Cargando…

The Dependence of Electrode Impedance on the Number of Performed EEG Examinations

In clinical practice, it is recommended to employ reusable electrodes for the registration of brain waves. Before registering EEG signals, the EEG technician checks the condition of all the electrodes, i.e., the occurrence of mechanical damage and the color of the electrode coating. It should be not...

Descripción completa

Detalles Bibliográficos
Autores principales: Górecka, Joanna, Makiewicz, Przemysław
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6603518/
https://www.ncbi.nlm.nih.gov/pubmed/31181738
http://dx.doi.org/10.3390/s19112608
_version_ 1783431523527032832
author Górecka, Joanna
Makiewicz, Przemysław
author_facet Górecka, Joanna
Makiewicz, Przemysław
author_sort Górecka, Joanna
collection PubMed
description In clinical practice, it is recommended to employ reusable electrodes for the registration of brain waves. Before registering EEG signals, the EEG technician checks the condition of all the electrodes, i.e., the occurrence of mechanical damage and the color of the electrode coating. It should be noticed that there is still no information on the permissible number of EEG examinations performed with one set of electrodes. After placement of the electrodes on the patient’s head, the scalp–electrode impedance is measured with the use of EEG equipment. When the scalp–electrode impedance achieves a value above 5 kΩ, it is necessary to replace the given electrode or to re-execute skin abrasion. The Electrochemical Impedance Spectroscopy (EIS) method was used in order to estimate the permissible number of EEG examinations performed with one set of electrodes. Ten new reusable electrodes were tested. Then, the tests were repeated after subsequent uses of those electrodes. The conducted tests led us to the conclusion that the permissible number of examinations performed with one set of electrodes is up to twenty except for the gold electrodes for which it is up to ten. Furthermore, the use of the EIS method revealed variability of impedance in the case of new electrodes.
format Online
Article
Text
id pubmed-6603518
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-66035182019-07-19 The Dependence of Electrode Impedance on the Number of Performed EEG Examinations Górecka, Joanna Makiewicz, Przemysław Sensors (Basel) Article In clinical practice, it is recommended to employ reusable electrodes for the registration of brain waves. Before registering EEG signals, the EEG technician checks the condition of all the electrodes, i.e., the occurrence of mechanical damage and the color of the electrode coating. It should be noticed that there is still no information on the permissible number of EEG examinations performed with one set of electrodes. After placement of the electrodes on the patient’s head, the scalp–electrode impedance is measured with the use of EEG equipment. When the scalp–electrode impedance achieves a value above 5 kΩ, it is necessary to replace the given electrode or to re-execute skin abrasion. The Electrochemical Impedance Spectroscopy (EIS) method was used in order to estimate the permissible number of EEG examinations performed with one set of electrodes. Ten new reusable electrodes were tested. Then, the tests were repeated after subsequent uses of those electrodes. The conducted tests led us to the conclusion that the permissible number of examinations performed with one set of electrodes is up to twenty except for the gold electrodes for which it is up to ten. Furthermore, the use of the EIS method revealed variability of impedance in the case of new electrodes. MDPI 2019-06-08 /pmc/articles/PMC6603518/ /pubmed/31181738 http://dx.doi.org/10.3390/s19112608 Text en © 2019 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Górecka, Joanna
Makiewicz, Przemysław
The Dependence of Electrode Impedance on the Number of Performed EEG Examinations
title The Dependence of Electrode Impedance on the Number of Performed EEG Examinations
title_full The Dependence of Electrode Impedance on the Number of Performed EEG Examinations
title_fullStr The Dependence of Electrode Impedance on the Number of Performed EEG Examinations
title_full_unstemmed The Dependence of Electrode Impedance on the Number of Performed EEG Examinations
title_short The Dependence of Electrode Impedance on the Number of Performed EEG Examinations
title_sort dependence of electrode impedance on the number of performed eeg examinations
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6603518/
https://www.ncbi.nlm.nih.gov/pubmed/31181738
http://dx.doi.org/10.3390/s19112608
work_keys_str_mv AT goreckajoanna thedependenceofelectrodeimpedanceonthenumberofperformedeegexaminations
AT makiewiczprzemysław thedependenceofelectrodeimpedanceonthenumberofperformedeegexaminations
AT goreckajoanna dependenceofelectrodeimpedanceonthenumberofperformedeegexaminations
AT makiewiczprzemysław dependenceofelectrodeimpedanceonthenumberofperformedeegexaminations