Cargando…
Investigating the effectiveness of monitoring relevant variations during IMRT and VMAT treatments by EPID-based 3D in vivo verification performed using planning CTs
PURPOSE: The goal of this study was to investigate the effectiveness of monitoring relevant variations during treatments for electronic portal imaging device (EPID)-based 3D in vivo verification performed using planning CTs. METHODS: Experiments on two simple phantoms (uniform and nonuniform phantom...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6599132/ https://www.ncbi.nlm.nih.gov/pubmed/31251751 http://dx.doi.org/10.1371/journal.pone.0218803 |
_version_ | 1783430898745606144 |
---|---|
author | Li, Yinghui Zhu, Jinhan Shi, Jinping Chen, Lixin Liu, Xiaowei |
author_facet | Li, Yinghui Zhu, Jinhan Shi, Jinping Chen, Lixin Liu, Xiaowei |
author_sort | Li, Yinghui |
collection | PubMed |
description | PURPOSE: The goal of this study was to investigate the effectiveness of monitoring relevant variations during treatments for electronic portal imaging device (EPID)-based 3D in vivo verification performed using planning CTs. METHODS: Experiments on two simple phantoms (uniform and nonuniform phantoms) and a thoracic phantom were analyzed in this study, and six relevant variations including the machine output, planning target volume (PTV) deformation, multileaf collimator (MLC) and Phantom shift (set-up errors), and gantry and couch angle shifts were evaluated. 3D gamma and dose-volume histogram (DVH) methods were used to evaluate the detection sensitivity of the EPID-based 3D in vivo dosimetry and the dose accuracy of the EPID reconstruction, respectively, as affected by the variations, and the results were validated by determining the consistency with TPS simulated results. RESULTS: The results of the simple phantoms showed that the gamma failure rates and DVH trend of EPID reconstructions were consistent with the results of TPS simulations for machine output and MLC shifts and inconsistent for phantom shift, gantry/couch angle shift and PTV deformation variations. The results of the thoracic phantom showed that CBCT-guided EPID reconstruction sensitively detected 3-mm Phantom shift in thoracic phantom and its gamma failure rates and DVH trend were consistent with the results of TPS simulations. CONCLUSION: The variations, such as machine output and MLC shift, that are phantom unrelated and cause changes in the beam of the linear accelerator can be sensitively detected by EPID-based 3D in vivo dosimetry and do not affect the accuracy of the EPID reconstruction dose. Planning CT will limit the detection sensitivity and the accuracy of the reconstruction dose of the EPID-based 3D in vivo dosimetry for phantom-related variations (such as Phantom shift and gantry/couch angle shift). EPID reconstruction combined with IGRT technology is a more effective method to monitor phantom shift variations. |
format | Online Article Text |
id | pubmed-6599132 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-65991322019-07-12 Investigating the effectiveness of monitoring relevant variations during IMRT and VMAT treatments by EPID-based 3D in vivo verification performed using planning CTs Li, Yinghui Zhu, Jinhan Shi, Jinping Chen, Lixin Liu, Xiaowei PLoS One Research Article PURPOSE: The goal of this study was to investigate the effectiveness of monitoring relevant variations during treatments for electronic portal imaging device (EPID)-based 3D in vivo verification performed using planning CTs. METHODS: Experiments on two simple phantoms (uniform and nonuniform phantoms) and a thoracic phantom were analyzed in this study, and six relevant variations including the machine output, planning target volume (PTV) deformation, multileaf collimator (MLC) and Phantom shift (set-up errors), and gantry and couch angle shifts were evaluated. 3D gamma and dose-volume histogram (DVH) methods were used to evaluate the detection sensitivity of the EPID-based 3D in vivo dosimetry and the dose accuracy of the EPID reconstruction, respectively, as affected by the variations, and the results were validated by determining the consistency with TPS simulated results. RESULTS: The results of the simple phantoms showed that the gamma failure rates and DVH trend of EPID reconstructions were consistent with the results of TPS simulations for machine output and MLC shifts and inconsistent for phantom shift, gantry/couch angle shift and PTV deformation variations. The results of the thoracic phantom showed that CBCT-guided EPID reconstruction sensitively detected 3-mm Phantom shift in thoracic phantom and its gamma failure rates and DVH trend were consistent with the results of TPS simulations. CONCLUSION: The variations, such as machine output and MLC shift, that are phantom unrelated and cause changes in the beam of the linear accelerator can be sensitively detected by EPID-based 3D in vivo dosimetry and do not affect the accuracy of the EPID reconstruction dose. Planning CT will limit the detection sensitivity and the accuracy of the reconstruction dose of the EPID-based 3D in vivo dosimetry for phantom-related variations (such as Phantom shift and gantry/couch angle shift). EPID reconstruction combined with IGRT technology is a more effective method to monitor phantom shift variations. Public Library of Science 2019-06-28 /pmc/articles/PMC6599132/ /pubmed/31251751 http://dx.doi.org/10.1371/journal.pone.0218803 Text en © 2019 Li et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Research Article Li, Yinghui Zhu, Jinhan Shi, Jinping Chen, Lixin Liu, Xiaowei Investigating the effectiveness of monitoring relevant variations during IMRT and VMAT treatments by EPID-based 3D in vivo verification performed using planning CTs |
title | Investigating the effectiveness of monitoring relevant variations during IMRT and VMAT treatments by EPID-based 3D in vivo verification performed using planning CTs |
title_full | Investigating the effectiveness of monitoring relevant variations during IMRT and VMAT treatments by EPID-based 3D in vivo verification performed using planning CTs |
title_fullStr | Investigating the effectiveness of monitoring relevant variations during IMRT and VMAT treatments by EPID-based 3D in vivo verification performed using planning CTs |
title_full_unstemmed | Investigating the effectiveness of monitoring relevant variations during IMRT and VMAT treatments by EPID-based 3D in vivo verification performed using planning CTs |
title_short | Investigating the effectiveness of monitoring relevant variations during IMRT and VMAT treatments by EPID-based 3D in vivo verification performed using planning CTs |
title_sort | investigating the effectiveness of monitoring relevant variations during imrt and vmat treatments by epid-based 3d in vivo verification performed using planning cts |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6599132/ https://www.ncbi.nlm.nih.gov/pubmed/31251751 http://dx.doi.org/10.1371/journal.pone.0218803 |
work_keys_str_mv | AT liyinghui investigatingtheeffectivenessofmonitoringrelevantvariationsduringimrtandvmattreatmentsbyepidbased3dinvivoverificationperformedusingplanningcts AT zhujinhan investigatingtheeffectivenessofmonitoringrelevantvariationsduringimrtandvmattreatmentsbyepidbased3dinvivoverificationperformedusingplanningcts AT shijinping investigatingtheeffectivenessofmonitoringrelevantvariationsduringimrtandvmattreatmentsbyepidbased3dinvivoverificationperformedusingplanningcts AT chenlixin investigatingtheeffectivenessofmonitoringrelevantvariationsduringimrtandvmattreatmentsbyepidbased3dinvivoverificationperformedusingplanningcts AT liuxiaowei investigatingtheeffectivenessofmonitoringrelevantvariationsduringimrtandvmattreatmentsbyepidbased3dinvivoverificationperformedusingplanningcts |