Quantification of Electromechanical Coupling to Prevent Inappropriate Implantable Cardioverter-Defibrillator Shocks

OBJECTIVES: This study sought to test specialized processing of laser Doppler signals for discriminating ventricular fibrillation (VF) from common causes of inappropriate therapies. BACKGROUND: Inappropriate implantable cardioverter-defibrillator (ICD) therapies remain a clinically important problem...

Descripción completa

Detalles Bibliográficos
Autores principales: Keene, Daniel, Shun-Shin, Matthew J., Arnold, Ahran D., Howard, James P., Lefroy, David, Davies, D. Wyn, Lim, Phang Boon, Ng, Fu Siong, Koa-Wing, Michael, Qureshi, Norman A., Linton, Nick W.F., Shah, Jaymin S., Peters, Nicholas S., Kanagaratnam, Prapa, Francis, Darrel P., Whinnett, Zachary I.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier Inc 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6597902/
https://www.ncbi.nlm.nih.gov/pubmed/31221358
http://dx.doi.org/10.1016/j.jacep.2019.01.025
_version_ 1783430666676862976
author Keene, Daniel
Shun-Shin, Matthew J.
Arnold, Ahran D.
Howard, James P.
Lefroy, David
Davies, D. Wyn
Lim, Phang Boon
Ng, Fu Siong
Koa-Wing, Michael
Qureshi, Norman A.
Linton, Nick W.F.
Shah, Jaymin S.
Peters, Nicholas S.
Kanagaratnam, Prapa
Francis, Darrel P.
Whinnett, Zachary I.
author_facet Keene, Daniel
Shun-Shin, Matthew J.
Arnold, Ahran D.
Howard, James P.
Lefroy, David
Davies, D. Wyn
Lim, Phang Boon
Ng, Fu Siong
Koa-Wing, Michael
Qureshi, Norman A.
Linton, Nick W.F.
Shah, Jaymin S.
Peters, Nicholas S.
Kanagaratnam, Prapa
Francis, Darrel P.
Whinnett, Zachary I.
author_sort Keene, Daniel
collection PubMed
description OBJECTIVES: This study sought to test specialized processing of laser Doppler signals for discriminating ventricular fibrillation (VF) from common causes of inappropriate therapies. BACKGROUND: Inappropriate implantable cardioverter-defibrillator (ICD) therapies remain a clinically important problem associated with morbidity and mortality. Tissue perfusion biomarkers, implemented to assist automated diagnosis of VF, sometimes mistake artifacts and random noise for perfusion, which could lead to shocks being inappropriately withheld. METHODS: The study tested a novel processing algorithm that combines electrogram data and laser Doppler perfusion monitoring as a method for assessing circulatory status. Fifty patients undergoing VF induction during ICD implantation were recruited. Noninvasive laser Doppler and continuous electrograms were recorded during both sinus rhythm and VF. Two additional scenarios that might have led to inappropriate shocks were simulated for each patient: ventricular lead fracture and T-wave oversensing. The laser Doppler was analyzed using 3 methods for reducing noise: 1) running mean; 2) oscillatory height; and 3) a novel quantification of electromechanical coupling which gates laser Doppler relative to electrograms. In addition, the algorithm was tested during exercise-induced sinus tachycardia. RESULTS: Only the electromechanical coupling algorithm found a clear perfusion cut off between sinus rhythm and VF (sensitivity and specificity of 100%). Sensitivity and specificity remained at 100% during simulated lead fracture and electrogram oversensing. (Area under the curve running mean: 0.91; oscillatory height: 0.86; electromechanical coupling: 1.00). Sinus tachycardia did not cause false positive results. CONCLUSIONS: Quantifying the coupling between electrical and perfusion signals increases reliability of discrimination between VF and artifacts that ICDs may interpret as VF. Incorporating such methods into future ICDs may safely permit reductions of inappropriate shocks.
format Online
Article
Text
id pubmed-6597902
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Elsevier Inc
record_format MEDLINE/PubMed
spelling pubmed-65979022019-07-11 Quantification of Electromechanical Coupling to Prevent Inappropriate Implantable Cardioverter-Defibrillator Shocks Keene, Daniel Shun-Shin, Matthew J. Arnold, Ahran D. Howard, James P. Lefroy, David Davies, D. Wyn Lim, Phang Boon Ng, Fu Siong Koa-Wing, Michael Qureshi, Norman A. Linton, Nick W.F. Shah, Jaymin S. Peters, Nicholas S. Kanagaratnam, Prapa Francis, Darrel P. Whinnett, Zachary I. JACC Clin Electrophysiol Article OBJECTIVES: This study sought to test specialized processing of laser Doppler signals for discriminating ventricular fibrillation (VF) from common causes of inappropriate therapies. BACKGROUND: Inappropriate implantable cardioverter-defibrillator (ICD) therapies remain a clinically important problem associated with morbidity and mortality. Tissue perfusion biomarkers, implemented to assist automated diagnosis of VF, sometimes mistake artifacts and random noise for perfusion, which could lead to shocks being inappropriately withheld. METHODS: The study tested a novel processing algorithm that combines electrogram data and laser Doppler perfusion monitoring as a method for assessing circulatory status. Fifty patients undergoing VF induction during ICD implantation were recruited. Noninvasive laser Doppler and continuous electrograms were recorded during both sinus rhythm and VF. Two additional scenarios that might have led to inappropriate shocks were simulated for each patient: ventricular lead fracture and T-wave oversensing. The laser Doppler was analyzed using 3 methods for reducing noise: 1) running mean; 2) oscillatory height; and 3) a novel quantification of electromechanical coupling which gates laser Doppler relative to electrograms. In addition, the algorithm was tested during exercise-induced sinus tachycardia. RESULTS: Only the electromechanical coupling algorithm found a clear perfusion cut off between sinus rhythm and VF (sensitivity and specificity of 100%). Sensitivity and specificity remained at 100% during simulated lead fracture and electrogram oversensing. (Area under the curve running mean: 0.91; oscillatory height: 0.86; electromechanical coupling: 1.00). Sinus tachycardia did not cause false positive results. CONCLUSIONS: Quantifying the coupling between electrical and perfusion signals increases reliability of discrimination between VF and artifacts that ICDs may interpret as VF. Incorporating such methods into future ICDs may safely permit reductions of inappropriate shocks. Elsevier Inc 2019-06 /pmc/articles/PMC6597902/ /pubmed/31221358 http://dx.doi.org/10.1016/j.jacep.2019.01.025 Text en Crown Copyright © Published by American College of Cardiology Foundation. All rights reserved. http://creativecommons.org/licenses/by/4.0/ This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Keene, Daniel
Shun-Shin, Matthew J.
Arnold, Ahran D.
Howard, James P.
Lefroy, David
Davies, D. Wyn
Lim, Phang Boon
Ng, Fu Siong
Koa-Wing, Michael
Qureshi, Norman A.
Linton, Nick W.F.
Shah, Jaymin S.
Peters, Nicholas S.
Kanagaratnam, Prapa
Francis, Darrel P.
Whinnett, Zachary I.
Quantification of Electromechanical Coupling to Prevent Inappropriate Implantable Cardioverter-Defibrillator Shocks
title Quantification of Electromechanical Coupling to Prevent Inappropriate Implantable Cardioverter-Defibrillator Shocks
title_full Quantification of Electromechanical Coupling to Prevent Inappropriate Implantable Cardioverter-Defibrillator Shocks
title_fullStr Quantification of Electromechanical Coupling to Prevent Inappropriate Implantable Cardioverter-Defibrillator Shocks
title_full_unstemmed Quantification of Electromechanical Coupling to Prevent Inappropriate Implantable Cardioverter-Defibrillator Shocks
title_short Quantification of Electromechanical Coupling to Prevent Inappropriate Implantable Cardioverter-Defibrillator Shocks
title_sort quantification of electromechanical coupling to prevent inappropriate implantable cardioverter-defibrillator shocks
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6597902/
https://www.ncbi.nlm.nih.gov/pubmed/31221358
http://dx.doi.org/10.1016/j.jacep.2019.01.025
work_keys_str_mv AT keenedaniel quantificationofelectromechanicalcouplingtopreventinappropriateimplantablecardioverterdefibrillatorshocks
AT shunshinmatthewj quantificationofelectromechanicalcouplingtopreventinappropriateimplantablecardioverterdefibrillatorshocks
AT arnoldahrand quantificationofelectromechanicalcouplingtopreventinappropriateimplantablecardioverterdefibrillatorshocks
AT howardjamesp quantificationofelectromechanicalcouplingtopreventinappropriateimplantablecardioverterdefibrillatorshocks
AT lefroydavid quantificationofelectromechanicalcouplingtopreventinappropriateimplantablecardioverterdefibrillatorshocks
AT daviesdwyn quantificationofelectromechanicalcouplingtopreventinappropriateimplantablecardioverterdefibrillatorshocks
AT limphangboon quantificationofelectromechanicalcouplingtopreventinappropriateimplantablecardioverterdefibrillatorshocks
AT ngfusiong quantificationofelectromechanicalcouplingtopreventinappropriateimplantablecardioverterdefibrillatorshocks
AT koawingmichael quantificationofelectromechanicalcouplingtopreventinappropriateimplantablecardioverterdefibrillatorshocks
AT qureshinormana quantificationofelectromechanicalcouplingtopreventinappropriateimplantablecardioverterdefibrillatorshocks
AT lintonnickwf quantificationofelectromechanicalcouplingtopreventinappropriateimplantablecardioverterdefibrillatorshocks
AT shahjaymins quantificationofelectromechanicalcouplingtopreventinappropriateimplantablecardioverterdefibrillatorshocks
AT petersnicholass quantificationofelectromechanicalcouplingtopreventinappropriateimplantablecardioverterdefibrillatorshocks
AT kanagaratnamprapa quantificationofelectromechanicalcouplingtopreventinappropriateimplantablecardioverterdefibrillatorshocks
AT francisdarrelp quantificationofelectromechanicalcouplingtopreventinappropriateimplantablecardioverterdefibrillatorshocks
AT whinnettzacharyi quantificationofelectromechanicalcouplingtopreventinappropriateimplantablecardioverterdefibrillatorshocks