Enhanced release of calcium phosphate additives from bioresorbable orthopaedic devices using irradiation technology is non-beneficial in a rabbit model: An animal study

OBJECTIVES: Bioresorbable orthopaedic devices with calcium phosphate (CaP) fillers are commercially available on the assumption that increased calcium (Ca) locally drives new bone formation, but the clinical benefits are unknown. Electron beam (EB) irradiation of polymer devices has been shown to en...

Descripción completa

Detalles Bibliográficos
Autores principales: Palmer, I., Clarke, S. A., Buchanan, F. J
Formato: Online Artículo Texto
Lenguaje:English
Publicado: 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6609887/
https://www.ncbi.nlm.nih.gov/pubmed/31346455
http://dx.doi.org/10.1302/2046-3758.86.BJR-2018-0224.R2
_version_ 1783432402189680640
author Palmer, I.
Clarke, S. A.
Buchanan, F. J
author_facet Palmer, I.
Clarke, S. A.
Buchanan, F. J
author_sort Palmer, I.
collection PubMed
description OBJECTIVES: Bioresorbable orthopaedic devices with calcium phosphate (CaP) fillers are commercially available on the assumption that increased calcium (Ca) locally drives new bone formation, but the clinical benefits are unknown. Electron beam (EB) irradiation of polymer devices has been shown to enhance the release of Ca. The aims of this study were to: 1) establish the biological safety of EB surface-modified bioresorbable devices; 2) test the release kinetics of CaP from a polymer device; and 3) establish any subsequent beneficial effects on bone repair in vivo. METHODS: ActivaScrew Interference (Bioretec Ltd, Tampere, Finland) and poly(L-lactide-co-glycolide) (PLGA) orthopaedic screws containing 10 wt% β-tricalcium phosphate (β-TCP) underwent EB treatment. In vitro degradation over 36 weeks was investigated by recording mass loss, pH change, and Ca release. Implant performance was investigated in vivo over 36 weeks using a lapine femoral condyle model. Bone growth and osteoclast activity were assessed by histology and enzyme histochemistry. RESULTS: Calcium release doubled in the EB-treated group before returning to a level seen in untreated samples at 28 weeks. Extensive bone growth was observed around the perimeter of all implant types, along with limited osteoclastic activity. No statistically significant differences between comparative groups was identified. CONCLUSION: The higher than normal dose of EB used for surface modification did not adversely affect tissue response around implants in vivo. Surprisingly, incorporation of β-TCP and the subsequent accelerated release of Ca had no significant effect on in vivo implant performance, calling into question the clinical evidence base for these commercially available devices. Cite this article: I. Palmer, S. A. Clarke, F. J Buchanan. Enhanced release of calcium phosphate additives from bioresorbable orthopaedic devices using irradiation technology is non-beneficial in a rabbit model: An animal study. Bone Joint Res 2019;8:266–274. DOI: 10.1302/2046-3758.86.BJR-2018-0224.R2.
format Online
Article
Text
id pubmed-6609887
institution National Center for Biotechnology Information
language English
publishDate 2019
record_format MEDLINE/PubMed
spelling pubmed-66098872019-07-25 Enhanced release of calcium phosphate additives from bioresorbable orthopaedic devices using irradiation technology is non-beneficial in a rabbit model: An animal study Palmer, I. Clarke, S. A. Buchanan, F. J Bone Joint Res Biomaterials OBJECTIVES: Bioresorbable orthopaedic devices with calcium phosphate (CaP) fillers are commercially available on the assumption that increased calcium (Ca) locally drives new bone formation, but the clinical benefits are unknown. Electron beam (EB) irradiation of polymer devices has been shown to enhance the release of Ca. The aims of this study were to: 1) establish the biological safety of EB surface-modified bioresorbable devices; 2) test the release kinetics of CaP from a polymer device; and 3) establish any subsequent beneficial effects on bone repair in vivo. METHODS: ActivaScrew Interference (Bioretec Ltd, Tampere, Finland) and poly(L-lactide-co-glycolide) (PLGA) orthopaedic screws containing 10 wt% β-tricalcium phosphate (β-TCP) underwent EB treatment. In vitro degradation over 36 weeks was investigated by recording mass loss, pH change, and Ca release. Implant performance was investigated in vivo over 36 weeks using a lapine femoral condyle model. Bone growth and osteoclast activity were assessed by histology and enzyme histochemistry. RESULTS: Calcium release doubled in the EB-treated group before returning to a level seen in untreated samples at 28 weeks. Extensive bone growth was observed around the perimeter of all implant types, along with limited osteoclastic activity. No statistically significant differences between comparative groups was identified. CONCLUSION: The higher than normal dose of EB used for surface modification did not adversely affect tissue response around implants in vivo. Surprisingly, incorporation of β-TCP and the subsequent accelerated release of Ca had no significant effect on in vivo implant performance, calling into question the clinical evidence base for these commercially available devices. Cite this article: I. Palmer, S. A. Clarke, F. J Buchanan. Enhanced release of calcium phosphate additives from bioresorbable orthopaedic devices using irradiation technology is non-beneficial in a rabbit model: An animal study. Bone Joint Res 2019;8:266–274. DOI: 10.1302/2046-3758.86.BJR-2018-0224.R2. 2019-07-05 /pmc/articles/PMC6609887/ /pubmed/31346455 http://dx.doi.org/10.1302/2046-3758.86.BJR-2018-0224.R2 Text en © 2019 Author(s) et al. https://creativecommons.org/licenses/by-nc/4.0/ This article distributed under the terms of the Creative Commons Attribution-Non Commercial 4.0 international (CC-BY-NC 4.0) licence (https://creativecommons.org/licences/by-nc/4.0/), which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed.
spellingShingle Biomaterials
Palmer, I.
Clarke, S. A.
Buchanan, F. J
Enhanced release of calcium phosphate additives from bioresorbable orthopaedic devices using irradiation technology is non-beneficial in a rabbit model: An animal study
title Enhanced release of calcium phosphate additives from bioresorbable orthopaedic devices using irradiation technology is non-beneficial in a rabbit model: An animal study
title_full Enhanced release of calcium phosphate additives from bioresorbable orthopaedic devices using irradiation technology is non-beneficial in a rabbit model: An animal study
title_fullStr Enhanced release of calcium phosphate additives from bioresorbable orthopaedic devices using irradiation technology is non-beneficial in a rabbit model: An animal study
title_full_unstemmed Enhanced release of calcium phosphate additives from bioresorbable orthopaedic devices using irradiation technology is non-beneficial in a rabbit model: An animal study
title_short Enhanced release of calcium phosphate additives from bioresorbable orthopaedic devices using irradiation technology is non-beneficial in a rabbit model: An animal study
title_sort enhanced release of calcium phosphate additives from bioresorbable orthopaedic devices using irradiation technology is non-beneficial in a rabbit model: an animal study
topic Biomaterials
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6609887/
https://www.ncbi.nlm.nih.gov/pubmed/31346455
http://dx.doi.org/10.1302/2046-3758.86.BJR-2018-0224.R2
work_keys_str_mv AT palmeri enhancedreleaseofcalciumphosphateadditivesfrombioresorbableorthopaedicdevicesusingirradiationtechnologyisnonbeneficialinarabbitmodelananimalstudy
AT clarkesa enhancedreleaseofcalciumphosphateadditivesfrombioresorbableorthopaedicdevicesusingirradiationtechnologyisnonbeneficialinarabbitmodelananimalstudy
AT buchananfj enhancedreleaseofcalciumphosphateadditivesfrombioresorbableorthopaedicdevicesusingirradiationtechnologyisnonbeneficialinarabbitmodelananimalstudy