwaveelastographysystemsusingthecombpushtechnique AT yoonjeonghee reproducibilityofliverstiffnessmeasurementsmadewithtwodifferent2dimensionalshearwaveelastographysystemsusingthecombpushtechnique AT leejeongmin reproducibilityofliverstiffnessmeasurementsmadewithtwodifferent2dimensionalshearwaveelastographysystemsusingthecombpushtechnique |